Chapter 10 - Care & Maintenance
10-9
10-5-5
Physical Inspection
Table 10-9 Physical Checks
Step
Item
Description
1
Labeling
Verify that all system labeling is present and in readable condition.
2
Scratches & Dents
Inspect the console for dents, scratches or cracks.
3
Control Panel
Inspect keyboard and control panel. Note any damaged or missing items.
4
Control Panel
Movement
Verify ease of control panel (Operator I/O Panel) movement in acceptable direction. Ensure that
it latches in position as required.
5
Wheels & Brakes
Check all wheels and casters for wear and verify operation of foot brake, to stop the unit from
moving, and release mechanism. Check all wheel locks and wheel swivel locks for proper
operation.
6
Cables & Connectors Check all internal cable harnesses and connectors for wear and secure connector seating.
Pay special attention to footswitch assembly and probe strain or bend reliefs.
7
Shielding & Covers Check to ensure that all EMI shielding, internal covers, air flow panels and screws are in place.
Missing covers and hardware could cause EMI/RFI problems while scanning.
8
External I/O
Check all connectors for damage and verify that the labeling is good.
9
Op Panel Lights
Check for proper operation of all operator panel key illuminations (flash once during system start-
up).
10
Monitor Light
Check for proper operation of any monitor lights if available.
11
External Microphone Check for proper operation of any external microphones by recording an audio test if available.
NOTICE