background image

GE M

EDICAL

 S

YSTEMS

 - K

RETZTECHNIK

DRAFT

V

OLUSON

® 730P

RO

 / 730P

RO

V (BT´04)

D

IRECTION

 KTI105947, R

EVISION

 

2

 

S

ERVICE

 M

ANUAL

Chapter 5 -  Components and Functions (Theory)

5-41

Section 5-10

Peripherals

5-10-1

General Information

The GEM Module contains the DVD/CD+(R)W Drive.
Additionally the Magneto Optical Drive and the ECG-preamplifier (MAN6) can be installed as an Option.

5-10-1-1

ECG-preamplifier (MAN6 - optional)

The ECG-preamplifier is used for acquiring an ECG-signal to be displayed with the ultrasound image. 
This optional peripheral serves for gaining an ECG-signal to mark the systolic and end diastolic 
moments in M-Mode and Doppler evaluations.

The ECG-preamplifier must not be used for ECG-diagnostics. It is not intended for use as a cardiac 
monitor and must not be used for an intraoperative application on the heart.

5-10-1-2

DVD/CD+(R)W Drive

Recordable and Re-writable DVD´s and CD´s are ideal for any storage purpose and offer complete 
security and reliability for important data.

5-10-1-3

Magneto-Optical Drive (optional)

Storage capacity by disk: 1.3GB, 640MB, 540MB, 230MB, 128MB

The MO-Drive allows to read and write any GIGAMO standard 1.3GB disk at twice the liner bit density. 
Additionally it retains full read/write compatibility with ISO/IEC 3.5 - inch disks ranging from original 
128MB to current 640MB.

MO disks are nearly indestructible and immune from the problems that plague magnetic media.
MO disks can be rewritten an unlimited number of times.

Figure 5-24  GEM incl. optional MOD and ECG

Read Speed: 

max. 12x DVD ROM (max. 16200 kByte/s); max. 40x CD ROM (max. 6000 kByte/s) 

Write Speed: 

DVD+R:  4x CLV (5520 kByte/s), DVD+RW:  2.4x CLV (3300 kByte/s), 
CD-R:  16x CLV (2400 kByte/s), CD-RW:  10x CLV (1500 kByte/s)

Access Time:  DVD: 140ms; CD: 120ms

optional ECG (MAN6)

DVD+(R)W Drive

optional 
Magneto-Optical Drive (MOD)

Содержание Voluson 730Pro

Страница 1: ...l GE Medical Systems Kretz Ultrasound 0123 Copyright 2004 2005 by General Electric Co Voluson 730Pro 730ProV systems with Serial number A31501 A33999 Voluson 730Pro 730ProV systems with Software version SW 4 x x Voluson 730Pro 730ProV systems that were upgraded to BT 04 Voluson 730Pro 730ProV systems upgraded to BT 05 SW 5 x x ...

Страница 2: ......

Страница 3: ... DANS UNE AUTRE LANGUE QUE L ANGLAIS C EST AU CLIENT QU IL INCOMBE DE LE FAIRE TRADUIRE NE PAS TENTER D INTERVENTION SUR LES ÉQUIPEMENTS TANT QUE LE MANUEL SERVICE N A PAS ÉTÉ CONSULTÉ ET COMPRIS LE NON RESPECT DE CET AVERTISSEMENT PEUT ENTRAÎNER CHEZ LE TECHNICIEN L OPÉRATEUR OU LE PATIENT DES BLESSURES DUES À DES DANGERS ÉLECTRIQUES MÉCANIQUES OU AUTRES DIESES KUNDENDIENST HANDBUCH EXISTIERT NUR...

Страница 4: ...A GEMS SOLICITAR ESTES MANUAIS NOUTRO IDIOMA É DA RESPONSABILIDADE DO CLIENTE FORNECER OS SERVIÇOS DE TRADUÇÃO NÃO TENTE REPARAR O EQUIPAMENTO SEM TER CONSULTADO E COMPREENDIDO ESTE MANUAL DE ASSISTÊNCIA TÉCNICA O NÃO CUMPRIMENTO DESTE AVISO PODE POR EM PERIGO A SEGURANÇA DO TÉCNICO OPERADOR OU PACIENTE DEVIDO A CHOQUES ELÉTRICOS MECÂNICOS OU OUTROS IL PRESENTE MANUALE DI MANUTENZIONE È DISPONIBIL...

Страница 5: ...GE MEDICAL SYSTEMS KRETZTECHNIK VOLUSON 730PRO 730PROV BT 04 DIRECTION KTI105947 REVISION 2 SERVICE MANUAL Important Precautions iii ...

Страница 6: ...e performed by licensed electrical contractors Other connections between pieces of electrical equipment calibrations and testing shall be performed by qualified GE Healthcare personnel In performing all electrical work on these products GE will use its own specially trained field engineers All of GE s electrical work on these products will comply with the requirements of the applicable electrical ...

Страница 7: ...sed A license to use operating or basic service software does not extend to or cover this software or related documentation If you are a GE Medical Systems employee or a customer who has entered into such a license agreement with GE Medical Systems to use this proprietary software you are authorized to use this Material according to the conditions stated in your license agreement However you do no...

Страница 8: ...GE MEDICAL SYSTEMS KRETZTECHNIK VOLUSON 730PRO 730PROV BT 04 DIRECTION KTI105947 REVISION 2 SERVICE MANUAL vi Important Precautions This page was intentionally left blank ...

Страница 9: ...ion Title Page 2 Chapter 2 Pre Installation pages 2 1 to 2 10 2 Chapter 7 Diagnostics Troubleshooting pages 7 1 to 7 32 2 Important Precautions pages i to vi 2 Chapter 3 Installation pages 3 1 to 3 52 2 Chapter 8 Replacement Procedures pages 8 1 to 8 24 2 Rev History LOEP pages vii to viii 2 Chapter 4 Functional Checks pages 4 1 to 4 46 2 Chapter 9 Replacement Parts pages 9 1 to 9 36 2 Table of Co...

Страница 10: ...GE MEDICAL SYSTEMS KRETZTECHNIK VOLUSON 730PRO 730PROV BT 04 DIRECTION KTI105947 REVISION 2 SERVICE MANUAL viii Revision History This page was intentionally left blank ...

Страница 11: ...Conventions 1 3 Conventions Used in Book 1 3 Standard Hazard Icons 1 4 Product Icons 1 5 Safety Considerations 1 7 Introduction 1 7 Human Safety 1 7 Mechanical Safety 1 7 Electrical Safety 1 8 Safe Practices 1 8 Probes 1 8 Labels Locations 1 9 Dangerous Procedure Warnings 1 9 Lockout Tagout Requirements For USA Only 1 10 Returning Shipping Probes and Repair Parts 1 10 Electromagnetic Compatibility...

Страница 12: ...irements 2 2 Voluson 730Pro 730ProV Power Requirements 2 2 Inrush Current 2 2 Site Circuit Breaker 2 3 Site Power Outlets 2 3 Unit Power Plug 2 3 EMI Limitations 2 4 Scan Probe Environmental Requirements 2 5 Time and Manpower Requirements 2 5 Facility Needs 2 6 Purchaser Responsibilities 2 6 Required Features 2 6 Desirable Features 2 7 Minimal Floor Plan Suggestion 2 8 Networking Pre installation ...

Страница 13: ...ion 3 6 System Voltage Settings 3 6 EMI Protection 3 6 Connection of Auxiliary Devices 3 7 Monitor Connection 3 9 Foot Switch Connection 3 9 ECG preamplifier Connection 3 10 Global Modem Connection 3 11 S VHS Video Recorder Connection 3 12 Mitsubishi HS MD3000 3 12 Sony SVO 9500MD 3 13 B W Video Printer Connection 3 14 Line Printer Connection 3 15 Digital Color Printer Connection 3 16 Bluetooth Pr...

Страница 14: ...i Job Jet 210 Printer Settings 3 37 Printer Remote Control Selection 3 38 System Configuration 3 39 System Specifications 3 39 Physical Dimensions of Voluson 730Pro 730ProV 3 39 Weight without Monitor and Peripherals 3 39 Acoustic Noise Output 3 39 Electrical Specifications 3 39 System Setup 3 40 How to enter Date and Time 3 41 How to enter Hospital Name 3 41 How to change Language and or EUM Lang...

Страница 15: ... Check 4 12 Spectral Doppler Mode Checks 4 13 Color Doppler Mode Checks 4 15 Volume Mode Checks 4 17 Pre Volume Mode Functions 4 17 Functions after the 3D Acquisition 4 18 Sub Menu 4 20 Using Cine 4 21 Activating Cine 4 21 Cine Split Function Multiple Format 4 21 Activating 2D Auto Cine 4 21 Spectral Doppler or M Cine Loop 4 21 Activating 3D Rotation Cine 4 21 Activating 4D Cine 4 21 Generic Measu...

Страница 16: ...Database Preset Configurations and Images 4 31 Save User Settings Only Application Settings 4 32 Load User Settings Only Application Settings 4 33 Save Full Backup Presets Configurations Application Settings 4 35 Load Full Backup Presets Configurations Application Settings 4 37 Delete Full Backup Presets Configurations Application Settings 4 39 Archiving Images 4 40 Software Configuration Checks 4...

Страница 17: ...Mode 5 11 Special B Mode Techniques 5 12 M Mode 5 12 D Mode Pulsed Wave and Continuous Wave Doppler 5 13 D Mode Autotrace draws PC calculated envelope to D Spectrum 5 13 CFM Mode Color Flow Mode 5 14 3D Mode Freezes after 1 volume sweep 5 14 Real Time 4D Mode nonstop volume rendering 5 14 XBeam CRI Mode CrossBeam Compound Resolution Imaging 5 14 Extern Video Mode display Video from Video Recorder ...

Страница 18: ...over Board 5 24 CPK Motherboard of GEF Module 5 24 CRS Signal Processor Board 5 25 CPP Power Supply Secondary Board Motor Power stage 5 25 BackEnd Processor 5 27 Block diagram CKV 5 28 BackEnd Board Descriptions 5 29 SBC Single Board Computer 5 29 PCI VGA Card 5 29 CKV DMA Controller Video Card 5 29 Hard Disk Drive 5 29 CPE Back Panel I O Card 5 29 CPM PC Motherboard Card 5 29 CPP Power Supply Sec...

Страница 19: ...k Drive Module GEM 5 45 Mechanical Descriptions 5 46 Physical Dimensions 5 46 Monitor 5 47 Top Console Positioning 5 47 Rotation of the Control Console 5 47 Assembly Drawing GW GEU Monitor 5 48 Air Flow Control 5 49 Air Flow Distribution 5 49 Air Flow Distribution Overview 5 49 Service Platform 5 50 Introduction 5 50 Access Security 5 50 iLinq Interactive Platform 5 51 Service Page 5 52 Introducti...

Страница 20: ...e Adjustments Overview 6 1 Purpose of Chapter 6 6 1 Regulatory 6 1 Monitor Adjustment 6 2 Monitor Calibration 6 2 Degauss 6 3 Color Temp Calibration 6 3 Contrast and Brightness 6 4 Geometry 6 4 Convergence 6 4 Color temperature 6 4 B Mode Quality 6 4 Control Console Transport Lock 6 5 Control Console 6 5 Transport Lock 6 5 Trackball Adjustment 6 6 ...

Страница 21: ...ey 7 6 Export Log s and System Data via Service Page 7 7 Dump file 7 7 How to use the Auto Tester program 7 8 Minimum Configuration to Boot Scan 7 11 Booting without using the Ultrasound Hardware 7 11 Minimum Configuration to Scan 7 12 Troubleshooting Trees and Instructions 7 13 System Does Not Power On Boot Up 7 13 Noise disturbs the Image 7 14 Trackball 7 15 System Does Not Power Off Shutdown 7 ...

Страница 22: ...ing Procedure 8 8 Full Backup Presets Configurations Appl Settings Loading Procedure 8 8 Introduction 8 8 Loading Procedure 8 8 Service Platform SSW Upgrade Procedure 8 9 Manpower 8 9 Tools 8 9 Upgrade Procedure 8 9 Electronic User Manual EUM Upgrade Procedure 8 11 Manpower 8 11 Tools 8 11 Preparations 8 11 EUM Upgrade Procedure 8 12 Software and Functional Checks after the Upgrade 8 14 Replacemen...

Страница 23: ... the new system 8 19 Replacement of the Monitor Task Lamp 8 22 Manpower 8 22 Tools 8 22 Preparations 8 22 Task Lamp Removal Procedure 8 22 Task Lamp Installation Procedure 8 22 Replacement of the Trackball top fixation ring 8 23 Manpower 8 23 Trackball top fixation ring Replacement Procedure 8 23 Replacement of Digipots and TGC Slider controls 8 23 Manpower 8 23 Tools 8 23 Cap and or Spring Replac...

Страница 24: ...itional Console Hardware 9 4 User Interface GEU Top Console 9 7 Monitor Monitor replacement parts 9 9 Disk Drives GEM 9 10 Main Power Module CPN 9 11 Main Board Module GEF 9 13 FrontEnd US Part 9 14 BackEnd Processor PC Part 9 16 Options and Upgrades 9 18 Miscellaneous Cables 9 19 Optional Peripherals and Accessories 9 25 Probes 9 29 2D Probes 9 29 9 13 1 2D Probes cont d 9 30 Real Time 4D Volume ...

Страница 25: ...ements for Care Maintenance 10 5 System Maintenance 10 6 Preliminary Checks 10 6 Functional Checks see also Chapter 4 10 7 System Checks 10 7 Peripheral Option Checks 10 8 Input Power 10 8 Mains Cable Inspection 10 8 Cleaning 10 8 General Cleaning 10 8 Physical Inspection 10 9 Optional Diagnostic Checks 10 10 Probe Maintenance 10 10 Probe Related Checks 10 10 Basic Probe Care 10 10 Basic Probe Cle...

Страница 26: ... Leakage Lead to Ground 10 17 Definition 10 17 Generic Procedure 10 18 Isolated Patient Lead Source Leakage Lead to Lead 10 18 Isolated Patient Lead Sink Leakage Isolation Test 10 19 Data Sheet for ECG Leakage Current 10 20 Probe Leakage Current Test 10 21 Definition 10 21 Generic Procedure 10 21 No Meter Probe Adapter Procedure 10 22 Data Sheet for Transducer Source Leakage Current 10 23 When The...

Страница 27: ...nal Checks Contains functional checks that are recommended as part of the installation or as required during servicing and periodic maintenance 5 Chapter 5 Components and Functions Theory Contains block diagrams and functional explanations of the electronics 6 Chapter 6 Service Adjustments Contains instructions on how to make available adjustments to the Voluson 730Pro 730ProV 7 Chapter 7 Diagnost...

Страница 28: ...0ProV BT 04 Console 100V JAPAN H48651BE Voluson 730Pro 730ProV BT 04 Console 230V KOREA H48651BF Voluson 730Pro 730ProV BT 04 Console 230V CHINA NOTICE This manual applies to Voluson 730Pro 730ProV systems with Serial number A31501 33999 Voluson 730Pro 730ProV systems with Software version 4 x x installed Voluson 730Pro 730ProV systems that were upgraded to BT 04 and or Voluson 730Pro 730ProV syst...

Страница 29: ...age Known or potential hazards are labeled in one of following ways NOTE Notes provide important information about an item or a procedure Information contained in a NOTE can often save you time or effort DANGER DANGER IS USED TO INDICATE THE PRESENCE OF A HAZARD THAT WILL CAUSE SEVERE PERSONAL INJURY OR DEATH IF THE INSTRUCTIONS ARE IGNORED WARNING WARNING WARNING IS USED TO INDICATE THE PRESENCE ...

Страница 30: ...t this chapter In addition to text several different graphical icons symbols may be used to make you aware of specific types of hazards that could cause harm Other hazard icons make you aware of specific procedures that should be followed Table 1 3 Standard Hazard Icons ELECTRICAL MECHANICAL RADIATION LASER HEAT PINCH Table 1 4 Standard Icons Indicating a Special Procedure be Used AVOID STATIC ELE...

Страница 31: ...e Class Label Used to indicate the degree of safety or protection IP Code IPX 1 IP Code IPX 7 Indicates the degree of protection provided by the enclosure per IEC 529 IPX 1 and IPX 7 indicates drip proof Footswitch Probes Equipment Type BF man in the box symbol IEC 878 02 03 indicates B Type equipment having even more electrical isolation than standard Type B equipment because it is intended for i...

Страница 32: ...jacent to On Off Standby switch left below the Control panel Mains ON Indicates the power on position of the mains power switch Rear of system at mains switch F1 ON switch of the isolation transformer for auxiliary devices Rear of system at the switch for auxiliary devices F2 OFF switch of the isolation transformer for auxiliary devices Rear of system at the switch for auxiliary devices F2 Protect...

Страница 33: ...f your equipment Rear side of the unit on each probe WARNING WARNING USE EXTREME CAUTION WHEN ELEVATING THE UNIT OR IF IT IS RAISED FOR A REPAIR OR MOVED ALONG ANY INCLINE IT MAY BECOME UNSTABLE WHICH COULD CAUSE THE UNIT TO TIP OVER WARNING WARNING ULTRASOUND PROBES ARE HIGHLY SENSITIVE MEDICAL INSTRUMENTS THAT CAN EASILY BE DAMAGED BY IMPROPER HANDLING USE CARE WHEN HANDLING AND PROTECT FROM DAM...

Страница 34: ...nto an approved electrical outlet with safety ground If an extension cord is used with the system make sure that the total current rating of the system does not exceed the extension cord rating The power outlet used for this equipment should not be shared with other types of equipment Both the system power cable and the power connector meet international electrical standards 1 3 4 2 Probes Follow ...

Страница 35: ...hout this manual Instructions contained in the warnings must be followed Figure 1 1 Labeling DANGER DANGEROUS VOLTAGES CAPABLE OF CAUSING DEATH ARE PRESENT IN THIS EQUIPMENT USE EXTREME CAUTION WHEN HANDLING TESTING AND ADJUSTING WARNING WARNING EXPLOSION WARNING DO NOT OPERATE THE EQUIPMENT IN AN EXPLOSIVE ATMOSPHERE OPERATION OF ANY ELECTRICAL EQUIPMENT IN SUCH AN ENVIRONMENT CONSTITUTES A DEFIN...

Страница 36: ...ial Section 1 4 Electromagnetic Compatibility EMC 1 4 1 What is EMC Electromagnetic compatibility describes a level of performance of a device within its electromagnetic environment This environment consists of the device itself and its surroundings including other equipment power sources and persons with which the device must interface Inadequate compatibility results when a susceptible device fa...

Страница 37: ...ual or if you require additional assistance please contact the local distributor or appropriate support resource as listed below NOTE Prepare vital system information see Section 7 2 on page 7 2 before you call System Type System Serial number also visible on label on back of the system Application Software version Backup version additional information about installed software WARNING WARNING DO N...

Страница 38: ...edical Systems Ultrasound Service Engineering 4855 W Electric Avenue Milwaukee WI 53219 Customer Answer Center 1 262 524 5300 1 262 524 5698 Fax 1 414 647 4125 Europe GE Medical Systems Kretztechnik GmbH Co OHG Service Department Ultrasound Tiefenbach 15 A 4871 Zipf Austria Customer Answer Center Tel 43 0 7682 3800 26 Fax 43 0 7682 3800 47 Tel 33 13083 1300 Asia GE Ultrasound Asia Service Departme...

Страница 39: ...iew 2 1 1 Purpose of Chapter 2 This chapter provides the information required to plan and prepare for the installation of a Voluson 730Pro 730ProV Included are descriptions of the facility and electrical needs to be met by the purchaser of the unit Table 2 1 Contents in Chapter 2 Section Description Page Number 2 1 Overview 2 1 2 2 General Console Requirements 2 2 2 3 Facility Needs 2 6 ...

Страница 40: ...r and ground for the proper operation of its Ultrasound equipment This dedicated power shall originate at the last distribution panel before the system The dedicated line shall consist of one phase a neutral not shared with any other circuit and a full size ground wire from the distribution panel to the Ultrasound outlet Please note that image artifacts can occur if at any time within the facility...

Страница 41: ...medical and test equipment needed to support this unit must also be present within 1 m 3 2 ft of the unit Electrical installation must meet all current local state and national electrical codes 2 2 2 5 Unit Power Plug If the unit arrives without a power plug or with the wrong plug you must contact your GE dealer or the installation engineer must supply what is locally required CAUTION POWER OUTAGE...

Страница 42: ...ound the unit Poor grounding is the most likely reason a unit will have noisy images Check grounding of the power cord and power outlet Replace all screws RF gaskets covers cores After you finish repairing or updating the system replace all covers and tighten all screws Any cable with an external connection requires a magnet wrap at each end Install the shield over the front of card cage Loose or ...

Страница 43: ...n takes time Begin Pre installation checks as soon as possible If possible allow six weeks before delivery for enough time to make necessary changes NOTICE SYSTEMS AND ELECTRONIC PROBES ARE DESIGNED FOR STORAGE TEMPERATURES OF 10 TO 50 degrees C WHEN EXPOSED TO LARGE TEMPERATURE VARIATIONS THE PRODUCT SHOULD BE KEPT IN ROOM TEMPERATURE FOR 10 HOURS BEFORE USE CAUTION Have two people available to d...

Страница 44: ...n listed or customer provided product or to place an approved product further from the system than the interface kit allows presents challenges to the installation team To avoid delays during installation such variances should be made known to the individuals or group performing the installation at the earliest possible date preferably prior to the purchase The ultrasound suite must be clean prior...

Страница 45: ...east 90 cm 3 ft wide Circuit breaker for dedicated power outlet is easily accessible Sink with hot and cold water Receptacle for bio hazardous waste like used probe sheaths Emergency oxygen supply Storage for linens and equipment Nearby waiting room lavatory and dressing room Dual level lighting bright and dim Lockable cabinet ordered by GE for its software and proprietary manuals ...

Страница 46: ...MEDICAL SYSTEMS KRETZTECHNIK RAFT VOLUSON 730PRO 730PROV BT 04 DIRECTION KTI105947 REVISION 2 SERVICE MANUAL 2 8 Section 2 3 Facility Needs 2 3 4 Minimal Floor Plan Suggestion Figure 2 1 Minimal Floor Plan ...

Страница 47: ...dded benefit transferring images in this manner frees up the on board monitor and peripherals enabling viewing to be done while scanning continues With DICOM images can be archived stored and retrieved faster easier and at a lower cost 2 3 5 2 DICOM Option Pre installation Requirements To configure the Voluson 730Pro 730ProV to work with other network connections the site s network administrator m...

Страница 48: ...GE MEDICAL SYSTEMS KRETZTECHNIK RAFT VOLUSON 730PRO 730PROV BT 04 DIRECTION KTI105947 REVISION 2 SERVICE MANUAL 2 10 Section 2 3 Facility Needs This page was intentionally left blank ...

Страница 49: ... 3 1 3 2 Installation Reminders 3 1 3 3 Receiving and Unpacking the Equipment 3 4 3 4 Preparing for Installation 3 6 3 5 Connection of Auxiliary Devices 3 7 3 6 Completing the Installation 3 20 3 7 Printer Installation 3 23 3 8 System Configuration 3 39 3 9 Available Probes 3 47 3 10 Software Option Configuration 3 48 3 11 Connectivity Installation Worksheet 3 49 3 12 Network IP Address Configurat...

Страница 50: ...e Operator I O Panel in it s extended position the console could tip over CAUTION Equipment damage possibility Turning the system on without acclimation after arriving at site may cause the system to be damaged Table 3 3 Acclimation Time C 60 55 50 45 40 35 30 25 20 15 10 5 0 5 10 15 20 25 30 35 40 F 140 131 122 113 104 96 86 77 68 59 50 41 32 23 14 5 4 13 22 31 40 hrs 8 6 4 2 0 0 0 0 0 0 0 2 4 6 ...

Страница 51: ...r outlet Do not use a three to two prong adapter This defeats safety grounding CAUTION Do NOT wear the ESD wrist strap when you work on live circuits and more than 30 V peak is present CAUTION Do not use a 20 Amp to 15 Amp adapter on the 120 Vac unit s power cord This unit requires a dedicated 16 A circuit CAUTION Do not operate this unit unless all board covers and frame panels are securely in pl...

Страница 52: ...ine by one person could result in injury or damage or both Figure 3 3 envelope at front panel of the crate NOTICE The device must only be transported in the original packaging Unpack the devices such a way that packaging can be reused A drill with size 20 torx bit and or a Phillips 2 screwdriver is needed to open the crate Table 3 4 Unpacking Procedure Step Task 1 Loosen the screws and remove top ...

Страница 53: ... foam packing material and plastic bag from the ultrasound unit and monitor Caution Two people are needed in the next step due to the weight of the equipment 5 Disengage the brakes and slowly move unit down the ramp top panel Note Packing crate and material should be stored for future use Table 3 4 Unpacking Procedure Step Task screws screws probes and accessories packing WARNING WARNING Do not op...

Страница 54: ... correct voltage The Voltage settings for the Voluson 730Pro 730ProV Scanner is found on the identification plate on the rear of the system 3 4 3 EMI Protection This unit has been designed to minimize the effects of Electo Magnetic Interference EMI Many of the covers shields and screws are provided primarily to protect the system from image artifacts caused by this interference For this reason it ...

Страница 55: ...IDEO AUDIO AUDIO IN S VHS VIDEORECORDER POWER VGA1 VGA2 RGB S VHS1 S VHS2 VHS1 VHS2 VHS S VHS CKV AUDIO OUT AUDIO IN SOUND IN CPE CPN ECG RS232 CONSOLE REMOTE BW PRINTER AUX SUPPLY LAMP USB PS2 RS 232 Line In MIC Parallel Port UI Disp Line Out USB VGA LAN1 USB GEM USB PC SW PRINTER REMOTE VIDEO IN POWER COLOR PRINTER POWER USB ECG Primary Power Supply Foot Switch Power supply MODEM POWER RS232 TEL...

Страница 56: ... washers see Figure 3 6 2 When the cover is loose on top pull the rear cover out and move upwards 3 Disconnect the GND ground cable from the back of the rear cover plate 4 Connect all Peripherals according to connection scheme see Figure 3 5 5 Connect the GND ground cable at the rear cover plate 6 Mount rear cover plate reattach screw and washers and plug in the caps Figure 3 6 Remove rear cover p...

Страница 57: ...ANUAL Chapter 3 Installation 3 9 3 5 1 Monitor Connection NOTE The monitor comes already installed with the system 3 5 2 Foot Switch Connection NOTE To adjust the Foot Switch refer to the Voluson 730Pro 730ProV Basic User Manual Chapter 17 3 3 System Setup Peripherals Figure 3 8 Foot Switch Connection Scheme ...

Страница 58: ...STEMS KRETZTECHNIK RAFT VOLUSON 730PRO 730PROV BT 04 DIRECTION KTI105947 REVISION 2 SERVICE MANUAL 3 10 Section 3 5 Connection of Auxiliary Devices 3 5 3 ECG preamplifier Connection Figure 3 9 ECG Connection Scheme ...

Страница 59: ...ction Figure 3 10 Global Modem Connection Scheme WARNING WARNING Never install the telephone wiring or use the modem during an electric storm there may be a remote risk of electrical shock from lightning CAUTION Use only the power adapter supplied with the modem and connect it as shown Use of any other power adapter will void the warranty and could damage the modem ...

Страница 60: ...5 S VHS Video Recorder Connection 3 5 5 1 Mitsubishi HS MD3000 CAUTION ONLY the specific GE Kretztechnik Version of this video recorder type may be used in connection with the system Figure 3 11 Mitsubishi Video Recorder Connection Scheme NOTICE Please use the proper connection set and remote cable see Section 9 12 Optional Peripherals and Accessories on page 9 25 ...

Страница 61: ...he proper connection set and remote cable see Section 9 12 Optional Peripherals and Accessories on page 9 25 Set DIP Switches on back of the VCR as shown Switch 1 OFF Audio is muted during fast playback Switch 2 OFF FF REW mode is used during searches Switch 3 ON VCR counter is reset upon cassette ejection Switch 4 OFF The VCR can be controlled by the system Switch 5 6 ON The Baud rate is set to 1...

Страница 62: ...ection 3 5 Connection of Auxiliary Devices 3 5 6 B W Video Printer Connection Figure 3 14 B W Video Printer Connection Scheme Set DIP Switches on back of the Black White printer as shown in the image all DOWN NOTICE Please use the proper connection set see Section 9 12 Optional Peripherals and Accessories on page 9 25 ...

Страница 63: ...3 7 1 Installing Line Printer HP 990cxi or HP 995c on page 3 24 Figure 3 15 Line Printer Connection Scheme CAUTION Please observe that the Line Printer HP 990cxi or HP 995c has to be located outside of the patient environment acc IEC 60601 1 UL 2601 1 NOTICE The switch of the printer has to be in ON position before starting the system Leave printer switch always in the ON position NOTICE Please us...

Страница 64: ...o lateral distances See Instruction Manual of the printer CAUTION The Printer Supply Voltage must be the same as the Output Voltage of the Voluson 730Pro 730ProV Power Out Connectors Power Supply NOTICE The switch of the printer has to be in ON position before starting the system Leave printer switch always in the ON position NOTICE Please use the proper connection set see Section 9 12 Optional Pe...

Страница 65: ...atient environment acc IEC 60601 1 UL 2601 1 CAUTION The used printer may not be a medical device The Bluetooth Printer Set and the Power Supply of the Bluetooth Printer Adapter is also not a medical device The equipment meets the requirements of the EN60950 Standard NOTICE The switch of the printer has to be in ON position before starting the system Leave printer switch always in the ON position ...

Страница 66: ... of the device For this purpose the System Setup BACKUP page see Figure 4 22 on page 4 31 has a STOP USB DEVICES button NOTE Unplugging or ejecting devices without first stopping them can often cause to crash and lose of valuable data By clicking the STOP USB DEVICES button the Unplug or Eject Hardware dialog is started Using this dialog the USB devices can be stopped before they are physically di...

Страница 67: ...from the system By clicking OK the Unplug or Eject Hardware dialog is active again Close this dialog by clicking CLOSE Afterwards select OK to reboot the system Figure 3 20 Stop a Hardware device Figure 3 21 Safe to Remove Hardware WARNING WARNING Do not connect or disconnect any external USB devices to or from the system while scanning a patient The appearing dialogs could distract you from the s...

Страница 68: ...stem including the Back end Processor starts and the software code is distributed to initiate the scanner Boot up time is about 2 minutes NOTE The mains outlet of the system for peripheral auxiliary equipment are commonly switched with the Standby switch The switch of printers has to be in ON position before starting the system However be aware some auxiliary equipment may switch itself to standby...

Страница 69: ...xiliary equipment need not to be switched ON OFF separately 3 Disconnection of the Mains Power Cable is necessary For example Repairing the scanner Unscrew the 2 screws and remove the pull out protection to disconnect the cable from the system or unplug the cable from the wall socket 4 Press once on the brakes to block the front wheels brakes on front wheels under the foot rest 5 Prior to disconne...

Страница 70: ...hat the transducer twist lock lever is at horizontal position 2 Insert the transducer connector on the receptacle guide pin until it touches the receptacle mating surface 3 Twist the transducer twist lock lever to vertical position to lock it in place Twist the lever to the horizontal position to disconnect the transducer 4 Open the right hand side door lay the cable into the intended cable holder...

Страница 71: ...tion instructions see Section 3 7 1 Installing Line Printer HP 990cxi or HP 995c on page 3 24 Section 3 7 2 Installing Digital Color Printer Sony UP D21MD or UP D23MD on page 3 26 Section 3 7 3 Printer Installation manually on page 3 28 Section 3 7 4 Adjustment of Printer Settings on page 3 33 CAUTION The Bluetooth Printer Connection Set as well as the color printer HP 5600 Series or Olivetti Job ...

Страница 72: ...ct Click left right trackball key Opening a context menu upper trackball key 3 Turn ON the printer then switch ON the power of the system and wait till the system has booted NOTE The power switch of the printer has to be in ON position before starting the system 3 7 1 1 Install the HP 990cxi or HP 995c printer software driver Perform the following steps if this printer was never installed on the V...

Страница 73: ...r HPDeskjet990c for HP 990cxi C Utilities PrinterDriver HPDeskjet995c for HP 995c and then click OK 7 Confirm the correct path and click NEXT to install the driver All necessary files are copied 8 Confirm the installation by clicking FINISH to close the Hardware Wizard 9 Close all open windows and restart the system turn off and on the system Figure 3 27 Search for Network path on hard disk Figure...

Страница 74: ...Opening a context menu upper trackball key 3 Turn ON the printer then switch ON the power of the system and wait till the system has booted NOTE The power switch of the printer has to be in ON position before starting the system 3 7 2 1 Install the UP D21MD UP D23MD printer software driver Perform the following steps if this printer was never installed on the Voluson 730Pro 730ProV 4 Connect the U...

Страница 75: ...n click OK 7 Confirm the correct path and click NEXT to install the driver All necessary files are copied 8 The Message Digital Signature Not Found appears Click YES 9 Confirm the installation by clicking FINISH to close the Hardware Wizard 10 Close all open windows and restart the system turn off and on the system Figure 3 30 Search for Network path Figure 3 31 Confirm correct path and finish the...

Страница 76: ...ol panel press the UTILITIES key 2 In the Utilities menu on the left side of the screen select SYSTEM to invoke the setup desktop 3 Select the SERVICE page The password window appears automatically 4 Enter the password SHE and click the ACCEPT button to display the Service Tools window 5 Click on the PRINTER button Figure 3 32 System Setup Service page Figure 3 33 Service Tools window Printer butt...

Страница 77: ... d 6 Click the ADD PRINTER button A warning message appears Please read this message carefully and click YES if you have skills to do this Click the NEXT button to start the Add Printer Wizard 7 Select the Local Printer deselect Automatically install Plug and Play printer and then click NEXT Figure 3 34 Add Printer and Printer Installation Properties Figure 3 35 Add Printer Wizard ...

Страница 78: ...ter Installation 3 7 3 Printer Installation manually cont d 8 Select the corresponding Printer Port e g Figure 3 36 USB001 and click NEXT 9 In the following window select the HAVE DISK button 10 Use the BROWSE button to search the Printer Driver path C Utilities PrinterDriver xxxx Figure 3 36 Select Printer Port Figure 3 37 Have Disk ...

Страница 79: ... REVISION 2 SERVICE MANUAL Chapter 3 Installation 3 31 3 7 3 Printer Installation manually cont d 11 Click OPEN 12 Verify the selected Printer Driver path and confirm with OK Figure 3 38 Select Printer Driver path C Utilities PrinterDriver Figure 3 39 verify selected Printer Driver path ...

Страница 80: ... Do not share this printer and confirm the Printer Sharing window Figure 3 40 by clicking NEXT 16 The Complete the Add Printer Wizard window appears on the screen 17 Complete the manual Printer Installation with the FINISH button 18 Close all open windows close the System Setup with SAVE EXIT and restart the system turn off and on the system Figure 3 40 Assign name and select Printer Sharing no Fi...

Страница 81: ...er from the pull down menu and click the EDIT SETTINGS button 6 Confirm the warning message with the YES button The Printer Properties appear To adjust the HP 990cxi HP 995c printer see Section 3 7 4 1 HP 990cxi HP 995c Printer Settings To adjust the UP D21MD UP D23MD Printer see Section 3 7 4 2 UP D21MD UP D23MD Printer Settings To adjust the Bluetooth printer see Section 3 7 4 3 HP 5600 Series O...

Страница 82: ...EATURES page for adjusting Two Side Printing if desired Orientation must be set to Portrait See right Figure 3 43 5 For saving the adjusted printer settings click APPLY and then OK 6 Select the PORTS page and select verify the correct USB printer port 7 For saving the adjusted printer settings click APPLY and then OK Finally close the Printers window with the close button and exit System Setup wit...

Страница 83: ...eration see Section 3 7 4 Adjustment of Printer Settings 2 Select the PAPER page and select Paper Size UPC 21L large UPC 21S small Orientation Landscape recommended when using large paper size High Speed check mark on NOTICE Settings for Paper Size MUST match with the used Paper large small and also the right color ink cartridge has to be used Otherwise you will get an error message at printing Fi...

Страница 84: ...up menu select a Color Balance Cyan 0 Magenta 0 Yellow 0 b Gamma Select Gamma 1 c Color Correction set Printer Hardware Color Correction d Lightness Sharpness 7 or 8 Dark 0 Gamma 12 Light 8 4 For saving the adjusted printer settings click APPLY and then OK Finally close the Printers window with the close button and exit System Setup with SAVE EXIT Figure 3 45 Graphics page Color Balance Gamma Sele...

Страница 85: ...ze settings click APPLY and then OK 5 In the next window click OK again Finally close the Printers window with the close button and exit System Setup with SAVE EXIT 6 Assign the HP 5600 Series respectively the Olivetti Job Jet 210 printer as Report Printer see Section 3 7 5 Printer Remote Control Selection on page 3 38 7 Print report page s containing measurements For operation see Basic User Manu...

Страница 86: ...the setup desktop 3 Select the PERIPHERALS page Remote Print A Select the desired Printer for the remote control PRINT A key Remote Print B Select the desired Printer for the remote control PRINT B key NOTE Optionally the Remote Control can be done by Foot switches Therefore select Print A or Print B in Foot Switch Left or Foot Switch Right section Report Printer Select the desired Report Printer ...

Страница 87: ...l mains outlets are co switched by the unit s mains switch via built in isolation transformer Output voltage for ST1 ST5 150V or 230V Table 3 5 Physical Dimensions of Voluson 730Pro 730ProV Height Width Depth 142 cm 55 9 inches 68 cm 26 8 inches 100 cm 39 4 inches Table 3 6 Weight of Voluson 730Pro 730ProV with Monitor and without other Peripherals Model Weight kg Weight lbs Voluson 730Pro 730ProV...

Страница 88: ...Load User Settings Only Save Load Delete Full Backup Network to set up all DICOM Sonoview and Network configuration nodes System Info shows which Software Hardware version is installed in the system To invoke the Setup procedure In general operations are done with the trackball and the trackball keys mouse emulation 1 Press the UTILITIES key on the Control Panel The menu area changes to the Utilit...

Страница 89: ...ystem Setup see Figure 3 49 1 Select the text box to enter a new Clinic Name with the keyboard 2 Click SAVE EXIT to save Settings and exit System Setup The clinic name will be copied into the Hospital ID in the information header 3 8 3 3 How to change Language and or EUM Language Select the GENERAL page in the System Setup see Figure 3 49 1 Select the desired language from the pop up menu 2 Click ...

Страница 90: ... In the System Setup desktop select the SERVICE page enter the password SHE and click ACCEPT 3 If not currently selected click the desired Video Norm field PAL 50Hz or NTSC 60Hz NOTE The Adjustment of the Video Norm will take about 3 minutes 4 Click SAVE EXIT to exit System Setup The system will reboot BT Version at BT 04 systems Changing the Video Norm is not possible by the user Please contact y...

Страница 91: ...uipment inclusive color video monitor connected to these outlets must not exceed 350VA Table 3 8 Approved Peripherals Device Manufacturer Model Video Signal B W Video Printer SONY UP 895MD NTSC PAL Digital Color Video Printer SONY UP D21MD USB Port Digital Color Video Printer SONY UP D23MD USB Port Video Cassette Recorder SONY SVO 9500MD SVO 9500 MDP NTSC PAL Mitsubishi HS MD3000U HS MD3000E NTSC ...

Страница 92: ...able 3 9 External I O Connector Descriptions Item Connector Name Table Number Description 1 VGA OUT Table 3 10 print out VGA signal with monitor printer 2 MODEM Table 3 11 RJ 11 with global adapter kit for modem connection 3 NETWORK Table 3 12 DICOM input output twisted pair RJ 45 10 100 megabit s 4 USB 1 Table 3 13 USB 2 0 port 5 USB 2 Table 3 13 USB 2 0 port 6 COMP VIDEO OUT Table 3 14 BNC Conne...

Страница 93: ...HS H Sync 14 VGA OUT1 VS V Sync Table 3 11 Modem RJ 11 Modular 6 Pin Pin No Output Signal Description 2 TEL L4 Telephone L4 3 TEL L2 Telephone L2 4 TEL L1 Telephone L1 5 TEL L3 Telephone L3 Others NC Non connection Table 3 12 Network Connector RJ45 Modular 8 Pin Pin No Output Signal Description 1 ETHER TD Ethernet RD 2 ETHER TD Ethernet RD 3 ETHER RD Ethernet TD 6 ETHER RD Ethernet TD Others NC No...

Страница 94: ...Connector located at Power Supply Module CPN rear side Pin No Input Signal Description 1 Signal GND 2 right switch normally open 3 left switch normally open 4 not connected not connected Table 3 16 Video Specifications VGA Connector Timing Parameter Value Visible Resolution 800 x 600 Horizontal Rate kHz 47 20 H Total cycle time µs 21 20 H Sync width µs 1 50 H Back Porch µs 2 94 H Active Video Time...

Страница 95: ...rizontal Total Line 64µs 1135 Pixel 63 56µs 910 Pixel Horizontal Active Display 50 50µs 902 Pixel 52 50µs 752 Pixel Horizontal Front Porch 1 96µs 35 Pixel 1 62µs 23 Pixel Horizontal Sync Width 4 62µs 82 Pixel 4 68µs 67 Pixel Horizontal Back Porch 6 52µs 116 Pixel 4 76µs 68 Pixel Vertical Total Lines 20ms 312 50 Lines 16 68ms 262 50 Lines Vertical Active Lines 18 18ms 284 Lines 15 22ms 239 50 Lines...

Страница 96: ...Voluson 730Pro 730ProV Basic User Manual Chapter 17 3 2 System Setup User Settings For information on configuring assignment of PRINT keys Foot Switch Save Destination VCR Source etc refer to the Voluson 730Pro 730ProV Basic User Manual Chapter 17 3 3 System Setup Peripherals For information on configuring Software Options refer to the Voluson 730Pro 730ProV Basic User Manual Chapter 17 3 4 System...

Страница 97: ...Worksheet Site System Information Services Destination Devices TCP IP Settings Device Type Manufacturer Name IP Address Port AE Title 1 2 3 4 5 6 7 8 9 10 11 12 Name AE Title Dept Serial Type Floor Room REV Comments CONTACT INFORMATION Name Title Phone E Mail Address Site IP Settings IP Address Subnet Mask Default Gateway Remote Archive Setup Remote Archive IP Remote Archive Name ...

Страница 98: ... display the Utilities menu 2 Select SYSTEM and then open the NETWORK page on the System Setup desktop screen 3 Click the NETWORK CONFIGURATION button read the message and confirm with YES 4 The Internet Protocol TCP IP Properties dialog page appears To specify a DICOM Address follow the instructions of described in the Basic User Manual Chapter 17 3 7 of the Voluson 730Pro 730ProV Figure 3 51 Net...

Страница 99: ...nnection must be re established manually after a shutdown or reboot 4 Select the CONNECT button to establish the connection to the remote machine If successful the DISCONNECT button becomes active Figure 3 53 Map Network Drive window NOTICE If there is an error during the connection a warning message appears inside the dialog In this case please verify the data in the dialog NOTICE If there alread...

Страница 100: ... that all relevant safety and user information is available during the operation and service of the complete system 3 13 1 Product Locator Installation NOTE The Product Locator Installation Card shown may not be same as the provided Product Locator card 3 13 2 User Manual s Check that the correct User Manual s for the system and software revision is included with the installation Specific language...

Страница 101: ...W or MO Disk At least one transducer See Probes on page 9 29 for an overview normally you should check all the transducers used on the system Table 4 1 Contents in Chapter 4 Section Description Page Number 4 1 Overview 4 1 4 2 Required Equipment 4 1 4 3 General Procedure 4 2 4 4 Functional Checks 4 7 4 5 Backup and Restore Database Preset Configurations and Images 4 31 4 6 Software Configuration C...

Страница 102: ...Switch ON the Circuit Breaker at the rear of the system CAUTION SYSTEM REQUIRES ALL COVERS Operate this unit only when all board covers and frame panels are securely in place The covers are required for safe operation good system performance and cooling purposes NOTICE Lockout Tagout Requirements For USA only Follow OSHA Lockout Tagout requirements by ensuring you are in total control of the Power...

Страница 103: ... Standby switch So the auxiliary equipment need not to be switched ON OFF separately 4 3 2 Power Off Shutdown NOTE After turning off a system wait at least ten seconds before turning it on again The system may not be able to boot if power is recycled too quickly 4 3 2 1 Scanner Shutdown 1 Press the ON OFF Standby switch left below the Control Panel 2 Switch OFF the Circuit Breaker at the rear of t...

Страница 104: ...monic Imaging key 7 Inter memory keys to save to Sonoview or send to DICOM server 8 Print A Print B Trigger key 9 Freeze Run key 10 Trackball and Trackball keys 11 Trackball Menu Navigation key 12 Annotation and Measurement keys 13 Menu Control digipots and toggle switch controls 14 Exit key 15 Navigation wheel Menu control element 16 Probe key 17 Patient Data Entry key 18 Utilities key 19 Keyboar...

Страница 105: ...ols By rotating resp switching they deliver digital pulses and can be selected by program call up They are displayed in the status area by their location their function and their actual value of setting NAVIGATION WHEEL and corresponding KEYBOARD SHORTCUT FUNCTIONS The available menus are shown in the menu area on the left side of the screen The current menu selection is highlighted Roll the navig...

Страница 106: ...urement results 3 Patient ID number GA Gestational Age 16 Status volume box saved images gray map 4 Probe Application 17 Status area of the Freeze key 5 Depth Frame rate 18 Status area of the trackball 6 Mechanical Index 19 Status area of digipots and flip switch controls 7 Thermal Index 20 Menu Softkey area 8 Sonographers Name 21 Bodymarks 9 Hospital Name Identification 22 Depth scale markers 10 ...

Страница 107: ...aging A Plane C Plane and VCI Static at systems upgraded to BT 05 only STIC at systems upgraded to BT 05 only TUI at systems upgraded to BT 05 only VOCAL II NOTE Some software may be considered standard depending upon system configuration If any Modes or Options are not part of the system configuration the check can be omitted NOTE Different menus are displayed depending on which Menu and which Mo...

Страница 108: ...from Single to DUAL or QUAD display mode Press the SINGLE format key or the 2D MODE key to change from Dual or Quad to Single display 5 Image Orientation Use the LEFT RIGHT respectively the UP DOWN keys on the control panel to alternate the image orientation The orientation marker on the screen shows the actual orientation 6 2D Automatic Optimization Pressing the AUTO key causes automatic optimiza...

Страница 109: ... High Resolution Zoom function 15 Harmonic Imaging Press the HI key on the control panel to switch on off the Coded Harmonic Imaging function in 2D Mode provided the active probe allows this function 16 FREQUENCY resp HARM FREQU in case of Harm Imaging To adjust the range of the receive frequency high resolution lower penetration mid resolution mid penetration or lower resolution high penetration ...

Страница 110: ...des of gray thereby increasing the adjustable range of contrast 27 PERSIST Persistence is a temporal filter that averages frames together This has the effect of presenting a smoother softer image This function is only available if CRI is switched off 28 CRI ENHANCE If this filter is set to high the XBeam CRI image is smoothed CRI Enhance setting off leads to a sharper impression of the XBeam CRI i...

Страница 111: ...htness of the entire M image 4 M Mode Depth Common with 2D Mode Depth 5 Invert This function inverts the M mode trace from up to down in the M mode display area The Invert function is only available with endovaginal probes 6 SPEED By touching up or down four different sweep speeds can be selected 7 FREQUENCY Common with 2D Mode Frequency resp Harm Frequ in case of Harm Imaging 8 FORMAT For selecti...

Страница 112: ...4 Functional Checks 4 4 2 1 MCFM Mode Check For further details refer to the Voluson 730Pro 730ProV Basic User Manual Chapter 6 4 By pressing the M control and the C control the MCFM mode is switched on in the preparation mode The M cursor with M Color window appears on the active 2D image Figure 4 9 MCFM Main and MCFM Sub Menu ...

Страница 113: ...te position to Gate size 2 Activation of Doppler Mode Press the right trackball key to activate the motion display Press the left trackball key to activate both Modes B D 3 Doppler Gain Rotate the PW MODE key to adjust the amplification of the entire spectrum 4 Steering The steering function is only possible with linear probes 5 RT TRACE Real Time Auto Trace The envelope curve of the Doppler spect...

Страница 114: ...switched on 11 ANGLE The angle cursor can be turned in both directions without stop By pressing the angle knob repeatedly the angle correction switches from 60 to 0 and to 60 12 SCALE To select the displayed measuring unit in relation to the zero line 13 FORMAT For selection of either one of three formats 14 FREQUENCY Serves for selection of the required transmit frequency for actual gate position...

Страница 115: ...djust the Power Doppler Gain Rotate the C MODE key to adjust the Tissue Doppler Gain 3 2D 2D C PD or TD Changes the Single image display to two simultaneous half images The left frame shows only the 2D Mode image The right frame shows the 2D Mode image with color information 4 Steering Beam Steering is only possible with linear probes in CFM and PD Mode 5 Invert The color of the color wedge invert...

Страница 116: ...the number of pulses to constitute one Color or Power Doppler line in the display 18 LINE D Determines the line density within the Color Box The lower the line density the larger the line distance and the size of the color pixels 19 C MAP PD TD Provides selectability of the color coding for an optimization of the display of blood flow similar to the post processing curves with grayscale 2D scans A...

Страница 117: ...C and TUI Options are only available at systems upgraded to BT 05 SW 5 x x Figure 4 12 Pre Volume Mode menus Table 4 7 Pre Volume Mode Functions Step Task Expected Results 1 3D RENDERING 3D Static volume acquisition rendered 3D image also in combination with PD or CFM Mode 1 SECT PLANES 3D Static volume acquisition resp 4D volume acquisition without rendered 3D image Volume Acquisition Static 3D 4...

Страница 118: ...l calculation of 3D rendered image 7 Quarter size display of Sectional Planes without 3D image or Quarter size display of Sectional Planes rendered 3D image Note The display depends on selected Acquisition and Visualization Mode 8 Dual size display of Sectional Planes rendered 3D image Note The display depends on selected Acquisition and Visualization Mode This format is not possible for Static 3D...

Страница 119: ...y repeatedly to choose the Reference image among A B or C Choosing a reference image automatically determines the control functions of the rotary controls and the trackball for the liberal adjustment of a sectional plane 6 Initial Condition Resets the rotations and translations of a volume section to the initial start position 7 3D Image Orientation To change the image orientation of the rendered ...

Страница 120: ...RI can be activated in all 3D 4D modes except Full size display to reduce speckle which interferes with the sectional planes A B and C Note The rendered 3D image is not affect 6 BACKGR Adjusts the contrast of the screen background from dark to bright Note This key is not available in Static 3D Sectional Planes mode 7 CONTR POS To set the start point of the gamma curve bending Graphic display on sc...

Страница 121: ...6 To start stop the Cine Loop playback press the RIGHT OR LEFT TRACKBALL KEY After stopping the sequence move the TRACKBALL to display the images one by one 4 4 6 4 Spectral Doppler or M Cine Loop Press FREEZE then roll the TRACKBALL to display the Cine Loop one by one The UPPER TRACKBALL KEY changes from the 2D Cine to the D Loop respectively M Loop NOTE The active Cine is displayed on the screen...

Страница 122: ...e DELETE MEAS key on the keyboard NOTE To alternate the control from one cursor to the other press CHANGE the upper trackball key To re adjust a traced line press UNDO the upper trackball key repeatedly General remarks to perform Generic Measurements By pressing the CALIPER key on the control panel the Generic Measurement function is switched on Selection of the desired measurement marks is done w...

Страница 123: ...LETE or press the BACKSPACE key on the keyboard The original caliper is displayed to restart the measurement To exit the measurement function press the CALIPER key or EXIT key on the control panel 4 4 7 3 Volume Measurements 1 Press the CALIPER key and then select the GENERIC VOLUME in the menu area 2 Select the appropriate item 3 Perform the measurement s using the TRACKBALL and SET right or left...

Страница 124: ...termination of the envelope curve requires a clear and low noise record of the Doppler spectrum Otherwise the reliability of the displayed measurement results may not be ensured 4 4 7 4 2 Manual Trace 1 Press the CALIPER key 2 Select the MANUAL TRACE item from the menu area A cursor appears on the screen 3 Move the cursor with the TRACKBALL to the start point of the measurement and press SET right...

Страница 125: ...MENU NAVIGATION key is illuminated the trackball is consequently assigned to select the menu items in the menu area on the left side of the screen Positioning of measurement marks is done with the TRACKBALL Entering and storage of measuring marks is done with SET right or left trackball key To change measuring marks before completion press CHANGE upper trackball key Depending on the setting in the...

Страница 126: ...ar 4 Clicking a Setting field causes loading of the preset The probe is initialized the main menu 2D mode and the ultrasound image appears on the monitor in write mode real time display 4 4 9 3 Deactivating the probe When deactivating the probe the probe is automatically placed in standby mode read mode 1 Press the FREEZE key 2 Gently wipe the excess gel from the face of the probe Refer to the Bas...

Страница 127: ...l topics Clipboard Sending Exams Printing Exams Images Export Exams Images Backup Exams Restore the Backup Exams DICOM Print Send Verifying and Pinging a Device Sending Images via e mail Browsing and Managing an Exam s stored Image Connectivity and Dataflow Concept and Creation Configuring Connectivity Services Destinations Input of comments and voice annotations Measure Distance and Ellipse Butto...

Страница 128: ... from the MOD Manual ejection methods are listed below in preferred order from best 1 to worst 3 a Press the EJECT switch on the MOD while system is ON b Press and hold the EJECT switch while the system is booting c Mechanical ejection Insert the end of a paper clip into the hole next to the EJECT switch while system power is OFF NOTICE The MOD Magneto Optical Drive is an Option Figure 4 16 Magnet...

Страница 129: ... mark the Low Level Format icon 2 Click the START button to start the formatting process 3 A message box appears on the screen Confirm with OK 4 When the formatting has been completed click OK to continue 5 Select the CLOSE button and press the EXIT key on the control panel to return to the Scan Mode To erase format the backup media DVD CD R W or MOD press the SONOVIEW key on the Control panel The...

Страница 130: ...llowing message 3 Press the EXIT key on the control panel to return to the Scan Mode 4 4 11 1 1 Adjusting the Write Speed of the DVD CD Recorder 1 Click the SETTINGS icon on the left side of the Sonoview screen 2 Select DVD CD RECORDER from the tool bar 3 Choose the Write Speed from the corresponding pop up menu NOTICE It is highly recommended to use the complete erase mode to avoid problems with ...

Страница 131: ...ts Configurations Application Settings on page 4 35 Section 4 5 4 Load Full Backup Presets Configurations Application Settings on page 4 37 Section 4 5 5 Delete Full Backup Presets Configurations Application Settings on page 4 39 The User Settings and or Full Backup can be saved to the following destinations D partition of internal hard disk DVD CD R W MOD if present Mapped Network Drive Z see Sec...

Страница 132: ... into the drive 2 Press the UTILITIES key on the control panel The menu area changes to the Utilities menu 3 Select the SYSTEM item from the menu area to activate the setup desktop screen 4 Select the BACKUP page 5 Click the SAVE button of the User Settings Only group 6 Choose the media e g DVD CD RW and click the SAVE button 7 Select the NEW FILE key and enter a file name without extension 8 Clic...

Страница 133: ...P page and click the LOAD button of the User Settings Only group 5 Choose the media e g DVD CD RW and click the LOAD key 6 Select the appropriate file and click OK The Load Backup Data window appears CAUTION The loading procedure overwrites the existing application settings on the local hard drive Make sure to insert the correct DVD CD or MO Additionally you can load the backup from D User Setting...

Страница 134: ...o start the loading procedure of the complete backup into the system NOTE Also only parts of a User Settings Only Backup can be loaded into the database to overwrite restore copy etc the database in the system 1 Click the sign to open the content tree 2 Click the button to copy the selected item into the Load Data field 3 To return selected items from the Load Data field to Backup Data field selec...

Страница 135: ...a NOTE Always Full Backup any presets configurations and application settings to HDD and or DVD or MO disk before upgrading the software and or application settings This ensures that if the presets need to be reloaded will be the same ones the customer was using prior to service 1 Press the UTILITIES key on the control panel The menu area changes to the Utilities menu 2 Select the SYSTEM item from...

Страница 136: ...ion 3 5 10 External USB Devices on page 3 18 Figure 4 28 directory structure of full backup data Figure 4 29 Other drive drop down menu NOTICE When the backup is saved to an external USB device the system has to be informed about the removal of the hardware For this purpose every last dialog of Full Backup Save and Full Backup Delete has a STOP USB DEVICES button see Figure 4 30 Figure 4 30 Please...

Страница 137: ...ion than the system the backup was created on is prohibited 2 Options can only be restored on the same Voluson 730Pro 730ProV system within the same major software version 3 When loading a backup into a system with a software version that has a higher major number 2 x x 3 x x 4 x x 5 x x the following items will not be restored A User Settings B Options C State of the Service Platform new model ty...

Страница 138: ...u area to activate the setup desktop screen 3 Select the BACKUP page and click the LOAD button of the Full Backup group 4 Choose the Destination e g Network Drive 5 Click on the backup to be restored additional information is displayed in the table 6 Select the NEXT button The following window will be displayed 7 Select the data to be restored to the Voluson 730Pro 730ProV system NOTE For descript...

Страница 139: ...ey on the control panel The menu area changes to the Utilities menu 2 Select the SYSTEM item from the menu area to activate the setup desktop screen 3 Select the BACKUP page and click the DELETE button of the Full Backup group 4 Choose the Destination e g HDD Hard disk 5 Click on the backup to be deleted additional information is displayed in the table 6 Select the DELETE button Figure 4 33 Start ...

Страница 140: ...ackball key SET The number of all exams the number of currently selected exams the number of images and the capacity of selected images are displayed automatically at the right and upper corner of the exams list 5 Click the BACKUP button 6 Select the destination for the backup NOTE For destination NET perform Section 3 12 1 Map Network Drive on page 3 51 first 7 After finishing the backup select w...

Страница 141: ... switch Foot Switch left and right are assigned as desired by the customer 8 Peripherals Check Save Destination assignment assigned as desired by the customer 9 Peripherals Check 2D Save key assignment assigned as desired by the customer 10 Peripherals Check Remote Print A Remote Print B Report Printer and DICOM Print Job assignment settings assigned as desired by the customer 11 Option Check that...

Страница 142: ...ss the VCR key on the Control Panel once The VCR Remote Control menu is displayed in the menu area as well as the status area on the screen 6 Press the RECORD digipot or select the RECORD item from the menu area to start recording A red dot is displayed in the VCR status area on the Title bar to indicate that recording has begun 7 Press the STOP digipot or select the STOP item from the menu area T...

Страница 143: ...trol Console forward grasp it at the front grip and rotate the console It is possible to rotate the Control Console up to 30 to the right WARNING WARNING Do not put your hand between the control console and the main unit when moving the console to the 0 position Danger of injuries Figure 4 38 Front Wheel with Brake Table 4 15 Brakes and Direction Lock Step Task Expected Result s 1 Flap the foot re...

Страница 144: ...The password window appears automatically 4 Enter the password SHE and click the ACCEPT button to display the Service Tools window 5 Click the MAINTENANCE REPORT button The following message box will be displayed 6 Fill in the requested information and click OK 7 Click the EXIT button on the Service Tools window and the EXIT button on the System Setup Service page NOTE After Hardware or Software m...

Страница 145: ... KRETZTECHNIK DRAFT VOLUSON 730PRO 730PROV BT 04 DIRECTION KTI105947 REVISION 2 SERVICE MANUAL Chapter 4 Functional Checks 4 45 4 9 2 Site Log Paper Documentation Table 4 16 Site Log Date Service person Problem Comments ...

Страница 146: ...GE MEDICAL SYSTEMS KRETZTECHNIK RAFT VOLUSON 730PRO 730PROV BT 04 DIRECTION KTI105947 REVISION 2 SERVICE MANUAL 4 46 Section 4 9 Site Log This page was intentionally left blank ...

Страница 147: ...t and subsystem function It also describes the Power Distribution System PDS and probes Table 5 1 Contents in Chapter 5 Section Description Page Number 5 1 Overview 5 1 5 2 General Information 5 2 5 3 Main board Chassis GEF Module 5 20 5 4 FrontEnd Processor 5 22 5 5 BackEnd Processor 5 27 5 6 Internal I O 5 31 5 7 Top Console 5 34 5 8 Monitor 5 39 5 9 External I O 5 40 5 10 Peripherals 5 41 5 11 ...

Страница 148: ... for 2D Mode and Color Doppler Imaging CFM PD and TD M Mode MCFM Imaging Doppler PW CW 3D Mode and Real Time 4D Imaging Different combinations of the above modes Figure 5 1 Voluson 730Pro 730ProV Major Components Monitor Gel Holder TGC Slider Controls Monitor Controls Control Console Front Grip Lever to rotate and lock the Control Console mounted at the front below the Console Probe Connectors Foo...

Страница 149: ...5 27 GEU Top Console User interface System I O with hard keys Section 5 7 on page 5 34 MONITOR Section 5 8 on page 5 39 GES External I O Connection Module Section 5 8 on page 5 39 GEM Removable Disk drive module MO Drive and ECG preamplifier optional Section 5 10 on page 5 41 CPN Primary Power supply and Isolation transformer for the peripherals Section 5 11 on page 5 42 GW System mechanical chass...

Страница 150: ...onal CRS RFI Board MID Proc SysCtrl MotCtrl PW Doppler CPM PC Motherboard Subboards Slot CPU several PC Cards CKV Video Card DMA Controller Card CPE Backplane Card CPP Powersupply Secondaryboard GEU User Interface Module Monitor GES External IO GEF Main board Chassis CPN MainPowerSupply Module Subboards 32xCPD GEM ECG Diskdrive Module GW Trolley housing mechanical framework The following component...

Страница 151: ...ety of linear curved phased array and pencil probes for various clinical applications Any three probes may be connected at the same time one pencil CW probe Medical application fields include Obstetrics Gynecology and Fertility Radiology Internal Medicine Neurology Cardiology Oncology Urology Orthopedics Pediatrics The system is designed for follow up expansion In addition to the initial operation...

Страница 152: ...t resolution excellent acoustic clutter rejection and an easy to operate user interface Coded Harmonics enhances near field resolution for improved small parts imaging as well as far field penetration It diminishes low frequency amplitude noise and improves imaging technically difficult patients It may be especially beneficial when imaging isoechoic lesions in shallow depth anatomy in the breast l...

Страница 153: ...ucture 2D echo For each pixel in the overlay the decision of whether to display color Doppler power gray scale echo information or a blended combination is based on the relative strength of return echoes from the soft tissue structures and from the red blood cells A high pass filter wall filter is used to remove the signals from stationary or slowly moving structures Tissue motion is discriminated...

Страница 154: ...3 dimensional structure in the body To complete survey a 3 dimensional structure in the body it is necessary to collect 2D images which span a volume containing the structure One way is to sweep the imaging cross section by translating it in a direction perpendicular to the cross section Another example method is to rotate the cross section about a line contained in the cross section The Voluson 7...

Страница 155: ...ro 730ProV Figure 5 5 Top Level Architecture PCI to Local I F DMA IPC Pentium 4 3 06 GHz 768 MB PCI 32 Bit 33 MHz CRT Monitor 15 Video Mgmt Dig VGA Out USB 2 0 HDD 80 GB DVD RW MO Drv LAN 10 100 MBit AGP VGA RS232 Aux VGA User I F PS Unit Probe MUX and Connectors Board CPU HVPS Motor Power Drive LVPS Front End Board CKV Board CRS RFI Beamformer Board CPR System Control Control Motor Control ...

Страница 156: ...nitude Calculator Logarithmic Amplifier Resample Edge Enhance Line Filter Blending Frame Filter CKV Video Card DMA Controller f HighSpeed DataTransfer into Slot CPU Mem Video Converter TMDS VGA RGB VHS etc VGA1 VGA2 S VHS2 VHS1 O VHS IN VHS2 O S VHS IN RGB S VHS1 PC Keyboard Trackball VGA Signal to V730EXPERT Touchscreen Signal is in fact analog but is re sampled and digitalized in Touchscreen Cab...

Страница 157: ...ques on page 5 12 FrameFilter Multibeam DeInterleave means Incoming Pixelorder shot1pix1 shot2pix1 shot3pix1 shot4pix1 shot1pix2 shot2pix2 shot3pix2 shot4pix2 is converted to the new order shot1pix1 shot1pix2 shot1pix3 shot2pix1 shot2pix2 shot2pix3 After DC cancelling the signal is mixed with RX Frequency and brought to LF Spectrum where the LowPassFilter cuts HF Mixer and Magnitude Calculator arr...

Страница 158: ...s are shot to the same Tissue location The Beams have different TX foci By means of Blending adaption of Brightnesses they are composed to one whole RX Line c XBeam CRI CrossBeam Compound Resolution Imaging Does not need any special functions of CRS Image is composed of more than one different direction steered images PC calculated 5 2 4 2 M Mode 1 CRS see 5 2 4 1 B Mode 2 CPR see 5 2 4 1 B Mode 3...

Страница 159: ...with ECG CineMode Memory for the ECG Curve is inside PC Memory Software has to take care that D Mode Image and ECG Curve are placed exactly one upon the other means have the same Cine Shift 4 CKV Video section see 5 2 4 1 B Mode 5 2 4 4 D Mode Autotrace draws PC calculated envelope to D Spectrum ECG Curve is similar to Autotrace Curve 1 CRS PRF generator see 5 2 4 1 B Mode After DC cancelling the ...

Страница 160: ...de Freezes after 1 volume sweep see 5 2 4 1 B Mode 5 2 4 7 Real Time 4D Mode nonstop volume rendering see 5 2 4 1 B Mode 5 2 4 8 XBeam CRI Mode CrossBeam Compound Resolution Imaging see 5 2 4 1 B Mode 5 2 4 9 Extern Video Mode display Video from Video Recorder 1 CRS Not used for Signal Processing 2 CPR Not used for Signal Processing 3 CKV DMA section Not used for Signal Processing 4 CKV Video sect...

Страница 161: ...sition box is at the same time the render box All information in the volume box is used for the render process In Real Time 4D mode a frame rate of up 16 volumes second at Voluson 730Pro V up to 10 volumes second is possible By freezing the acquired volumes size can be adjusted manipulated manually as known from the Voluson 3D Mode BT Version Some Software Options may be considered standard or are...

Страница 162: ...y the examiner in multi planar display Construction of a virtual shell around the contour of the lesion The wall thickness of the shell can be defined The shell can be imagined as a layer of tissue around the lesion where the tumor vascularization takes place Automatic calculation of the vascularization within the shell by 3D color histogram by comparing the number of color voxels to the number of...

Страница 163: ...me Contrast Imaging Volume Contrast Imaging utilizes 4D transducers to automatically scan multiple adjacent slices and delivers a real time display of the ROI This image results from a special rendering mode consisting of texture and transparency information VCI improves the contrast resolution and therefore facilitates finding of diffuse lesions in organs VCI has more information from multiple sl...

Страница 164: ... can be upgraded by implementing the CW Dopplerboard CRW 5 2 6 2 ECG Preamplifier MAN6 internal digital version For details see Section 5 10 1 1 ECG preamplifier MAN6 optional on page 5 41 5 2 6 3 MOD Magneto Optical Drive For details see Section 5 10 1 3 Magneto Optical Drive optional on page 5 41 5 2 6 4 Scan Freeze Foot switch 5 2 6 5 Global Modem The Multi Tech global modem is a standard modem...

Страница 165: ...During Power On Briefly Flashing These LEDs indicates communication speeds above 14 kbs If one of this LEDs are ON during communication it will stay ON until the modem is reset or connected the next time At speeds below 14 kbs these LEDs are OFF Red 33 V 34 Mode Red 14 V 32bis Mode Red OH Off hook ON when dialing online or answering a call Flashes if puls dialing Off when modem not in use Red TR T...

Страница 166: ...s GEF Module The GEF Module contains the Front End processor and the Back End processor and the Secondary Power supply for the full GEF Chassis Additionally GEF Module is the connection point of the internal I O wiring Figure 5 10 Mechanic of GEF Module Front End Secondary Power Board CPP Back End Front Rear with Internal I O Audio Video View from Right with Internal I O PC part Top View ...

Страница 167: ...N TxDelayMatrix Transmitter RxAmp RxDelayMatrix Analog Mixer Amps Filter WMF AntiAliasing PRF SampADC16Bit SIN COS Powered for B C D modes Multi Beam DeInterleave Subtraction Filter for THI Digital TGC DC Cancelling Mixer with NCO FIR Low Pass B Mode Magnitude Calculator Logarithmic Amplifier Resample Edge Enhance Line Filter Blending Frame Filter CW Data serial D Q 16Bit each US Data SIN COS Syst...

Страница 168: ...ansmitter channel used 128 Receiver channels CPR contains 32 pieces of CPD see Section 5 4 1 3 CPD Beamformer Sub board on page 5 23 Figure 5 12 CPU Beamformer CPR Power 192 TX RX MOT_signals CTRL signals CPU Probe Board Power 192 TX RX MOT_signals CTRL signals Power 192 TX RX MOT_signals CTRL signals TX RX Probe 1 Probe 2 Probe 3 CW Probe CPK Motherboard PowerSupply SIN COS Mot Strobes CTRL signa...

Страница 169: ...13 Block diagram CPD Sep 8 2 Pushpull Output stage ASIC BFIC 2 1 Mux Antialiasing Filter TGC ADC 10 Bit Antialiasing Filter Antialiasing Filter 8 Sep Sep Sep 8 El_0 El_128 El_3 El_131 Tx_0p Tx_0n Rx_0 Rx_1 Rx_2 Rx_3 2 1 Mux 2 1 Mux 2 1 Mux Tx_0 Tx_1 Tx_6 Tx_7 El_1 El_129 El_2 El_130 to ASIC n 1 From ASIC n 1 Driver Driver EN_A EN_B Driver Driver EN_A EN_B Tx_1p Tx_1n Driver Driver EN_A EN_B Driver...

Страница 170: ...fier Filter WMF Anti Aliasing PRF SampADC16Bit 5 4 1 5 CPZ Cover Board Transfers the analogue transmitting receiving signals between the following boards CPU CPR CRW 192 Transmitter 128 Receiver channels 16 CW Doppler channels switched by relays at CPZ 5 4 1 6 CPK Motherboard of GEF Module Following boards are direct connected to the CPK CPU Probe connector Board CPR Beam former Board CRW CW Doppl...

Страница 171: ...etition Frequency Generator Beamformer Configuration Tx transmitting Frequency TxRx transmitting receiving Focus timing TxRx transmitting receiving Apodization Line number lateral Position contains TX Power DAC C Doppler CFM 1 CFM processing Color Doppler shot lines transfer to CKV 2 Doppler Processing DSP Audio DAC 3 Beamformer Configuration Data EH_DATA 16 EH_ADR 8 B Mode Data D Motor Control Ma...

Страница 172: ...NIK RAFT VOLUSON 730PRO 730PROV BT 04 DIRECTION KTI105947 REVISION 2 SERVICE MANUAL 5 26 Section 5 4 FrontEnd Processor FAN 10 24V 15 Watt adjustable by software TX_POW 90V TX_POW 90V TX_POW2 TX_POW2 Figure 5 14 Block diagram CPP ...

Страница 173: ...S VHS2 VHS1 O VHS IN VHS2 O S VHS IN RGB S VHS1 PC Keyboard Trackball VGA Signal to V730EXPERT Touchscreen Signal is in fact analog but is re sampled and digitalized in Touchscreen Cable does not exist at V730PRO GEU User Interface Module PowerSupply Cable 5V 12V USB Cable GEM ECG Diskdrive Module External Diskdrives ECG Module CPN MainPowerSupply Module CPY Powerswitch Plugboard CCF Powersupply P...

Страница 174: ...I F 1 us Write slot 1 1 us Write slot 0 DMA 0 Demand Mode 2 us Read slot UART 16C550 Ext Bus I F DMA 1 Block Mode Local Bus 32 Bit 33MHz VGA Retrace Interrupt B MB C IQ MC IQ PW CW Read Write Address generator DMA 0 I F B C DMA 1 I F MB MC PW CW Host PC Access PLX PCI 9054 Fifo Buffer Memory SDRAM 32 Bit 66 MHz VCR Remote To From U S Front end H W 32 Bit 16 7 MHz Video I O FPGA RAM DVI TMDS I F CV...

Страница 175: ...Card DMA Controller f High Speed Data Transfer into Slot CPU Memory VGA Output 2 Channels for the System Main Monitor and external device Video Converter VGA RGB to VHS S VHS RGB Display External Playback Video and adding Overlay graphics to it see also Figure 5 16 Block diagram CKV on page 5 28 5 5 2 4 Hard Disk Drive Minimum 80GB IDE Stores the system programs and Image filing patient data Repor...

Страница 176: ... 7 CPP Power Supply Secondary Board Motor Power stage used for Supply of both FrontEnd and BackEnd DC DC Converter 59VDC to following output voltages 3 3V 5V 12V 15V FAN 10 24V 15 Watt adjustable by software TX_POW 90V TX_POW 90V Motor Sinus2 Powerstage Amplifier Motor Cosinus2 Powerstage Amplifier TX_POW2 TX_POW2 see also Section 5 4 1 8 on page 5 25 ...

Страница 177: ...cable VGA1 VGA2 RGB S VHS1 S VHS2 VHS1 VHS2 VHS S VHS AUDIO OUT AUDIO IN SOUND IN R L CKV CPE Monitor CPN PC USB PS2 ECG RS232 CONSOLE REMOTE BW PRINTER LAN1 RS 232 Line In MIC AUX SUPPLY Parallel Port UI Disp Line Out USB VGA GES External IO Board LAMP CPN Primary Power Supply FootSwitch StdbySwitch User PowerON OFF GEU User Interface Module PS2 Adaptercable USB GEM USB MODEM GEM Disk Drive Modul...

Страница 178: ...m PAL 1Vss 75 Ohm NTSC 7 VHS 2 Video 2 OUT 1Vss 75 Ohm CCIR 1Vss 75 Ohm FCC 8 VHS Video IN 1Vss 75 Ohm PAL CCIR 1Vss 75 Ohm NTSC FCC 9 S VHS S VHS IN 10 Audio IN Sound Audio IN R L Sound 11 Remote BW Printer Remote control for BW Printer 12 AUX Supply Power Supply for Module GEM 13 Console Connector for Console 14 CPN Connector for Power Supply CPN input 15 Lamp Connector for external lamp 16 RS23...

Страница 179: ...r Name Description 1 USB GEM Connector for Disk Drive Module GEM 2 Parallel Port Parallel port for PC Line Printer 3 Audio MIC Connector for Microphone Line OUT Connector for Soundcard 4 USB USB port connectors 5 LAN1 Connector for Network twisted pair RJ 45 10 100 megabit s 6 UI Disp NOT USED in the Voluson 730Pro 730ProV 7 USB USB port connector 8 RS 232 Connector for Global Modem 9 VGA no funct...

Страница 180: ...ontroller Slide pots TGC with zero raster position Rotary Encoders with integrated push buttons PS 2 compatible Trackball 2 with standard PC interface PS 2 compatible Qwerty Keyboard with standard PC interface LED indicators with 3 intensity levels off 50 100 2 Speaker used for Doppler and voice replay DC DC Converter Converts 12V DC input voltage to 5V DC output voltage for supplying UI component...

Страница 181: ...alc 3D 4D HI PD CW auto BF TD 1 2 3 4 5 6 7 8 9 0 _ Back space Esc F1 F2 F3 F4 F5 F6 F7 Delete Line Delete Abc Delete Arrow Delete Meas PrtSc Pause Tab Q W E R T Y U I O P Caps Lock A S D F G H J K L Enter Shift Z X C V B N M Shift Ctrl Fn Alt Ins Del 101 102 103 104 105 106 107 109 110 111 121 122 120 123 124 126 127 128 131 132 133 134 135 136 138 140 141 142 144 145 146 148 151 153 155 156 157 ...

Страница 182: ...istribution 111 Dual Format H Dual Screen format horizontal distribution not yet implemented 120 Mode_Switch1 M Mode Motion mode 121 Mode_Switch2 PD Mode Power Doppler 122 Mode_Switch3 PW Mode Pulsed Wave Doppler 123 CW Continuous Wave Doppler 124 HI Harmonic Imaging 126 Text Image documentation to write onto the screen 127 Bodymark Bodymark display to enter Bodymark symbols 128 ClrScreen to clear...

Страница 183: ...emented 160 Save 2D to save send 2D images and sequences 162 NOT USED 163 Exit Exit 164 Utilities Utilities Menu 165 Invert Invert 166 Left Right image orientation of 3D image 167 UP Down up down image orientation 168 Mirror left right image orientation 169 Init reset to initial position in volume mode 170 Steering Beam steering 172 NOT USED 173 OTO Automatic Optimization 174 4D Mode 4D Mode ON OF...

Страница 184: ...h function 187 Soft_Flip 3 DOWN Soft Flip 3 Toggle switch function 188 Soft_Flip 4 UP Soft Flip 4 Toggle switch function 189 Soft_Flip 4 DOWN Soft Flip 4 Toggle switch function 190 Menu Selector Wheel Switch Menu Navigation wheel 191 Change Menu Change Menu 192 Measure_User Basic Measurements NOTICE Key codes which are not listed in the table are not available at Voluson 730Pro 730ProV Table 5 5 V...

Страница 185: ...lso used to select the function in the Main Menu or to save the setting in the Sub Menu Push for 3sec b DECREASE ᇪ LEFT Use this button to move down the OSD selection menu and adjust the attribute of the monitor while in OSD mode Pressing this button out of the OSD menu allows you to decrease the level of contrast of the display screen c INCREASE ᇬ RIGHT Use this button to move up the OSD selectio...

Страница 186: ... External I O Panel Connectors Table 5 6 External I O Connector Description Item Connector Name Description 1 VGA OUT print out VGA signal with monitor printer 2 MODEM RJ 11 with global adapter kit for modem connection 3 NETWORK DICOM input output twisted pair RJ 45 10 100 megabit s 4 USB 1 USB port 5 USB 2 USB port 6 COMP VIDEO OUT BNC Connector Color Video Output 1 2 3 4 5 6 ...

Страница 187: ...cation on the heart 5 10 1 2 DVD CD R W Drive Recordable and Re writable DVD s and CD s are ideal for any storage purpose and offer complete security and reliability for important data 5 10 1 3 Magneto Optical Drive optional Storage capacity by disk 1 3GB 640MB 540MB 230MB 128MB The MO Drive allows to read and write any GIGAMO standard 1 3GB disk at twice the liner bit density Additionally it reta...

Страница 188: ... the current grows to high the switch will automatically break the power From the Mains Switch the AC power is routed via an Inrush Current Limiter to a internal outlet connector for the Mains Transformer 5 11 1 2 Major Functions of CPN6 Inrush Current limiter Power factor correction transformer for Sinus load for the mains voltage Power down Circuitry Standby Switch The CPN6 module generates 57VD...

Страница 189: ...ircuit breaker 2 5 5A F3 F4 F5 F6 are the fuses for the input voltage for the switching power supplies generating the DC Supply voltage for the Secondary power supply inside GEF Module 5 11 1 4 Fuses inside CPN6 F1 on CCF board fuses the surge current limiter circuit NOTE If this fuse is blown the NTC limiting the surge current will remain hot during system operation and if the system is switched ...

Страница 190: ...ns Power Input module the AC power is routed via an Inrush Current Limiter to a internal outlet connector for the Mains Transformer 5 11 2 2 Major Functions of CPN80 Inrush Current limiter Power factor correction transformer for Sinus load for the mains voltage Power down Circuitry Standby Switch The CPN module generates 57VDC 2V as an input voltage for the Secondary Power supply of the GEF module...

Страница 191: ... on CCF board fuses the surge current limiter circuit NOTE If this fuse is blown the NTC limiting the surge current will remain hot during system operation and if the system is switched off on within a few seconds the surge current could be to high Because of this reason Fuse F1 on CPN Rear Panel or the Hospital circuit fuse could be blown 5 11 3 Disk Drive Module GEM The check points for the driv...

Страница 192: ...0PROV BT 04 DIRECTION KTI105947 REVISION 2 SERVICE MANUAL 5 46 Section 5 12 Mechanical Descriptions Section 5 12 Mechanical Descriptions 5 12 1 Physical Dimensions Figure 5 30 Physical Dimensions 668 607 5 851 8 1041 1 1415 678 29 337 5 399 5 456 516 560 5 ...

Страница 193: ...12 2 Monitor Tilt 11 forwards and backwards Swivel 90 rotation 5 12 3 Top Console Positioning 5 12 4 Rotation of the Control Console Horizontal Access The control panel offers 30 of horizontal adjustment to the right Figure 5 31 Top Console Positioning Figure 5 32 Locking lever under Control Console 30 Locking Lever for locking and unlocking the control console ...

Страница 194: ...S KRETZTECHNIK RAFT VOLUSON 730PRO 730PROV BT 04 DIRECTION KTI105947 REVISION 2 SERVICE MANUAL 5 48 Section 5 12 Mechanical Descriptions 5 12 5 Assembly Drawing GW GEU Monitor Figure 5 33 Assembly Drawing GW GEU Monitor ...

Страница 195: ...is draw air into the scanner through the filter grid and pushes it through the card rack Air holes in the left and right side of the rack allow the air to move down through the card rack Air exits the scanner through holes the Main Air Outlet at the left side panel 5 13 1 1 Air Flow Distribution Overview Figure 5 34 Console Views Figure 5 35 Air Flow Control Main Air Outlet CPN Air Outlet Main Air...

Страница 196: ...nd service costs This web enabled technology provides linkage to e Services e Commerce and the iCenter making GE s scanners more e enabled than ever The Service Platform contains specific software hardware test modules system setup update etc for Voluson systems see Section 5 15 Service Page on page 5 52 5 14 2 Access Security The Service Platform has different access and security user levels Each...

Страница 197: ...Interactive Platform Many of the services of the Service Platform come from its integration with iLinq If an iLinq checkout was performed on the Voluson 730Pro 730ProV the ILINQ selection item is available in the menu area on the left side of the screen of the Utilities menu By selecting this item the Netscape browser starts the iLinq Home Page Figure 5 36 iLinq Home Page ...

Страница 198: ...Voluson systems only 5 15 2 Access Security The service page has different access and security user levels Each user is only granted access to the tools that are authorized for their use 5 15 3 Service Login 1 Press the UTILITIES key on the Control Panel The menu area changes to the Utilities menu 2 Select the SYSTEM item from the menu area to activate the setup desktop screen 3 Select the SERVICE...

Страница 199: ...n of customer activities It records all user actions scanning entries performing Calculations review of Patient Reports etc It is possible to safe record as file on HDD But also export to DVD CD or MOD can be done to allow replay of the records on other units Operation see Section 7 5 How to use the Auto Tester program on page 7 8 Figure 5 38 Service Tools window NOTICE For intermittent problems t...

Страница 200: ...EXPORT SYSTEM DATA button on the Service Tools page to export dump files and text files the full Service Database informations about probes boards Software Options Service Actions and the Event Log File to the DVD CD Drive or the optional MO Drive Operation see Section 7 4 2 Export Log s and System Data on page 7 5 5 15 3 6 Maintenance Report Any modification upgrade and maintenance action should ...

Страница 201: ...k the EXIT button on the Service Tools window and the EXIT button on the System Setup Service page NOTE After Hardware or Software modifications normally the Maintenance Report message box see Figure 5 41 appears automatically on the screen 5 15 3 7 Monitor Test 1 To perform the Monitor test select the MONITOR TEST button The following message appears on the screen 2 Read the displayed instruction...

Страница 202: ...r Installation manually on page 3 28 5 15 3 9 Update 5 15 3 9 1 NLS Not for use in the field 5 15 3 9 2 EUM is for updating the Electronic User Manual Operation see Section 8 6 Electronic User Manual EUM Upgrade Procedure on page 8 11 5 15 3 9 3 UIS is for updating the Ultrasound Application Software Operation see Section 8 2 5 Software Installation Procedure via Service Page on page 8 5 5 15 3 9 ...

Страница 203: ... out of specification Although some tests may be optional they should only be performed by qualified personnel Section 6 2 Regulatory Verify where applicable that any regulatory information or tests required by national law are present and accounted for and any regulatory tests required by national law are performed and documented Table 6 1 Chapter 6 Contents Section Description Page Number 6 1 Ov...

Страница 204: ...Menu or to save the settings in the Sub Menu Push for approx 4 sec Left Decrease Use this button to move down the OSD selection menu and adjust the attribute of the monitor while in OSD menu Pressing this button out of the OSD menu allows you to decrease the level of contrast and brightness of the display screen Right Increase Use this button to move down the OSD selection menu and adjust the attr...

Страница 205: ...nu 4 Confirm the Monitor Test message with YES The Monitor Test main menu appears on the screen 5 Select the Color calibration icon and press the right left trackball key once monitor is white 6 Press and hold the MENU middle key on the monitor controls for approximately 4 seconds till the OSD menu appears on the screen 7 Press the RIGHT button 3 times to the COLOR TEMP and then press the MENU key...

Страница 206: ...on to decrease increase the value 3 Press the MENU key again to toggle to brightness 4 Adjust the BRIGHTNESS by pressing the LEFT or RIGHT button to decrease increase the value 6 3 1 4 Geometry Check the raster lines of the test pattern by referring to Figure 6 2 Expected results 1 Squares are not distorted 2 Raster lines are straight 3 No keystone and pincushion distortion 4 Horizontal and vertic...

Страница 207: ...d position 6 4 2 Transport Lock There is a locking lever for locking and unlocking the control console mounted at the front below the control console When preparing the system for transport the lock has to be engaged in order to secure the console against uncontrolled rotation The lock catches in when the console is rotated to its center 0 position Figure 6 3 Locking Lever under Control Console WA...

Страница 208: ...fixing ring by turning it counterclockwise 2 Adjust the trackball for smooth running by rotating the black securing ring CCW smooth run CW tighten run 3 Mount the outer fixing ring by turning it clockwise see Figure 6 4 Figure 6 4 Trackball with fixing ring Figure 6 5 Trackball Securing Ring NOTICE Avoid tightening of thread caused by improper mounting of securing ring rotate counterclockwise to r...

Страница 209: ...run at the application level 7 1 2 Overview There may be a time when it would be advantageous to capture trouble images and system data logs for acquisition through remote diagnostics InSite or to be sent back to the manufacturer for analysis There are different options to acquire this data that would give different results Table 7 1 Contents in Chapter 7 Section Description Page Number 7 1 Overvi...

Страница 210: ...ect the SYSTEM item and then enter the SERVICE INFO page System Type System Serial number also visible on label on the back of the system Application Software version Backup Version File Name additional information e g Hardware ID SLOT_CPU version etc Figure 7 1 System Setup System Info page e g V730 Pro Serial number Application Software Move the scroll bar downwards to review additional informat...

Страница 211: ...le see Figure 7 5 on page 7 7 Address Module Address Application Software version Memory Dump number complete 4 Press the ALT D key on the alphanumeric keyboard to save export a screenshot of the point in time when ALT D was pressed the current System state as Full Backup on demand and logs and dump files already stored on the D export folder on the D export folder or on DVD CD Operation see Secti...

Страница 212: ...n 730Pro 730ProV upwards and check the status of the Diagnostic LED s LED 6 LED 5 LED 4 LED 1 LED 2 LED 3 In case of problems check the above voltages with a Digital Volt Meter DVM to Ground 15V 15V 3 3V 5V 12V LED 1 2 and 5 ECG connector 8 1 9 15 Pin 15 5V Pin 13 15V Pin 11 15V LED 4 AUX Supply GEM Power connector LED 6 Footswitch connector on Power Supply Module CPN on Backpanel of GEF Box Pin 1...

Страница 213: ...ll Backup of the System state fullbackup fb1 saved on D export by pressing the ALT D key on the alpha numeric keyboard 7 4 1 Capturing a screen The following is the generic process to capture any screen from the scanner 3 Select the stored image s in Sonoview Exam List and EXPORT the image s to DVD CD R W or MO Disk jpg bmp or tiff 7 4 2 Export Log s and System Data There are two possibilities to ...

Страница 214: ... investigated by R D It is not intended to replace or enhance the existing Full Backup functionality Data can be stored on the hard disk D export fullbackup fb1 or you can export them to DVD CD including the D export folder which contains logs and dump files for details see Section 7 4 2 2 1 NOTE Successive ALT D keystrokes overwrite existing snapshots at destination HDD 1 Press the ALT TAB key on...

Страница 215: ...0 4 Click on the EXPORT SYSTEM DATA button to export dump files see Section 7 4 2 2 1 text files the Event Log File the full Service Database and information about probes boards software options and service actions to DVD CD Drive or the optional MO drive 7 4 2 2 1 Dump file Every time an error message like Figure 7 5 is produced a dump file and a text file containing the error dump and the error ...

Страница 216: ...Tester program 1 Press the UTILITIES key and select the SYSTEM from the menu area on the left side of the screen 2 Select the SERVICE page on the screen The password window appears automatically 3 Enter the password SHE and click the ACCEPT button to display the Service Tools window 4 Activate the Auto Tester program by clicking the START button Figure 7 6 Service Tools window Start Auto Tester ...

Страница 217: ...oard 7 Activate the Auto Tester program by clicking the Record icon on the displayed screen 8 Start scanning You can scan normally and everything will be recorded to the program up to several hours NOTE It is important that you are recording the processes where the errors normally occur 9 Stop the program by pressing the PAUSE key on the right upper corner of the keyboard Figure 7 7 Message Box Fi...

Страница 218: ...r program cont d The following screen will appear 10 Insert an empty DVD CD R W in the Drive and select the DVD CD Burn icon 11 Enter a Filename 12 After clicking OK the following message boxes will appear 13 After the DVD CD write is finished click the OK button and close the Auto Tester program Figure 7 9 Auto Tester Finished Figure 7 10 Enter a Filename Figure 7 11 CD Burn Process DVD CD Burn i...

Страница 219: ...ving a Full Backup The remote service capabilities are also present 1 Power On Boot Up the system as described in Section 3 6 1 on page 3 20 3 Click on CONTINUE The Probe Selection menu appears on the monitor 4 Press the FREEZE key The application starts but does not show any Ultrasound information 5 Backup all images via SonoView see Section 4 5 6 Archiving Images on page 4 40 6 Save a Full Backu...

Страница 220: ... IBASE Standard Board Computer If the PC Board is a Kontron SBC the PS2 cable a has to be connected directly to the PC part of the GEF box b UI Disp connector for User Interface Display NOT used at Voluson 730Pro 730ProV c VGA Monitor d Console e CPN Primary Power Supply f Standby switch g Monitor Power Supply Connect mains power cable to the system and to an appropriate mains power outlet Connect...

Страница 221: ...rst items checked if the scanner doesn t boot up If all three conditions are present the scanner should be in a standby mode with the ON OFF switch illuminated with amber color Follow the flow chart to determine the reason the scanner doesn t boot up No No Switch in ON Position Fan Noise Yes Check if the power switch over the mains plug of the Primary Power Supply isn t switched off Check the conn...

Страница 222: ...7 14 Section 7 7 Troubleshooting Trees and Instructions 7 7 2 Noise disturbs the Image Figure 7 16 Noise disturbs the Image Troubleshooting Noise is shown in Image YES NO Is there any electrical devices that affect the Scanner Keep the Scanner away from those electrical devices use another wallsocket Check probes ...

Страница 223: ...TI105947 REVISION 2 SERVICE MANUAL Chapter 7 Diagnostics Troubleshooting 7 15 7 7 3 Trackball Figure 7 17 Trackball Troubleshooting The trackball is low sensitive Remove the ball from the Trackball unit and clean it YES NO Does the trackball move at all Replace the trackball unit ...

Страница 224: ...ctions 7 7 4 System Does Not Power Off Shutdown Figure 7 18 Power Off Shutdown Troubleshooting System has been switched off System starts with the shut down procedure System is down Yes system switched off automatically in approx 3min Primary Power CPY inside Supply is defekt Check the Standby switch and the cable If ok Primary Power Supply is defekt No Yes No ...

Страница 225: ...olored streaks appear in image Check for presence of magnetic sources near the monitor Eliminate the sources and then degauss the monitor Screen image is not centered or sized properly Adjust the picture location picture size picture rotation or pincushion distortion Some video modes do not fill the screen to the edge of the monitor There is no single answer to solve the problem This phenomenon ma...

Страница 226: ...Troubleshooting Unable to record to VCR Check that all signal and power supply cable connections to the VCR unit are OK YES NO Is the VCR properly connected Is the VCR tape put into the device Put the tape into the device and rewind it NO YES Replace VCR Is recording unable even by manual operation YES Check the signal cable connections between VCR and Internal I O NO YES Video Output of System o ...

Страница 227: ...nter Troubleshooting Figure 7 20 Printer Troubleshooting Unable to use Printer s Check that all signal and power supply cable connections between Printer and Scanner are OK YES NO Is the printer s properly connected Are there print papers left Insert the printer papers NO YES Is proper configuration for the Print key s set Configure the Print key s NO See next page ...

Страница 228: ...and Instructions 7 7 7 Printer Troubleshooting cont d Figure 7 21 Printer Troubleshooting cont d Still unable to use the Printer s Check the signal cables between GEF and Backpanel YES NO Is printing unavailable even by manual operation Still no image on the print out Video Output of System o k NO YES Printer o k Replace the Printer Replace CKV NO YES ...

Страница 229: ...VIEW key on the control panel see Figure 7 22 3 Click the Open icon to display the list of exams 4 Select exam s and backup them to MO 5 Choose MO Drive 6 The images which you have chosen during backup should be visible Figure 7 22 To backup exams to DVD CD R W resp MO Disk NOTICE In case of any problems check status of the Diagnostic LED s and voltages of the AUX Supply connector Disk Drive Modul...

Страница 230: ...t a network cable between your Scanner and your PC Try to ping from the Scanner to the IP address on the PC If OK the hardware connection inside the Scanner is OK 7 7 10 2 GE remote service connection The customer gets visual information about the network status The network status is defined as follows NOTE The status is checked every 10 seconds regardless whether you re in read or write mode BLAC...

Страница 231: ...ok and save this user setting once again B_Gain Out Of Range press ok and save this user setting once again B_Reject Out Of Range press ok and save this user setting once again B_TxFocus not calculated and B_SHOT_PART_ON restart the system Backup error while writing Error during writing of backup data check storage destination for Full Backup e g DVD CD not empty insufficient rights on target Netw...

Страница 232: ...port session use a new CDR to write data to CD Can t open file restart the system can t open MotCtrl RS232 Serial connection restart the system Can t open restart the system Can t open C V730 Distribution Hardware patterndat bin restart the system can t write XilinxFiles ini restart the system CD write error use a new CDR Cine2D_CtrlBlock vSetState value of m_GIP_eCineType undefined restart the sy...

Страница 233: ...d save this user setting once again Different software error Backup data was made on another system with a different software version This backup cannot be restored by the user Disc full use a new CDR for writing data to CD Display Rect Region fails restart the system Division by zero restart the system done is low restart the system Downgrade error Backup data was made with a software version hig...

Страница 234: ...and is on ERROR_MSG_NO_ACK check VCR cables and try again ERROR_MSG_NO_CASSETTE put cassette into drive of VCR ERROR_MSG_NO_RESPONSE check VCR cables casette and try again ERROR_MSG_WRITE_PROTECTED remove cassette from VCR and put writeable cassette into drive of VCR Execute privileged instruction restart the system FallSmooth Out Of Range press ok and save this user setting once again File Could ...

Страница 235: ...amic cast failed restart the system GeoDescriptionTissueCF Constructor unknown mode Parameter handed over restart the system GeoDescriptionTissueCF operator handed over argument is not of same type dynamic cast failed restart the system Hardware doesn t support CW mode pencil probe CW Hardware not available HW problem hardware error on restart the system HardwareRelatedSoftware_Windows in write ha...

Страница 236: ...AcousticUnit wrong ProbeAcousticUnitID restart the system missing ProbeGeneral data wrong ProbeID restart the system missing ProbeScanFunc wrong ProbeScanFuncID restart the system MotCtrl RS232 communication timeout restart the system MotCtrl No Referenzposition signal Confirm the message by pressing the OK button disconnect and reconnect the active probe and use it again If the message appears ag...

Страница 237: ...e system PW_BaseLinePos Out Of Range press ok and save this user setting once again PW_BurstCalcBlock UserProgApplication out of range restart the system PW_CW_FFT_FactBlock DSC_ScrollX_Zoom darf nicht kleiner als eins sein restart the system PW_Dynamic Out Of Range press ok and save this user setting once again PW_Reject Out Of Range press ok and save this user setting once again PWCW DSC Memory ...

Страница 238: ...stem RepresentationManager vResizeBufferLength caused exception replay buffer is not empty restart the system RepresentationManager vSetGeoDescription called within Transaction restart the system RepresentationManager vSetReplayParams called within Transaction restart the system Restore error Error while reading backup data Backup data are probably damaged Try again or load another backup RiseSmoo...

Страница 239: ...too small restart the system Start frequency too small restart the system StoragePool Not sufficient replay storage for B CF Mode available restart the system StoragePool Not sufficient replay storage for ECG available restart the system StoragePool Not sufficient replay storage for M Mode available restart the system StoragePool Not sufficient replay storage for MC or PW Mode available restart th...

Страница 240: ...volume files from other storage medium unknown Xilinx Version restart the system unrecordable disc in drive try again with another disk Unsupported color mode restart the system Verify error Error while checking backup data Backup data are probably damaged Try again or load another backup ViewerConMngr vAssembleCF_DFE attempt to get ECG_Consumer_2D or ECG_Draw_2D Block from ECGViewer Objects faile...

Страница 241: ...al and electronic equipment must not be disposed as unsorted municipal waste and must be collected separately Please contact the manufacturer or other authorized disposal company for information concerning the decommission of your equipment Table 8 1 Chapter 8 Contents Section Description Page Number 8 1 Overview 8 1 8 2 Ultrasound Application Software UIS Installation Procedure 8 2 8 3 User Setti...

Страница 242: ...e purpose of the regulation is to protect employees in the transportation industry as well as the people who will receive or open this package NOTE The US Department of Transportation DOT has ruled that items what were saturated and or dripping with human blood that are now caked with dried blood or which were used or intended for use in patient care are regulated medical waste for transportation ...

Страница 243: ...EM INFO page on the System Setup desktop screen to see which Software Hardware version is installed in the unit 4 Select the OPTION page on the System Setup desktop screen to see which Options are installed Figure 8 1 Version check Figure 8 2 Option page NOTICE Please print out the OPTION page or write down all the option codes which are shown in the Key fields currently installed Software version...

Страница 244: ...n the DICOM SONOVIEW CONFIGURATION button in the NETWORK page b Click on the NETWORK CONFIGURATION button in the NETWORK page Figure 8 3 DICOM Configuration NOTICE Please print out the DICOM Configuration dialog page or write down the DICOM designations AE Title Station name Retry Count Retry Interval Figure 8 4 Internet Protocol TCP IP NOTICE Please print out the Internet Protocol TCP IP Properti...

Страница 245: ...nstallation Procedure via Service Page 1 Insert the System DVD into the drive 2 Press the UTILITIES key and then select SYSTEM item from the Utilities menu 3 Select the SERVICE page 4 Type in the password SHE and click ACCEPT 5 Click on the UIS button for updating the Ultrasound Application Software Figure 8 5 System Setup Service Figure 8 6 Service Tools UIS button ...

Страница 246: ...ocedure via Service Page cont d 6 Confirm the following message with YES 7 A new window pops up on the screen and the InstallShield starts extracting files 8 Click the NEXT button In the Welcome to Kretz V730 Software Installation window The contents of the Software package are being extracted Figure 8 7 Update UIS message Figure 8 8 Install Shield Wizard Extracting Files Figure 8 9 Welcome to Sof...

Страница 247: ...ISH key The system restarts automatically 10 If desired load the appropriate factory settings see Section 8 3 on page 8 or the full backup see Section 8 4 on page 8 8 11 If necessary update the Service Software SSW as described in Section 8 5 on page 8 9 12 Afterwards update the Electronic User Manual EUM as described in Section 8 6 on page 8 11 13 Perform Software and Functional checks as describ...

Страница 248: ...ntains following data Patient demographic and exam data database containing the patient data and measurements SonoView image data NOT available when saving to the internal hard disk DVD CD or MOD User Settings databases and files containing gray curves and the user settings Image transfer settings DICOM settings e g DICOM servers AE Title Station Name etc Measure Setup Settings user specific measu...

Страница 249: ... Upgrade Procedure 1 Insert the System DVD or Upgrade CD into the drive 2 Press the UTILITIES key on the control panel once to display the Utilities menu 3 Select the SYSTEM item from the Utilities menu which is displayed on the left side of the screen 4 Select the SERVICE page 5 Enter the password SHE and click the ACCEPT 6 The Service Tools menu appears on the screen 7 Click on the SSW button fo...

Страница 250: ...VE button 9 Browse for the install bat file and click OK 10 Confirm the following message with YES 11 A new window pops up on the screen and the InstallShield starts extracting files 12 Press any key to continue to finish Service Platform installation procedure 13 Close the SERVICE page with EXIT and close the System Setup with SAVE EXIT Figure 8 12 Browse for Folder Figure 8 13 Update SSW message...

Страница 251: ... 2 Insert the System DVD or Upgrade CD into the drive 3 Press the UTILITIES key on the control panel once to display the Utilities menu 4 Select the SYSTEM item from the menu area to invoke the setup desktop on the screen 5 Select the SERVICE page 6 Enter the password SHE and click the ACCEPT 7 The Service Tools menu appears on the screen NOTICE The electronic user manual EUM must never be opened ...

Страница 252: ...n 3 Browse for the EUMSetup_en_V730Pro exe file and click OK 4 Confirm the following message with YES to proceed the update The contents of this package are being extracted 5 Confirm the following warning message with OK to proceed the update 6 Confirm the Maintenance Complete message InstallShield Wizard has finished performing maintenance operations on V730 Pro User Manual with the FINISH button...

Страница 253: ...s of the InstallShield Wizard confirm the messages with the NEXT key 12 After successfully installation of the Electronic User Manual click FINISH to exit the wizard 13 Close the SERVICE page with EXIT and close the System Setup with SAVE EXIT 14 Restart the system turn off and on the system 15 After rebooting the system fill in the requested information into the System Status Messenger box and cl...

Страница 254: ...he OPTION page see Figure 8 2 on page 8 3 If necessary customize the settings according to the printout 5 Check and if necessary match the NETWORK page settings according to the printout DICOM Configuration dialog page see Figure 8 3 on page 8 4 Network Configuration dialog page see Figure 8 4 on page 8 4 6 Restart the system and perform basic functional checks to ensure system is functioning norm...

Страница 255: ...e to display the Utilities menu 2 Select the SYSTEM item from the menu area to invoke the setup desktop on the screen 3 Select the OPTION page where you can see which options are installed in the system D Demo Option is activated for demo and expires on the date shown in the Valid column I Inactive Option is not activated P Permanent Option is permanently activated purchased NOTICE Not all Options...

Страница 256: ...ling a Demo Key or a Permanent Key 1 Position the cursor inside the input field desired and press the right left trackball key 2 If one exists clear edit the current key code 3 Enter the encrypted serial code with the keyboard and then click on SUBMIT The code will be checked 4 Click the SAVE EXIT button NOTE After activating a key code restart turn of and on the Voluson 730Pro 730ProV system ...

Страница 257: ...e and Images from System to System 8 9 1 Introduction NOTICE It is possible to backup Sonoview Image data to a Mapped Network Drive Z WARNING WARNING Please make sure that the server you are connecting to is trustworthy and reliable For details contact your local system administrator If you backup Sonoview data to this server all the patients demographic data will be copied to this server ...

Страница 258: ...ess the SONOVIEW key on the Control panel 2 Click on the OPEN icon on the upper left side of the screen and verify that Drive HDD is selected 3 Using the TRACKBALL and the right trackball key SET select the first exam of the list 4 To select all available exams click on SELECT TO END and then on the BACKUP button 5 In the Backup window select the destination NET 6 After finishing the backup select...

Страница 259: ...e this name of the old system s Backup Folder e g A09008 and then click CANCEL 8 9 2 5 Restore all Exams of the old system to the new system 1 On the system new system that should receive data and images press the SONOVIEW key on the Control panel 2 Select the SETTINGS icon on the left side of the screen 3 In the displayed window click on the CHANGE BACKUP FOLDER ON NETWORK DRIVE button 4 Choose f...

Страница 260: ...tton 10 In the Backup Restore window click on the RESTORE button NOTICE The folder which is selected to be renamed e g A09008 that contains the backup data from a different system in this case the backup data of the old system it is not copied but simply renamed Thus the same data can be shared between two systems by renaming the respective backup folders to the serial number of the accessing syst...

Страница 261: ... hard disk is replaced with the exam in the backup The system will ask again if another identical exam is found during the restore process Yes to all All identical exams are replaced without further notice No The exam on the hard disk is not replaced by the exam in the backup The system will ask again if another identical exam is found during the restore process No to all No identical exams are re...

Страница 262: ... 1 and 2 stubby length 30mm 8 10 3 Preparations 1 Power Off Shutdown the system see Section 3 6 2 on page 3 21 8 10 4 Task Lamp Removal Procedure 1 Use the stubby screwdriver and unscrew the 2 screws which fixes the cover of the task lamp 2 Screw out the task lamp and remove it 8 10 5 Task Lamp Installation Procedure 1 Screw in the new task lamp into the lamp socket 2 Mount the cover of the task l...

Страница 263: ... counterclockwise 2 Mount the fixation ring by turning it clockwise Section 8 12 Replacement of Digipots and TGC Slider controls 8 12 1 Manpower One person 5 min 8 12 2 Tools small sized slotted screwdriver or tweezers 8 12 3 Cap and or Spring Replacement Procedure 1 Remove the cap for Rotation digipots or for Slider potentiometer TGC 2 Mount the cap for Rotation digipots or for Slider potentiomet...

Страница 264: ...CAL SYSTEMS KRETZTECHNIK RAFT VOLUSON 730PRO 730PROV BT 04 DIRECTION KTI105947 REVISION 2 SERVICE MANUAL 8 24 Section 8 12 Replacement of Digipots and TGC Slider controls This page was intentionally left blank ...

Страница 265: ...le 9 1 Contents in Chapter 9 Section Description Page Number 9 1 Overview 9 1 9 2 List of Abbreviations 9 2 9 3 Parts List Groups 9 3 9 4 Housing GW and additional Console Hardware 9 4 9 5 User Interface GEU Top Console 9 7 9 6 Monitor Monitor replacement parts 9 9 9 7 Disk Drives GEM 9 10 9 8 Main Power Module CPN 9 11 9 9 Main Board Module GEF 9 13 9 10 Options and Upgrades 9 18 9 11 Miscellaneo...

Страница 266: ... Switch Board CPZ Cover Board CRB USB to IDE Converter Board CRS Signal Processing Board CRW CW Doppler Board optional EUM Electronic User Manual FRU 1 Replacement part available in parts hub FRU 2 Replacement part available from the manufacturer lead time involved GEF Main Board Module Ultrasound FrontEnd and PC Boards Backend Processor GEM Disk Drive module with or without MO Drive and MAN GES I...

Страница 267: ...N Primary power module 500 510 570 Main Board Module GEF FrontEnd US Part FrontEnd US Part cont d Back Processor PC Part Table 9 9 on page 9 13 Table 9 10 on page 9 14 Table 9 11 on page 9 15 Table 9 12 on page 9 17 GEF Main Board Module Ultrasound FrontEnd PC Boards Backend Processor 600 Options and Upgrades Table 9 13 on page 9 18 700 Miscellaneous Cables Table 9 14 on page 9 19 800 Optional Per...

Страница 268: ...7 REVISION 2 SERVICE MANUAL 9 4 Section 9 4 Housing GW and additional Console Hardware Section 9 4 Housing GW and additional Console Hardware Figure 9 2 Housing GW SNO greater equal A32514 SNO less equal A32513 101 102 103 104 105 106 107 108 109 110 111 112 113 114 115 116 117 118 119 ...

Страница 269: ...eels used in serial production SNO less equal A32513 2 1 109 Rear Wheel Ø175 mm x 32 KTZ211081 Rear wheel non steerable 2 1 110 Steerable Wheel KTZ211080 Front wheel steerable 2 1 111 Blind Cap for rear screws KTZ208109 covers housing screws to make them invisible 2 1 112 Monitor Mounting Plate Voluson 730Pro 730ProV KTZ134110 Monitor mounting plate 1 1 113 Rear Handle for Trolley SNO A32514 KTZ13...

Страница 270: ...ction Cables to the V730 Main Unit internal Rear Panel 1 1 125 Hinge for Foot rest KTZ220018 Hinge for Foot rest 2 1 126 Top Cover of Trolley KTZ208119 Top Cover of Trolley 1 1 127 Standby Switch 24V 15A KTZ207125 Standby Switch left below the control panel 1 1 128 Cable Guide for Probes KTZ14M787 Cable Guide for Probes for SNO A32513 only 4 1 129 Pull out Protection for Mains Power cable CPN6 KTZ...

Страница 271: ...SON 730PRO 730PROV BT 04 DIRECTION KTI105947 REVISION 2 SERVICE MANUAL Chapter 9 Renewal Parts 9 7 Section 9 5 User Interface GEU Top Console Figure 9 4 User Interface GEU 201 202 203 206 207 208 205 204 209 210 211 212 213 214 215 216 217 218 ...

Страница 272: ...cludes 8 pieces with premounted brackets 8 1 208 Caps for TGC Slide pot 8 pcs KTZ214818 Caps for TGC Slide pot set includes 8 pieces with premounted brackets 8 1 209 Alpha numeric keyboard V730Pro KTZ208217 Alpha numeric keyboard for Voluson 730Pro 730ProV 1 1 210 Keytop Kit Voluson 730Pro 730ProV KTZ207558 Keytop Kit Voluson 730Pro 730ProV 1 1 211 Hard key Board for Voluson 730Pro 730ProV KTZ2082...

Страница 273: ... Cover used in serial production SNO less equal A32513 1 1 252 Monitor Base Cover KTZ134474 Monitor Base Cover silver colored used in serial production SNO greater equal A32514 can replace KTZ134365 1 1 253 Monitor Mounting Set KTZ154713 Monitor Mounting Set fixing part for Monitor only used with ANY Monitor KTZ212115 1 1 254 Set Monitor Front Housing KTZ208445 Set Monitor Front Housing incl cover...

Страница 274: ...303 Fan axial 62 x 62 x 14 mm KTZ207602 Fan for Air Cooling of GEM SCSI Drive Module 1 1 304 Disk Drive IDE MOD 1 3GB internal KTZ207269 Magneto Optical Drive 1 3GB internal no own cabinet 1 1 305 GEM10 10A MO Drive DVD R W KTZ195887 MO Drive and DVD R W 1 2 306 GEM11 11B MO Drive DVD R W KTZ196000 MO Drive and DVD R W can replace GEM10 KTZ195887 SW4 0 3 or higher is required 1 2 307 GEM 20A DVD R...

Страница 275: ...RAFT VOLUSON 730PRO 730PROV BT 04 DIRECTION KTI105947 REVISION 2 SERVICE MANUAL Chapter 9 Renewal Parts 9 11 Section 9 8 Main Power Module CPN Figure 9 7 Main Power Module CPN 401 402 403 404 405 406 407 408 409 410 411 412 413 414 415 ...

Страница 276: ...ith CPN80 KTZ207486 1 1 407 CCF100 P1 Power Primary Board KTZ209339 Power Primary Board only used with CPN80 KTZ207486 1 1 408 Mains Power switch and Input connector KTZ209340 Mains Power Switch and Input Connector only used at CPN80 KTZ207486 1 1 409 Fuse 10 Ampere 6 3 x 32 KTZ208239 Electric Current Overflow Protection GEF only used at CPN6 4 1 410 Fuses 16 Ampere 6 3 x 32 10 pcs KTZ209341 used ...

Страница 277: ...AL Chapter 9 Renewal Parts 9 13 Section 9 9 Main Board Module GEF Figure 9 8 Main Board Module GEF Table 9 9 Main Board Module GEF Item Part Name Part Number Description Qty FRU 501 Fan for GEF box 2 fan KTZ154678 Fan for GEF box 2 fan 1 1 502 Fan for GEF box single KTZ154679 Fan for GEF box single 1 1 502 501 ...

Страница 278: ...ection between CPU CPR and CW Board contains Relays for Signal to CW Board 1 1 513 CRS4 P3 Signal Processing Board KTZ195998 Signal Processing Board OBSOLETE please order CRS4d KTZ196080 or CRS5 KTZ196074 1 1 514 CRS4d P3 Signal Processing Board KTZ196080 Signal Processing Board replaces CRS4 KTZ195998 1 1 515 CRS5 P3 Signal Processing Board KTZ196074 Signal Processing Board can replace CRS4 KTZ19...

Страница 279: ...art cont d Table 9 11 FrontEnd US Part cont d Item Part Name Part Number Description Qty FRU 518 CPK80c P5 Motherboard V730 KTZ207456 Electrical Signal and Supply Connection for all boards including PC Motherboard CPM 1 1 519 CPP81 82 P3 Power Supply Board KTZ207274 Power Supply Board Output Power 900 W 1 1 520 CPU5 P5 Module Board BYM KTZ195636 Probe Connector Board Module Board 1 1 518 519 520 ...

Страница 280: ...TECHNIK RAFT VOLUSON 730PRO 730PROV BT 04 DIRECTION KTI105947 REVISION 2 SERVICE MANUAL 9 16 Section 9 9 Main Board Module GEF 9 9 2 BackEnd Processor PC Part Figure 9 11 Back Processor PC Part 571 572 573 574 575 576 577 578 ...

Страница 281: ...for all Voluson 730 V730 V730Expert and V730Pro systems KTZ196003 Universal Hard disk System Boot DVD see Table 9 13 is required 1 1 575 SBC BT 03 04 KTZ208570 Standard Single Board PC IBASE IB820 Pentium IV 3 06GHz 400MHz FrontSideBus PCI Bus 768MB Ram can replace KTZ209332 but system C image MUST be loaded In this case System Boot DVD see Table 9 13 is required 1 1 576 SBC BT 03 04 K KTZ209332 S...

Страница 282: ... BT 05 only KTZ196075 bootable DVD for System HDD recovery SW5 0 1 Contents SP4 with newest MS patches updated Linux rescue partition System C Image Windows 2000 SBC for BT 05 upgrades supplied UISApp Backup EUM Database Repair Tool etc N 1 604 CW Doppler Upgrade Kit H48641AA CW Doppler Upgrade Kit incl upgrade instructions N N A 605 Voluson 730Pro 730ProV BT 04 BT 05 Upgrade Kit H48651FP Contents...

Страница 283: ...Part Number Description Qty FRU 701 Audio Adapter KTZ208439 Set contains two adapters Audio Adapter Line IN OUT used in combination with SBC Board KTZ208570 1 1 Audio Adapter Line IN OUT used in combination with SBC Board KTZ209332 702 Adapter slot parallel serial KTZ208148 Adapter Parallel port serial COM1 1 1 703 Adapter Cable PS2 KTZ208137 Adapter for GEU PS2 cable only used in combination with...

Страница 284: ...yellow black red black 1 1 705 Power Cable for Harddisk KTZ212401 Power cable for Harddisk 4pin 230mm 1 1 706 Cable ATX 12Volt KTZ212449 HDD power connector to Pentium 4 power connector 1 1 707 Cable Stereo Jack Chinch KTZ212074 Cable from PC Sound StereoJack to External Rear Panel 1 1 708 Data Cable HDD ATA66 100 KTZ208147 Data cable for IDE Hard disk 40pin 1 1 Table 9 14 Miscellaneous Cables Ite...

Страница 285: ...EM GEF KTZ195464 Electrical Power Supply for Drives 5V 12V 1 1 711 KUG5 Remote Cable KTZ195606 VCR Remote Control Cable 1 1 713 KVS2 Cable Serial Interface KTZ195717 electrical Signal Connection between CPS and PIC_MG Slot CPU 1 1 714 KVX1 Network Cable KTZ212016 Cable from external rear Panel to the Voluson 730Pro 730ProV Main Unit internal rear Panel 1m 1 1 715 Monitor Power Connection Cable KTZ...

Страница 286: ...ower Cord Europe 230V KTZ212317 Power Cord Europe 230V 240V 1 1 717 Power Cord Japan Hosp grade KTZ212448 Power Cord Japan Hospital Grade 1 1 718 Power Cord UK KTZ212441 Power Cord United Kingdom 240V 1 1 719 Power Cord USA Hosp Grade KTZ212402 Power Cord USA Hospital Grade 1 1 Table 9 14 Miscellaneous Cables Item Part Name Part Number Description Qty FRU ...

Страница 287: ...panel used in combination with SBC Board KTZ208570 2 1 USB for PC Slot Connector on Backpanel Cables are connected to PC Board Leads the USB signals to the PC Backpanel used in combination with SBC Board KTZ209332 721 USB Cable A to B from GEM to PC part on GEF KTZ212125 USB Cable from Disk Drive Module GEM to connector USB GEM on the PC part of main board module GEF 1 1 722 VGA cable HD15 HD15 KT...

Страница 288: ...s Cable TMDS Digital Video Cable from SBC to CKV used in combination with SBC Board KTZ208570 1 1 Cable TMDS Digital Video Cable from SBC to CKV used in combination with SBC Board KTZ209332 724 Modem Data Cable KTZ212199 Data Cable for optional Global Modem 1 1 725 Modem Power Connection Cable KTZ212032 Power Connection Cable for optional Global Modem 1 1 Table 9 14 Miscellaneous Cables Item Part ...

Страница 289: ...ON 730PRO 730PROV BT 04 DIRECTION KTI105947 REVISION 2 SERVICE MANUAL Chapter 9 Renewal Parts 9 25 Section 9 12 Optional Peripherals and Accessories Figure 9 13 Optional Peripherals and Accessories 801 802 803 804 805 806 807 808 809 810 811 812 ...

Страница 290: ...P D23MD 1 1 805 Bluetooth Printer HP 5600 Series KTZ211503 H46631L USB Port Bluetooth Connection Set is required 1 1 806 Bluetooth Printer Olivetti Job Jet 210 KTZ211503 H46631L USB Port Bluetooth Connection Set is required License product of HP5600 1 1 807 Bluetooth Connection Set KTZ196002 H46631M Bluetooth Connection Set HP 5600 Series only 1 1 808 Sony VCR SVO 9500 MD KTZ211317 H46801D NTSC 1 ...

Страница 291: ...RU 812 Foot switch MFT7 KTZ195446 H46681D Foot switch 1 1 813 ECG preamplifier MAN 6 KTZ154644 H46681H consists of ECG preamplifier and patient connection cable 1 1 814 Global Modem Analog 2245794 H48641C Global Modem analog Common Part with L9 and V7 use modem cable KTZ212199 KTZ212032 1 1 815 USB RS232 Connection kit PRY KTZ195858 H46681S Converter from USB to RS 232 Serial Port Connection Modul...

Страница 292: ...n Manual Voluson 730Pro 730ProV English H48641BJ KTI105939 1 N Instruction Manual Voluson 730Pro 730ProV German H48641BK KTI105940 1 N Instruction Manual Voluson 730Pro 730ProV Spanish H48641BL KTI105941 1 N Instruction Manual Voluson 730Pro 730ProV Portuguese H48641BM KTI105942 1 N Instruction Manual Voluson 730Pro 730ProV Italian H48641BN KTI105943 1 N Instruction Manual Voluson 730Pro 730ProV F...

Страница 293: ...d array transducer frequency range of 2 5 MHz Applications Abdominal Obstetrics Gynecology Pediatrics 1 903 4C A KTZ156952 H46701AA BT Version electronic broadband curved array transducer frequency range of 2 5 MHz Applications Abdominal Obstetrics Gynecology This probe is only applicable if system were upgraded to BT 05 SW5 x x 1 904 IC5 9 KTZ195386 H46701F electronic endocavity broadband curved ...

Страница 294: ... Orthopedics 1 907 SP6 12 KTZ195362 H46701B electronic broadband linear array transducer frequency range of 6 12 MHz electronically steerable Applications Small Parts Peripheral Vascular Pediatrics Orthopedics 1 908 SP10 16 KTZ195531 H46701C electronic broadband linear array transducer frequency range of 10 16 MHz electronically steerable Applications Small Parts Orthopedics 1 909 PA2 5P KTZ195773...

Страница 295: ...621X Real time 4D broadband electronic curved array transducer with a frequency range of 2 5MHz Applications Abdominal Obstetrics Gynecology Interventional Radiology 1 924 RAB4 8L KTZ156846 H48621Z Real time 4D broadband electronic curved array transducer with a frequency range of 4 8MHz Applications Abdominal OB Gyn Pediatrics Interventional Radiology 1 925 RSP5 12 KTZ195763 H46701W Real time 4D ...

Страница 296: ...cations Gynecology Fertility Obstetrics Urology 1 927 RIC5 9H KTZ156920 H48651DA Real time 4D broadband electronic curved array transducer with a frequency range of 4 8MHz Applications Abdominal OB Gyn Pediatrics Interventional Radiology 1 928 RRE6 10 KTZ195534 H46701S Real time 4D Multi Plane electronic broadband curved array transrectal transducer with a frequency range of 6 10 MHz Applications ...

Страница 297: ...robes Item Part Name Part Number Description Qty FRU 931 PCW4 0 KTZ195540 H46701L single element Continuous Wave CW Doppler pencil probe with a nominal operating frequency of 4 0 MHz no B image Applications Cardiology Pediatrics 1 932 SCW2 0 KTZ195538 H46701K single element Continuous Wave CW Doppler pencil probe with a nominal operating frequency of 2 0 MHz no B image Applications Cardiology supr...

Страница 298: ...FT VOLUSON 730PRO 730PROV BT 04 DIRECTION KTI105947 REVISION 2 SERVICE MANUAL 9 34 Section 9 14 Biopsy Needle Guides Section 9 14 Biopsy Needle Guides Figure 9 20 Biopsy Needle Guides 850 851 852 853 854 855 856 857 858 859 860 861 862 ...

Страница 299: ...le guide for probe RAB2 5 RAB4 8P needle diameter 1 mm 1 4 mm 2 2 mm N 856 PEC68 H46721P reusable Biopsy needle guide for probe RSP5 12 RSP6 16 needle diameter 1 mm 1 4 mm 2 2 mm can be replaced by PEC75 N 857 PEC69 H46721S reusable Biopsy needle guide for probe RRE6 10 needle diameter 1 4 mm N 858 PEC71 H46721D reusable Biopsy needle guide for probe AB2 7 needle diameter 1 mm 1 4 mm 2 2 mm N 859 ...

Страница 300: ...GE MEDICAL SYSTEMS KRETZTECHNIK RAFT VOLUSON 730PRO 730PROV BT 04 DIRECTION KTI105947 REVISION 2 SERVICE MANUAL 9 36 Section 9 14 Biopsy Needle Guides This page was intentionally left blank ...

Страница 301: ...h the procedures before performing a task Table 10 1 Contents in Chapter 10 Section Description Page Number 10 1 Overview 10 1 10 2 Why do Maintenance 10 2 10 3 Maintenance Task Schedule 10 2 10 4 Tools Required 10 5 10 5 System Maintenance 10 6 10 6 Using a Phantom 10 11 10 7 Electrical Safety Tests 10 11 10 8 When There s Too Much Leakage Current 10 24 Ultrasound INSPECTION CERTIFICATE 10 25 CAU...

Страница 302: ...aintained on the site Your GE service representative can help you with establishing performing and maintaining records for a quality assurance program Please contact us for coverage information and or price for service Section 10 3 Maintenance Task Schedule 10 3 1 How often should care maintenance tasks be performed The Care Maintenance Task Schedule see page 10 3 specifies how often your Voluson ...

Страница 303: ...nd Probe holder with a fluid detergent in warm water on a soft damp cloth Be careful not to get the cloth too wet so that moisture does not enter the entire system Mechanical parts Clean and inspect the mechanical function of wheels casters brakes and swivel locks as well as side door foot rest front and rear handle and monitor holder Remove Dust and Coupling gel Mobile Unit Check Daily Control Pa...

Страница 304: ...y your facilities QA program Endocavity Probe Leakage Current Checks Twice Annually Measurement Accuracy Checks Also after corrective maintenance or as required by your facilities QA program Probe Phantom Checks Check axial and lateral resolution see Basic User Manual Technical specifications Check Gain and TGC changes vary the focus and check reaction on screen Also after corrective maintenance o...

Страница 305: ...justable wrist strap 3M 3051 conductive ground cord Anti Static Vacuum Cleaner 46 194427P278 46 194427P279 120V 230V Safety Analyzer 46 285652G1 DALE 600 KIT or equivalent for electrical tests SVHS VCR Cassette E7010GG E7010GF 60 minute 120 minute SVHS VCR Head Cleaner see VCR user manual for requirements 3 5 MOD MEDIA E8381AA E8381AB KTZ207077 blank 128 M disk blank 230 M disk blank 1 3 GB MO dis...

Страница 306: ...y have any problems or questions about the equipment 2 Paperwork Fill in the top of the Periodic Maintenance PM Inspection Certificate Note all probes and system options 3 Power up Turn the system power on and verify that all fans and peripherals turn on Watch the displays during power up to verify that no warning or error messages are displayed 4 Probes Verify that the system properly recognizes ...

Страница 307: ...e basic system controls that affect this mode of operation 5 Doppler Modes Verify basic Doppler Mode operation PW and CW if available Check the basic system controls that affect this mode of operation 6 3D Mode Verify basic 3D Mode operation Check the basic system controls that affect this mode of operation 7 Applicable Software Options Verify the basic operation of all optional modes such as Real...

Страница 308: ... DICOM Verify that DICOM is functioning properly Send an image to a DICOM device 6 Footswitch Verify that the footswitch is functioning as programed Clean as necessary 7 DVD and or MO Drive Verify that the DVD drive and or optional MO drive reads writes properly export recall images in Sonoview 8 Modem Verify modem remote connection 9 ECG Verify basic operation with customer Table 10 7 Mains Cable...

Страница 309: ...elease mechanism Check all wheel locks and wheel swivel locks for proper operation 6 Cables Connectors Check all internal cable harnesses and connectors for wear and secure connector seating Pay special attention to footswitch assembly and probe strain or bend reliefs 7 Shielding Covers Check to ensure that all EMI shielding internal covers air flow panels and screws are in place Missing covers an...

Страница 310: ...cates the probe cannot be used Do a visual check of the probe pins and system sockets before plugging in a probe 10 5 7 3 Basic Probe Cleaning Refer to the Basic User Manual of Voluson 730Pro 730ProV for details on cleaning NOTE To help protect yourself from blood borne diseases wear approved disposable gloves These are made of nitrile derived from vegetable starch to prevent allergic latex reacti...

Страница 311: ...e injury or death in sensitive patients High leakage current can also indicate degradation of insulation and a potential for electrical failure Do not use probes or equipment having excessive leakage current To minimize the risk that a probe may shock someone the customer should Not use a probe that is cracked or damaged in any way Check probe leakage current once a year on surface probes twice a ...

Страница 312: ...e tests are grounding continuity chassis leakage current probe leakage current and ECG leakage current All measurements are made with an electrical safety analyzer Table 10 11 Chassis Leakage Current Limits Accessible Metal Surfaces Country Normal Condition Open Ground Reverse Polarity Open Neutral USA N A 0 3 mA 0 3 mA N A Other 0 1 mA 0 5 mA 0 5 mA 0 5 mA Table 10 12 Type BF Applied Part Leakage...

Страница 313: ...let tester and noting the combination of lights that are illuminated Any problems found should be reported to the hospital immediately and the receptacle should not be used NOTE No outlet tester can detect the condition where the Neutral grounded supply conductor and the Grounding protective earth conductor are reversed If later tests indicate high leakage currents this should be suspected as a po...

Страница 314: ...teps to test the ground wire resistance 1 Turn the Voluson 730Pro 730ProV unit OFF 2 Plug the unit into the meter and the meter into the tested AC wall outlet 3 Plug the black chassis cable into the meter s CHASSIS connector and attach the black chassis cable clamp to an exposed metal part of the Voluson 730Pro 730ProV unit 4 Set the meter s FUNCTION switch to the RESISTANCE position 5 Set the met...

Страница 315: ...uld be made with the unit ON and OFF with the power line polarity Normal and Reversed Record the highest reading of current When using the Microguard or a similar test instrument its power plug may be inserted into the wall outlet and the equipment under test is plugged into the receptacle on the panel of the meter This places the meter in the grounding conductor and the current flowing from the c...

Страница 316: ...cribed for respective test points shown in Table 10 14 8 Keep a record of the results with other hard copies of PM data kept on site Table 10 14 Chassis Leakage Current Test Condition TEST CONDITION 1 Mounting screw for probe receptacle 2 Wheel support 3 Mounting screw for CRT housing 4 Mounting screw for peripheral plugged into unit 5 Mounting screw for other peripheral powered by unit ...

Страница 317: ...e Measurements should be made with the ground open and closed with power line polarity normal and reversed and with the ultrasound console Off and On For each combination the operating controls such as the lead switch should be operated to find the worst case condition Table 10 15 Typical Data Sheet for Chassis Source Leakage Current Unit Power Tester Polarity Switch Tester Neutral or Ground Switc...

Страница 318: ... the power condition combinations found in the table Record the highest leakage current measured Figure 10 4 Test Circuit for Measuring Non Isolated Patient Leads Table 10 16 Testing Power Conditions ECG Power Meter s Polarity Switch Meter s Neutral Switch ON NORM CLOSED ON NORM OPEN ON REVERSE CLOSED ON REVERSE OPEN OFF NORM CLOSED OFF NORM OPEN OFF REVERSE CLOSED OFF REVERSE OPEN POWER OUTLET H ...

Страница 319: ... meter s function selector to the LEAD ISO Select the ALL position on the lead selector Depress the rocker switch to ISO TEST to test lead isolation NOTE It is not necessary to test each lead individually or power condition combinations as required in previous tests CAUTION Line voltage is applied to the ECG leads during this test To avoid possible electric shock hazard the system being tested mus...

Страница 320: ... Source Maximum Allowance Limit GROUND OPEN GROUND CLOSED Patient Lead to Ground Leakage Current Test and Patient Lead to Lead Leakage Current Test 115V 10uA 10uA 220 240V 500uA 10uA Table 10 18 Maximum Allowance Limit for ECG Leakage Current AC Power Source Maximum Allowance Limit Patient Lead Isolation Current Test 115V 20uA 220 240V 5mA Table 10 19 Typical Data Sheet for ECG Leakage Current ECG...

Страница 321: ...easurements should be made with the ground open and closed with power line polarity normal and reversed and with the unit Off and On For each combination the probe must be active to find the worst case condition NOTE Each probe will have some amount of leakage current dependent on its design Small variations in probe leakage currents are normal from probe to probe Other variations will result from...

Страница 322: ...3 Plug the external probe into the meter s EXTERNAL connector 4 Set the meter s FUNCTION switch to EXTERNAL position 5 Connect the probe for test with the connector of the console 6 Add the saline probe and the imaging area of the probe into the saline bath 7 Have unit power ON for the first part turn it OFF for the second half 8 Depress the ISO TEST rocker switch and record the highest current re...

Страница 323: ...ertificate CAUTION Equipment damage possibility Never switch the Polarity and the status of Neutral when the unit is powered ON Be sure to turn the unit power OFF before switching them using the POLARITY switch and or the NEUTRAL switch Otherwise the unit may be damaged Table 10 20 Typical Data Sheet For Transducer Source Leakage Current Transducer Tested Unit Power Tester Power Polarity Switch Te...

Страница 324: ...isolate if the fault lies with the probe or the scanner NOTE Each probe will have some amount of leakage dependent on its design Small variations in probe leakage currents are normal from probe to probe Other variations will result from differences in line voltage and test lead placement The maximum allowable leakage current for body surface contact probe differs from inter cavity probe Be sure to...

Страница 325: ... 1 Frequency Scan Format Model Number Serial Number Probe 2 Frequency Scan Format Model Number Serial Number Probe 3 Frequency Scan Format Model Number Serial Number Probe 4 Frequency Scan Format Model Number Serial Number Probe 5 Frequency Scan Format Model Number Serial Number Probe 6 Frequency Scan Format Model Number Serial Number Probe 7 Frequency Scan Format Model Number Serial Number Probe ...

Страница 326: ...utlet correct ground wiring config System Ground Continuity Chassis Source Leakage Current Probe Chassis Source Leakage Current Wheel Chassis Source Leakage Current CRT Patient Lead Source Leakage Lead to Ground Patient Lead Source Leakage Lead to Lead Patient Lead Source Leakage Isolation Peripheral 1 Leakage Current Peripheral 1Ground Continuity Peripheral 2 Leakage Current Peripheral 2Ground Co...

Страница 327: ...2 FrontEnd Processor 5 21 Top Level Architecture 5 9 Voluson 730Pro 730Pro V 5 10 Voluson 730Pro 730Pro V Basic 5 4 Bluetooth Printer Connection Scheme 3 17 Board Description BackEnd 5 29 FrontEnd 5 22 Boot Up 3 20 4 2 C Calculation Measurements Functional Checks 4 25 General remarks to perform 4 25 Cine Functional Checks 4 21 Color Flow Modes Functional Checks 4 15 Compliance 1 11 Connection Exte...

Страница 328: ... Load 4 37 Loading Procedure 8 8 Save 4 35 Functional Checks 4 1 2D Mode 4 8 Brakes 4 43 Calculation Measurements 4 25 Cine Mode 4 21 Color Flow Modes 4 15 Control Panel 4 4 Doppler Modes 4 13 ECG Checkout 4 42 Generic Measurements 4 22 M Mode 4 11 Mechanical 4 43 MO Drive 4 28 Monitor Display 4 6 Peripherals 4 42 Probes Connector Usage 4 26 Required Equipment 4 1 Volume Modes 4 17 Fuses on CPE Bo...

Страница 329: ...39 Power Distribution CPN6 Primary Power Supply 5 42 CPN80 Primary Power Supply 5 44 Disk Drive Module GEM 5 45 Power Off Shutdown 3 21 4 3 Power On 3 20 4 2 Power Requirements electrical 2 2 Power Stability Requirements 2 3 Preparing the Unit for Installation 3 6 Printer Installation 3 23 Printer Installation manually 3 28 Printer Remote Control Selection 3 38 Printer Settings Adjustment 3 33 Pro...

Страница 330: ... DVD CD R W 7 21 Error Messages 7 23 MO Drive 7 21 Monitor 7 17 Network 7 22 Screen Captures 7 5 Voltage Points 7 4 Troubleshooting Trees Noise disturbs the Image 7 14 Printer 7 19 System does not Power OFF Shutdown 7 16 System does not Power ON Boot Up 7 13 Trackball 7 15 Unable to Record to VCR 7 18 U Ultrasound Application Software UIS Loading Procedure 8 2 Unpacking the Equipment 3 4 Upgrade P...

Страница 331: ......

Страница 332: ... Systems Telex 3797371 P O Box 414 Milwaukee Wisconsin 53201 U S A Asia Pacific Latin America North America GE Ultraschall Tel 49 0 212 28 02 208 Deutschland GmbH Co KG Beethovenstraße 239 Postfach 11 05 60 D 42655 Solingen Germany ...

Отзывы: