GE M
EDICAL
S
YSTEMS
- K
RETZTECHNIK
DRAFT
V
OLUSON
® 730P
RO
/ 730P
RO
V (BT´04)
D
IRECTION
KTI105947, R
EVISION
2
S
ERVICE
M
ANUAL
Chapter 6 - Service Adjustments
6-3
6-3-1-1
Degauss
Degaussing refers to the process of removing magnetic-field effects from the monitor. Operation of the
monitor within a magnetic field may adversely effect color purity. Degaussing can be used to correct the
problem.
1.) Press and hold the
MENU
(middle) key (see:
Figure 6-1
) on the monitor controls for approximately
4 seconds till the OSD menu appears on the screen.
2.) Press the
RIGHT
button 2 times to the DEGAUSS and then press the
MENU
key again to select it.
3.) Wait till DEGAUSS is completed.
6-3-1-2
Color Temp - Calibration
1.) Press the
UTILITIES
key on the Control Panel and then select the SYSTEM item.
2.) Select the
SERVICE
page, enter the password
SHE
and click the ACCEPT button.
3.) Select the MONITOR TEST button in the “Service Tools” menu.
4.) Confirm the “Monitor Test” message with YES.
The Monitor Test main menu appears on the screen.
5.) Select the “Color calibration” icon and press the
right/left trackball
key once (monitor is white).
6.) Press and hold the
MENU
(middle) key on the monitor controls for approximately 4 seconds till the
OSD menu appears on the screen.
7.) Press the
RIGHT
button 3 times to the COLOR TEMP and then press the
MENU
key to select it.
8.) Press the
RIGHT
button until CALIBRATION is highlighted and then press the
MENU
key again.
NOTE:
The calibration cycle will immediately start for Red, Green and Blue. Wait till it´s completion
message “It has completed...” appears. During the Calibration process do not touch any key.
9.) To return to the Main menu, press the
upper trackball
key.
10.)To exit the Monitor Test program, press
EXIT
.
NOTE:
Contrast and Brightness (see:
Section 6-3-1-3
) has to be adjusted after color calibration.
NOTICE
The monitor should automatically degauss itself each time power is applied if you wait at least 10
seconds before you turn power back ON.
Figure 6-2 Monitor Test - Main menu
EXIT
Color calibration
icon