3-38
Section 3-6 - System Configuration
7) Select the patient ID page type from the
Patient ID Page Type
dropdown list, as follows:
•
Type A:
displays patient ID, last and first name, weight, height, BSA, BP and other fields.
•
Type B:
displays all Type A fields except patient ID.
•
Type C:
displays all Type A fields except patient ID, weight, height and BSA. It also displays
the sonographer’s name.
•
Type D:
displays all Type A fields except patient ID, weight, height, BSA and BP.
8) Select the patient data to be displayed in the patient information area of the scanning screen title
bar from the
Title Additional Patient Info
dropdown list.
9) Select the type of media, to which data is stored during backup, from the
Backup Device Is
dropdown list.
10) Select
Eject MO Disk on Shutdown
to have the system automatically eject the backup media
when the unit is shut down.
11) Enter the location of ASCII files in the user (U:\path\) partition of the internal hard disk in the
Export
To Excel Path
field.
12) Select
Enter From Halves to Quad
to enable the user to toggle between viewing a single image,
two images or four images on the screen simultaneously (Quad View).
13) Select
Preview Cine Before Store
to display cineloops before they are stored.
•
When the FlexiView option is installed the time period (in seconds) after which a cineloop is
automatically stored in the archive without interrupting the user’s monitoring must be set in the
FlexiView Time Interval[sec]
field.
Figure 3-23
on page 3-37 indicates the recommended
default factory setup.
14) Trackball to the
OK
button and press
Select
.