3-62
Section 3-7 - Connectivity Setup
3-7-5-2
Validating Communication with the Workstation
1) Press the
Archive
button on the control panel. The
Patient List
window is displayed, as shown below:
Figure 3-44 Patient List Window
2) Select the
Local Archive-Int. HD + DICOM Server
option from the drop-down list at the bottom of
the window. If the system will not accept the dataflow, or displays an error message, there is an
error in dataflow communication.
3) Begin an exam for a new patient and store some images. The storage process will take longer as
each loop is now stored in both raw data and DICOM format.
4) Begin an exam for a new patient. Once the second patient is active, the first patient’s exam is closed
and sent to the spooler to be sent over the network.