Chapter 3 - Installation
3-61
5) If the connection was successfully established, trackball to the
Dicom Storage Server
line and
press
Select
, then trackball to the
Check
button and press
Select
.
•
If the connection is established, a green checkmark is displayed.
•
If the connection cannot be established, a red cross is displayed.
Properties of the DICOM storage server configuration are displayed on the right side of the tab, as reference.
6) If the connection was successfully established, trackball to the
Dicom Print
line and press
Select
,
then trackball to the
Check
button and press
Select
.
•
If the connection is established, a green checkmark is displayed.
•
If the connection cannot be established, a red cross is displayed.
7) If all connections were established successfully, click
OK
to exit the configuration procedure. If not,
review the entire procedure to locate the error.