Chapter 3 - Installation
3-71
3-8-6
Packing the Unit into the Wooden Shipping Crate
Note:
For details of item numbers in this procedure, refer to the following figures:
•
Figure 3-1
on page 3-4
•
Figure 3-3
on page 3-6
•
Figure 3-4
on page 3-7
•
Figure 3-5
on page 3-7.
1) Turn OFF the unit by pressing the ON/OFF button on the control console.
2) Turn OFF the AC circuit breaker on the rear of the unit.
3) Disconnect the mains power plug from the wall outlet.
4) Unplug any probes connected to the unit, and store them in their original cases.
5) Disconnect and remove all external cables from the right and left rear panels and the ECG cable
connected to the front panel.
6) Verify the presence of all the external cables listed in
Table 3-4 on page 3-13
, and pack them in
carton A (item 19).
7) Release and unplug the power cable connector from the AC input at the rear of the unit, and loop
it on the hook located on the rear cover.
8) Unplug the AC power cable from the monitor.
9) Verify the presence of all the accessories listed in
Table 3-5 on page 3-14
, and pack them in carton
B (item 19).
10) Pack the User Manual and any other manuals in an antistatic bag (item 4).
11) Lay the case base on the floor, with the two wood strips facing upwards.
12) Cover the top side of the base with the antistatic sheet 2.2 m x 3m (7.2 ft x 10 ft) (item 2).
13) Lay the top cover/ramp (item 3) on the floor and push it over the narrow side of the case base, in
front of the unit.
Note:
The top cover serves first as a ramp on which to roll the unit onto the crate base, and later
as the crate top cover.
14) Turn the unit’s front wheels inwards and parallel to the unit and lock their swivel movement by
pressing the brake pedal down to the right.
15) Push the unit using its front handle and roll it over the ramp on top of the antistatic sheet that was
previously placed on the crate base, until the front wheels touch the halting wood strips on each
side of the crate base.
16) Move the rear wheels inwards and parallel to the unit and lock all four wheels by pressing the brake
pedal down to the left.
17) Remove the top cover (ramp) (item 3) and put aside.
18) Place three Silica gel bags (item 5) in the space under the unit.
19) Pull the unit’s control console (monitor) up by pressing the release grip located under the center of
the unit’s front handle.
20) Cover the two cartons containing the external cables and accessories with antistatic foam, and lay
them in the space below the keyboard and the unit’s top cover.