Chapter 7 - Diagnostics/Troubleshooting
7-65
7-3-9-6-6 Show History
The
Show History
option in the
View Power
dialog box displays selected test results by
year
,
month
and
day
.
1) Access the
MUX
tab of the
FB Diagnostics
dialog box, as described in the
Accessing the MUX
Diagnostic Test Options
section, on page 7-47.
2) Trackball to the
Advanced board tests
button and press
Select
. The
Advanced System Tests
options are displayed, as shown in
Figure 7-34
on page 7-58.
3) Trackball to the
Monitor
button and press
Select
. The
View Power
options are displayed, as shown
in
Figure 7-36
on page 7-60.
4) Trackball to the
View History
button and press
Select
. The following options are displayed.
Figure 7-41 View History Options
5) Select the test and date that you want to view from the appropriate drop-down list, then trackball to
the
View
button and press
Select
. The relevant graphs are displayed showing activity over time in
values of
Temperature
,
%
, or
Voltage
.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com