GE PROPRIETARY
TO
GE
D
IRECTION
5308917-100, R
EVISION
8
LOGIQ P3 S
ERVICE
M
ANUAL
8-76
Section 8-7 - Power Block - LOGIQ P3
Section 8-7 Power Block - LOGIQ P3
8-7-1
Transformer Assembly (FRU No. 5408517) Cable Guard Clamp (FRU
No:5219747)
This is a description on how to remove and replace the Transformer Assy.
8-7-1-1
Tools
•
Common Phillips screwdrivers
8-7-1-2
Needed Manpower
•
1 persons, 15 minutes
8-7-1-3
Preparations
•
Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system.
NOTE:
For 110V System Configuration use 110V wired Plug (with 3 Amps Fuse).
NOTE:
For 230V System Configuration use 230V Wired plug (With 2 Amps Fuse).
NOTE:
To replace Cable Guard Clamp Donot remove the transformer,remove power cord and press
the cable clamp from both the sides the clamp will come out side.
8-7-1-4
Removal Procedure
1) Remove rear cover. Refer
2) Unscrew two screws (a,b) as marked in
3) Remove GND cable (which ever connected near to “a” ). Refer
Figure 8-83 Removing Transformer Assembly
a
b
GND
cable
Содержание LOGIQ P3
Страница 2: ...GE PROPRIETARY TO GE DIRECTION 5308917 100 REVISION 8 LOGIQ P3 SERVICE MANUAL 1 2 ...
Страница 9: ...GE PROPRIETARY TO GE DIRECTION 5308917 100 REVISION 8 LOGIQ P3 SERVICE MANUAL vii ...
Страница 98: ...GE PROPRIETARY TO GE DIRECTION 5308917 100 REVISION 8 LOGIQ P3 SERVICE MANUAL 3 32 Section 3 10 Paperwork ...
Страница 194: ...GE PROPRIETARY TO GE DIRECTION 5308917 100 REVISION 8 LOGIQ P3 SERVICE MANUAL 7 26 Section 7 6 Troubleshooting ...
Страница 294: ...GE PROPRIETARY TO GE DIRECTION 5308917 100 REVISION 8 LOGIQ P3 SERVICE MANUAL 8 100 Section 8 14 Loading Software ...