GE PROPRIETARY
TO
GE
D
IRECTION
5308917-100, R
EVISION
8
LOGIQ P3 S
ERVICE
M
ANUAL
7-4
Section 7-3 - USB Quick Save
Section 7-3
USB Quick Save
7-3-1
Overview
There may be times when the customer or field engineer will want to directly save images into USB
memory. This is accomplished by saving individual Cine clips (moving images- avi format) or still images
(jpg format) directly to a USB memory disk by pressing a Print Key.
The P3 key is the factory default print key to accomplish the USB Quick Save. However, the default is
for the Image Area only or the customer may have customized the P3 Key function.
7-3-2
Check and Record the P3 Key Function
Check the function of the Print 3 Key in the event that the customer may have made some customized
settings.
1.) Click
Utilit
y on the keyboard.
2.) Select
Connectivity
from the Utilities Menu.
3.) Select the
Buttons
tab on the Connectivity screen.
4.) In the
Physical Print Buttons
field, select Print3.
The Connectivity/Buttons Screen will be displayed like the one shown in
.
P3 is the factory default USB Quick Save key. If it is not set to Image Area, proceed to step 5 to record
the customer’s customized settings.
Figure 7-2 Buttons Set Up Screen
Содержание LOGIQ P3
Страница 2: ...GE PROPRIETARY TO GE DIRECTION 5308917 100 REVISION 8 LOGIQ P3 SERVICE MANUAL 1 2 ...
Страница 9: ...GE PROPRIETARY TO GE DIRECTION 5308917 100 REVISION 8 LOGIQ P3 SERVICE MANUAL vii ...
Страница 98: ...GE PROPRIETARY TO GE DIRECTION 5308917 100 REVISION 8 LOGIQ P3 SERVICE MANUAL 3 32 Section 3 10 Paperwork ...
Страница 194: ...GE PROPRIETARY TO GE DIRECTION 5308917 100 REVISION 8 LOGIQ P3 SERVICE MANUAL 7 26 Section 7 6 Troubleshooting ...
Страница 294: ...GE PROPRIETARY TO GE DIRECTION 5308917 100 REVISION 8 LOGIQ P3 SERVICE MANUAL 8 100 Section 8 14 Loading Software ...