GE PROPRIETARY
TO
GE
D
IRECTION
5308917-100, R
EVISION
8
LOGIQ P3 S
ERVICE
M
ANUAL
8-8
Section 8-2 - Disassembly/Re-assembly of LOGIQ P3
8-2-5
LCD Arm Cover (FRU P/N: 5310093)
This is a description on how to remove and replace the LCD Arm cover set.
8-2-5-1
Tools
•
Common Phillips screwdrivers
8-2-5-2
Needed Manpower
•
1 person, 15 minutes
8-2-5-3
Preparations
•
Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system.
8-2-5-4
Removal procedure
1) Remove LCD Power Cable cover & LCD Signal Cable.
2) Remove LCD Arm cover Assembly .as shown below
Figure 8-8 LCD Arm cover & monitor neck cover
8-2-5-5
Mounting procedure
1.) Install the new parts in the reverse order of removal.
8-2-5-6
Functional Checkout Procedure
Functional Test
Debrief Script
Do Visual Inspection and ensure
no scratches observed and
fitment is done properly
Service Manual Direction 5308917-100, Section 8-2-4. Equipment passes all required
tests and is ready for use.
Arm Cover
Remove two screws of
Arm Cover
a
b
Содержание LOGIQ P3
Страница 2: ...GE PROPRIETARY TO GE DIRECTION 5308917 100 REVISION 8 LOGIQ P3 SERVICE MANUAL 1 2 ...
Страница 9: ...GE PROPRIETARY TO GE DIRECTION 5308917 100 REVISION 8 LOGIQ P3 SERVICE MANUAL vii ...
Страница 98: ...GE PROPRIETARY TO GE DIRECTION 5308917 100 REVISION 8 LOGIQ P3 SERVICE MANUAL 3 32 Section 3 10 Paperwork ...
Страница 194: ...GE PROPRIETARY TO GE DIRECTION 5308917 100 REVISION 8 LOGIQ P3 SERVICE MANUAL 7 26 Section 7 6 Troubleshooting ...
Страница 294: ...GE PROPRIETARY TO GE DIRECTION 5308917 100 REVISION 8 LOGIQ P3 SERVICE MANUAL 8 100 Section 8 14 Loading Software ...