Probe Overview
LOGIQ 500 Basic Users Manual
2233658–100 Rev. 0
15–17
Disinfecting probes (cont’d)
Ultrasound transducers can easily be damaged by improper
handling and by contact with certain chemicals. Failure to
follow these precautions can result in serious injury and
equipment damage.
S
Do not immerse the probe into any liquid beyond the level
specified for that probe. Never immerse the transducer
connector or probe adapters into any liquid.
S
Avoid mechanical shock or impact to the transducer and do
not apply excessive bending or pulling force to the cable.
S
Transducer damage can result from contact with
inappropriate coupling or cleaning agents:
S
Do not soak or saturate transducers with solutions
containing alcohol, bleach, ammonium chloride
compounds or hydrogen peroxide
S
Avoid contact with solutions or coupling gels
containing mineral oil or lanolin
S
Avoid temperatures above 60
°
C.
S
Inspect the probe prior to use for damage or degeneration
to the housing, strain relief, lens and seal. Do not use a
damaged or defective probe.
WARNING
Содержание Logiq 500
Страница 102: ...Exam Applic Preset Selection LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 4 8 This page left blank intentionally ...
Страница 179: ...Mixed Mode Display Formats LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 5 78 This page left blank intentionally ...
Страница 206: ...Annotating an Image LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 6 28 This page left blank intentionally ...
Страница 238: ...General Calculations LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 8 12 This page left blank intentionally ...
Страница 317: ...Advanced Cardiac Calculations LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 10 38 This page left blank intentionally ...
Страница 329: ...Urology Basic Calculations LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 12 4 This page left blank intentionally ...
Страница 364: ...Advanced Recording Option LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 13 36 This page left blank intentionally ...
Страница 555: ...Assistance LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 16 44 This page left blank intentionally ...