Recording Images
LOGIQ 500 Basic Users Manual
2233658–100 Rev. 0
13–26
Disk Verification
In order to protect images from being deleted by other
LOGIQ
500 systems, the disk is always verified by the system
before it can be used.
If the MO disk is not inserted into the drive or an unformatted
MO disk is inserted, the following message is displayed:
“MEDIA IS EITHER UNMOUNTED OR UNFORMATTED.”
When a disk is inserted into a drive, the system first reads the
MO (disk ID) and System ID from the media. These two IDs
are displayed at the bottom of the monitor while the system
verifies that they are registered in the verification list on the
system hard drive.
Storing Images
Display the Archive sub-menu. Select Store Image in the
sub-menu. The following message is displayed during the
storage process:
“In Progress. Please wait.”
System operation is interrupted for five seconds before scan
operation can continue. The image storage operation is then
performed in the “background”.
Image Comments
With the comment cursor in the home position, four characters
can be entered as a comment that is stored with the patient
name, id and date of the scan. Refer to Figure 13–16 on
13–27
for examples.
Optional Method
In addition to the Store Image function in the Soft Menu, the
Record 1 & 2 keys can be programmed in the Set Up/System
Parameter Sub-Menu page 5.
By selecting Image Archive for the Record1 B&W/Color or
Record2 B&W/Color selection, these keys automatically store
the B/W or Color image to MOD when pressed.
Содержание Logiq 500
Страница 102: ...Exam Applic Preset Selection LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 4 8 This page left blank intentionally ...
Страница 179: ...Mixed Mode Display Formats LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 5 78 This page left blank intentionally ...
Страница 206: ...Annotating an Image LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 6 28 This page left blank intentionally ...
Страница 238: ...General Calculations LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 8 12 This page left blank intentionally ...
Страница 317: ...Advanced Cardiac Calculations LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 10 38 This page left blank intentionally ...
Страница 329: ...Urology Basic Calculations LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 12 4 This page left blank intentionally ...
Страница 364: ...Advanced Recording Option LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 13 36 This page left blank intentionally ...
Страница 555: ...Assistance LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 16 44 This page left blank intentionally ...