Probe Overview
LOGIQ 500 Basic Users Manual
2233658–100 Rev. 0
15–9
Probe Usage
For details on connecting, activating, deactivating,
disconnecting, transporting and storing the probes, refer
to
3–29.
Care and Maintenance
Inspecting probes
Inspect the probe’s lens, cable, and casing after each use.
Look for any damage that would allow liquid to enter the probe.
If any damage is found, do not use the probe until it has been
inspected and repaired/replaced by a GE Service
Representative.
Keep a log of all probe maintenance, along with a picture of any
probe malfunction.
Environmental Requirements
Probes should be operated, stored, or transported within the
parameters outlined below.
Operational
Storage
Transport
Temperature
10
_
- 40
_
C
50
_
- 104
_
F
–10
_
- 60
_
C
14
_
- 140
_
F
–40
_
- 60
_
C
–40
_
- 140
_
F
Humidity
30-85%
non-condensing
30-90%
non-condensing
30-90%
non-condensing
Pressure
700-1060hPa
700-1060hPa
700-1060hPa
Table 15–5. Probe Environmental Requirements
After Each Use
.
Содержание Logiq 500
Страница 102: ...Exam Applic Preset Selection LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 4 8 This page left blank intentionally ...
Страница 179: ...Mixed Mode Display Formats LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 5 78 This page left blank intentionally ...
Страница 206: ...Annotating an Image LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 6 28 This page left blank intentionally ...
Страница 238: ...General Calculations LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 8 12 This page left blank intentionally ...
Страница 317: ...Advanced Cardiac Calculations LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 10 38 This page left blank intentionally ...
Страница 329: ...Urology Basic Calculations LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 12 4 This page left blank intentionally ...
Страница 364: ...Advanced Recording Option LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 13 36 This page left blank intentionally ...
Страница 555: ...Assistance LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 16 44 This page left blank intentionally ...