Safety Precautions
LOGIQ 500 Basic Users Manual
2233658–100 Rev. 0
2–17
Best practices while scanning (cont’d)
Controls
Mode
Control
Affect
Default Setting
All
Acoustic Output
Direct.
Significant
The middle setting is a factory preset
determined to be a reasonable setting for all
exams. Use presets to set the output
preferred by scan mode and exam
combination.
B
Focus Comb
Indirect. Minor
Off.
B/M/CFD
Depth
(FOV)
Indirect. Minor
Probe-dependent operator
preset.
B/M
Focal Zone Position
and Number
Indirect. Minor
Probe-dependent system
preset.
B/CFD
Scan Area
Indirect.
Off.
Doppler
M/D Cursor
Indirect. Minor
Off.
Doppler
Doppler Sample
Volume Gate Length
Indirect. Minor
Application-dependent system preset.
Doppler
Velocity Scale
Indirect. Minor
Application-dependent operator preset.
PWD/CFD
Freq. Hi/Low
Indirect. Minor
Low.
CFD
Scan Area
Indirect. Minor
Off.
B/M/CFD
Zoom
Indirect. Minor
Off.
Table 2–3. Controls Affecting Acoustic Output
Acoustic Output Default Levels
In order to assure that an exam does not start at a high output
level, the LOGIQ
500 initiates scanning at a reduced default
output level. This reduced level is preset programmable and
depends upon exam category, application preset and probe
selected. It takes effect when the system is powered on or New
Patient is selected.
Содержание Logiq 500
Страница 102: ...Exam Applic Preset Selection LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 4 8 This page left blank intentionally ...
Страница 179: ...Mixed Mode Display Formats LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 5 78 This page left blank intentionally ...
Страница 206: ...Annotating an Image LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 6 28 This page left blank intentionally ...
Страница 238: ...General Calculations LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 8 12 This page left blank intentionally ...
Страница 317: ...Advanced Cardiac Calculations LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 10 38 This page left blank intentionally ...
Страница 329: ...Urology Basic Calculations LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 12 4 This page left blank intentionally ...
Страница 364: ...Advanced Recording Option LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 13 36 This page left blank intentionally ...
Страница 555: ...Assistance LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 16 44 This page left blank intentionally ...