Adding Color
LOGIQ 500 Basic Users Manual
2233658–100 Rev. 0
5–29
Common Controls
Description
After initial transmission/reception, CFM processing is generally
separate from gray scale processing.
Acoustic output affects the transmit power for both B-Mode and
Color Flow Mapping signals.
TGC, Depth, B/M Gain and Reverse function only on the
B-Mode image displayed.
The M/D Cursor key is not relevant to CFM scanning; it only
affects the Doppler/M-Mode.
Dual format keys (L–R) work the same as in dual B-Mode, but
display two B-Mode and CFM images on the left and right side
of the screen.
NOTE: If “L” and “R” keys are pressed simultaneously, both left
and right B-Mode images will be active. Pressing CFM will then
provide a live B-Mode image on the left and a live B-Mode with
CFM on the right.
Select
If the color flow mode is enabled while in Doppler, the Select
key switches the Doppler/CFM controls between Doppler and
CFM.
.
Содержание Logiq 500
Страница 102: ...Exam Applic Preset Selection LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 4 8 This page left blank intentionally ...
Страница 179: ...Mixed Mode Display Formats LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 5 78 This page left blank intentionally ...
Страница 206: ...Annotating an Image LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 6 28 This page left blank intentionally ...
Страница 238: ...General Calculations LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 8 12 This page left blank intentionally ...
Страница 317: ...Advanced Cardiac Calculations LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 10 38 This page left blank intentionally ...
Страница 329: ...Urology Basic Calculations LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 12 4 This page left blank intentionally ...
Страница 364: ...Advanced Recording Option LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 13 36 This page left blank intentionally ...
Страница 555: ...Assistance LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 16 44 This page left blank intentionally ...