System Care and Maintenance
LOGIQ 500 Basic Users Manual
2233658–100 Rev. 0
16–18
Removing
To remove the console air filters:
1.
Lift up on the Velcro tab securing the air filter.
2.
Pull out the filter.
To remove the power supply air filter:
1.
Pull out and up on the filter cover.
2.
Pull the filter screen out.
3
4
1
2
1
Single Filter Screen Removal
2
Dual Filter Screen Removal
3
Air Filter
4
Retainer
Figure 16–3. Single Filter Screen Removal
Cleaning
To clean the filter:
1.
Be sure to shake the filter in an area away from the
system.
2.
Wash the filter in a mild soapy solution, rinse and air
dry or dry with a cloth.
NOTE: Allow wet filter to dry thoroughly before installing.
3.
Slide the filter back into the system.
4.
Secure the filter with the Velcro tab or cover.
.
Содержание Logiq 500
Страница 102: ...Exam Applic Preset Selection LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 4 8 This page left blank intentionally ...
Страница 179: ...Mixed Mode Display Formats LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 5 78 This page left blank intentionally ...
Страница 206: ...Annotating an Image LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 6 28 This page left blank intentionally ...
Страница 238: ...General Calculations LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 8 12 This page left blank intentionally ...
Страница 317: ...Advanced Cardiac Calculations LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 10 38 This page left blank intentionally ...
Страница 329: ...Urology Basic Calculations LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 12 4 This page left blank intentionally ...
Страница 364: ...Advanced Recording Option LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 13 36 This page left blank intentionally ...
Страница 555: ...Assistance LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 16 44 This page left blank intentionally ...