HEAT OUTPUT
Stand by
Moderate Use (8 cas‐
es / a 10 hour day)
Typical Use (13 cas‐
es / a 10 hour day)
Maximum Use (maxi‐
mum peak power dur‐
ing exam)
Room
Core System
kW
BTU/hr
kW
BTU/hr
kW
BTU/hr
kW
BTU/hr
In room AW TFT monitor 0.12
409
Typical injector
0.09
320
LD Monitor
0.5
1706
Exam room total
1.3
4440
Ctrl
Room
AW work station
0.35
1201
2 AW TFT monitors
0.24
818
Printer
0.31
1054
LD Monitor (5)
0.5
1706
Ctrl room total
1.4
4779
Tech
Room
LD Cabinet
1.0
3412
3kVA Cabinet UPS -
model 9130 (4)
0.37
1257
Tech Room Total
1.37
4669
Typical configuration without fluoro
UPS
10.27
35173
15.15
51640
18.33
62493
23.39
79732
Note (1): Air flow requirements 1200 m
3
/h (706 CFM)
Note (2): For more details, consult appropriate pre–installation manual
Note (3): For UPS 20 kVA option refer to
Note (4): On batteries heat output = 0.52 kW / 1755 BTU
Note (5): The 2nd optional LD monitor is not necessarily in the control room. It may be installed in the exam room (outside patient
vicinity).
CAUTION
Make sure there is a ventilation air flow, preferably ensured by natural air flow,
otherwise by enforced ventilation, so that hydrogen concentration is below 1%
(according to Standard IEC 62040-1-2).
3.2 Fluoro UPS Option
Illustration 4-1:
Innova
TM
IGS 620, Innova
TM
IGS 630 Preinstallation Manual
Direction 5694389-1-1EN, Revision 2