Illustration 2-47: Table baseplate
NOTE:
Prepare the floor such that the Table baseplate will be flush with the floor surface.
For alternative table bolts or seismic area, refer to template drawing shown in
Illustration
Gantry and table mounting holes contained in
.
3.1.4 Hole dimension and preferred location in concrete floor
In the examination room, the Innova Frontal Positioner is not placed on a computer floor but
directly put on concrete floor, the location of the cable access needs to be carefully planned.
Otherwise, if the cable run is located under the concrete floor, the cables will have to come
through the floor and in this case you will need two holes, one for the Frontal Positioner and the
other for the patient table.
The diameter of both holes must be the same 225 mm or 9 in.
Innova
TM
IGS 620, Innova
TM
IGS 630 Preinstallation Manual
Direction 5694389-1-1EN, Revision 2
110
3 Room Structural Requirements