(1): The detector should be stored at 10 to 40 °C (50 to 104 °F) and less than or equal
to 90% RH in the plastic wrapped shipping box. (This should include two bags of
desiccant as well). The lowest temperature (e.g. 10 ºC (50 °F)) and humidity is
preferable. If they are to be stored outside of their shipping box or in the inner
shipping box without plastic wrapping they should be stored at 20 °C (68 °F) or less
and 30% RH or less. In terms of transportation, do not expose to temperatures below
–20 ºC (–4 °F) in its shipping box for more than 15 hours. The detector will reach the
ambient temperature after 20 to 25 hours. The detector should not be allowed to
reach temperatures less than –10 ºC (14 °F) or irreparable damage to the detectors
scintillator will occur. Care must be taken when removing a detector from a shipping
box. If the detector has been subject to cold temperatures for an extended period the
detector in the box should be allowed to sit in the plastic wrapped box to reach room
temperature. This will prevent condensation from occurring. Condensation on the
detector can cause irreparable damage to the electronics. Storage 10 to 40 ºC (50 to
104 °F); 10 to 90 % RH, 250 day storage transportation –20 to +60 ºC (–4 to 140 °F)
and 10 to 80% RH. The Detector chiller is shipped within GE Healthcare packaging.
NOTE:
(2) Special Humidity Instructions: The following parts can be can be shipped in
standard shipment conditions with the requirement that on arrival to installation site,
and before supplying power to these parts, they shall be kept in an environmental
relative humidity equal or lower than their specified capability, and that’s for a
minimum of 48 hours.
•
3kVA UPS 110 V & 220 V (Maximum specified relative humidity capability = 90%)
•
LDM Monitor (Maximum specified relative humidity capability = 90%)
•
CVI Injector (Maximum specified relative humidity capability = 85%)
•
Diamentor (Maximum specified relative humidity capability = 80%)
•
1MP MX191 LCD Monitor (Maximum specified relative humidity capability = 80%)
Table 1-5: Atmospheric Pressure - Storage and transport
SUB-SYSTEM COMPONENTS
ATMOSPHERIC PRESSURE (hPa)
MIN
MAX
Frontal Positioner
525
1013
Lateral Positioner
525
1013
Table Omega IV, Omega V
48
1220
TSUI (TSSC, Smart box, Smart handle, TPD)
500
1010
Footswitch
500
1060
Monitor LCD Eizo
540
1053
Monitor LDM
200
1050
Cabinets C1 Frt, C1 Lat, C2
500
1060
UPS 3KVA
697
1130
Power Distribution Box (PDB) CE
500
1060
Power Distribution Box (PDB) UL
500
1060
Innova
TM
IGS 620, Innova
TM
IGS 630 Preinstallation Manual
Direction 5694389-1-1EN, Revision 2
44
4 Product Storage and Handling Requirements