3
Operating instructions
48
Ettan IPGphor 3 Instrument User Manual 11-0034-58 Edition AB
3.5.4
Pausing or stopping a protocol
To pause or stop a protocol in progress, press the
STOP
key:
To continue the protocol, press the
START
key. To cancel the protocol, press the
STOP
key again. This will cause the instrument to respond as though the end of
an uninterrupted protocol had been reached as described above.
3.5.5
Printing a log of instrument conditions
To document and save run protocols and data during a run, use the
Ettan IPGphor 3 Control Software.
An abbreviated log illustrates the format.
Ettan IPGphor 3 Control Software may also be used to log and graph and print the
output.
In addition the output may be exported to a text or spreadsheet file such as
Microsoft Excel.
P a u s e d i n s t e p 1
P r e s s S TA R T t o c o n t i n u e
--- IPGphor 3
Serial No: 4711
Firmware version: A2 02/20/04
------------------------------------------
Date: Mar 14, 2005
------------------------------------------
17.52.21 User cttcjg logged on
17.52.21
Protocol: Test
No. of strips: 12
IEF Params: 50uA/Strip at 20 C
Step 1: Step 100V 0:01 Hr
Step 2: Grad 500V 0:01 Hr
Step 3: Grad 8000V 0:01 Hr
Step 4: Step 8000V 0:01 Hr
17.52.21 Step 1 started
17.53.21 Step 2 started
17.54.22 Step 3 started
17.55.23 Step 4 started
17.56.25 Run finished
Содержание Ettan IPGphor 3
Страница 1: ...GE Healthcare Ettan IPGphor 3 Instrument User Manual Ettan...
Страница 2: ......
Страница 4: ......
Страница 54: ...4 Care and maintenance 54 Ettan IPGphor 3 Instrument User Manual 11 0034 58 Edition AB...
Страница 62: ...6 Customer service information 62 Ettan IPGphor 3 Instrument User Manual 11 0034 58 Edition AB...
Страница 63: ......