ERROR TYPE
– The deviation from which the
transfer characteristic is to be
calculated. This can be one of the
following:
% Span
- a percentage of the Input signal span.
% Rdg
- a percentage of the Input signal reading.
•
LINEARITY
- The transfer characteristic from
Reference to Input signal. This can be
one of the following:
Linear
- a proportional response.
Square Root
- often found in flow sensors.
•
TOLERANCE
– The test limits for the deviation from the
transfer characteristic.
06 June 2013
K0541 issue 1
Page 88 of 175