Table 8-17 Calibration pressures
Ranges: gage
Nominal applied pressure
mbar (psi)
-FS †
Zero
+FS
< 700 mbar (10.0 psi)
-FS
0
+FS
> 700 mbar (10.0 psi)
-900 (-13.1)
0
+FS
†For a three-point calibration, do not apply more than -
90% of the specified FS for the unit.
Ranges: absolute
Nominal applied pressure
mbar (psi)
Zero
+FS
350 mbar (5.00 psi)
< 1.0 (0.02)
+FS
2 bar (30.0 psi)
< 5.0 (0.07)
+FS
7 bar (100.0 psi)
< 20.0 (0.29)
+FS
20 bar (300.0 psi)
< 50.0 (0.73)
+FS
350 bar (5000 psi)
Use atmospheric
pressure as zero.
+FS
06 June 2013
K0541 issue 1