User Manual
Druck DPI620 Genii
8
Calibration Procedures
Note
: GE Measurement and Control can provide a
calibration service that is traceable to international
standards.
Note
: GE Measurement and Control recommend returning
the instrument to the manufacturer or an approved service
agent for calibration. If using an alternative calibration
facility, check that it uses the standards that follow.
Before Starting
8.1
Use only original parts supplied by the manufacturer. To do
an accurate calibration, use the following:
• Calibration equipment specified in Table 8-1.
• Stable temperature environment: 21 ± 1°C (70 ± 2°F).
Before starting a calibration procedure, it is recommended
leaving the equipment in the calibration environment for a
minimum of two hours.
06 June 2013
K0541 issue 1