User Manual
Druck DPI620 Genii
Procedures (CH1): Frequency Amplitude (source)
8.7
Do the procedure as follows:
Note: The procedure that follows calibrates the “mark” value
of the square wave frequency output. The “space” value is
fixed and is approximately -120 mV.
1.
Connect the applicable calibration equipment (Ref:
2.
Let the equipment get to a stable temperature
(minimum: 5 minutes since the last power on).
3.
Set-up the equipment as follows:
Source frequency = 0 (For DC output);
Waveform = Square
4.
Use the calibration menu (Ref: Section 8.1) to do a two-
point calibration.
Point 1 = 0.2 V, point 2 = 20 V
06 June 2013
K0541 issue 1
Page 154 of 175