User Manual
Druck DPI620 Genii
4.
Check the calibration is correct
•
Select the applicable Resistance (measure) function
(Ref Section 2.1153).
•
Make a 4-wire connection to the applicable standard
resistor (Ref: Table 8-11) and measure the value.
•
Check the error is within limits (Ref: Table 8-11)
Table 8-11 Resistance (measure) error limits
Standard
Resistor
(
Ω
)
Resistor
uncertainty
(
Ω
)
Permitted
DPI 620 error
(
Ω
)
0 (short
-
0.02
100
0.002
0.032
200
0.004
0.044
300
0.006
0.056
400
0.008
0.068
1000
0.02
0.30
2000
0.04
0.41
4000
0.08
0.64
06 June 2013
K0541 issue 1
Page 157 of 175