![GE Druck DPI 620 Genii Скачать руководство пользователя страница 103](http://html.mh-extra.com/html/ge/druck-dpi-620-genii/druck-dpi-620-genii_user-manual_128379103.webp)
User Manual
Druck DPI620 Genii
Executing Methods
6.12
Not all HART® devices have the same methods available.
The function, purpose and execution of each method can
be different. Methods may include the following:
•
Self-test
•
Loop test
•
Sensor trim
•
D/A trim
Execute a method as follows:
1.
Select the method name. A screen opens with
information about the selected method. Four buttons
are displayed on the screen:
HELP
Displays a description of the method.
ABORT
Exits from the procedure.
OK
Accepts inputs and proceeds to the
next step.
SWITCH APP:
Returns to Druck DPI 620 Genii screen
(to change channel function settings
without interrupting the method
procedure).
Note:
Some methods may make the HART® device output a
certain current.
A warning will appear on the screen before the device
enters this mode.
06 June 2013
K0541 issue 1
Page 103 of 175