User Manual
Druck DPI620 Genii
Figure 2-23 CH1 Thermocouple Source CH2 Switch Test
Figure 2-23 shows a thermocouple switch test.
1.
Set the applicable software options:
•
The TC is set to source a temperature
•
The
UTILITY
is set to
SWITCH TEST
. The AUTOMATION is
set to
RAMP
2.
Complete the electrical connections.
3.
It is a CH1 function, the switch connections must be on
CH2.
4.
For the Ramp process, set
START
and
STOP
values that
are applicable to the switch value.
5.
To get an accurate switch value, set a long
TRAVEL
period.