K0430 Issue No. 2
Current Output Range (24 mA)
Use the following procedure for calibrating the current output range.
(1)
Switch the instrument on, enter
CALIBRATION
mode and select
INT RANGES
from
the
CALIBRATION
menu as shown on page 43.
(2)
Select
CURRENT OUT
range for calibration:
(3)
Connect a digital milliammeter to the instrument as shown below.
Note:
This procedure requires an external supply.
55mA
Vin
mA in
mA Sink
max 30V
max
mA
-
+
Voltage
source
Calibration
Содержание Druck DPI 615 IS
Страница 3: ...K0430 Issue No 2 ii ...
Страница 66: ...57 K0430 Issue No 2 Hydraulic Calibrator Versions 3 ...
Страница 67: ...K0430 Issue No 2 58 This page intentionally left blank ...
Страница 73: ...K0430 Issue No 2 64 This page intentionally left blank ...
Страница 74: ...65 K0430 Issue No 2 Appendix 1 DATA LOG FILE EXAMPLE ...
Страница 75: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com ...
Страница 78: ...69 K0430 Issue No 2 Appendix 2 ATEX Certificate of Conformity ...
Страница 79: ...K0430 Issue No 2 70 intentionally left blank ...
Страница 80: ...71 K0430 Issue No 2 ...
Страница 81: ...K0430 Issue No 2 72 ...
Страница 82: ...73 K0430 Issue No 2 ...
Страница 83: ...K0430 Issue No 2 74 ...
Страница 84: ...75 K0430 Issue No 2 ...
Страница 85: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com ...
Страница 86: ...77 K0430 Issue No 2 ...
Страница 87: ...K0430 Issue No 2 78 ...
Страница 88: ...79 K0430 Issue No 2 ...
Страница 89: ...K0430 Issue No 2 80 ...
Страница 90: ...81 K0430 Issue No 2 ...
Страница 91: ...K0430 Issue No 2 82 ...
Страница 92: ...83 K0430 Issue No 2 ...
Страница 93: ...K0430 Issue No 2 84 ...
Страница 94: ...85 K0430 Issue No 2 ...
Страница 95: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com ...