K0430 Issue No. 2
14
Current to Pressure Converter (I-P) Task
A - External pressure source
B - Pressure to current device (24V)
C - Barrier
D - Safe area
E - External supply
•
Use the
up
and
down
cursor keys to adjust the loop current to the required
value. Alternatively, press
ENTER
and use cursor keys to enter a finite value.
Cursor keys can then be used to nudge the output either up or down.
If required, change pressure units with INPUT key. A flashing CHECK LOOP
message indicates either an open circuit supply loop (or no external supply).
Pressure Switch Test (P-SWITCH) Task
A - External pressure source
B - Pressure regulator
C - Pressure switch under test
•
Contact state will be shown on display. When contacts close, buzzer sounds.
•
To run switch test, close vent valve and press the
RUN
(F1) key.
•
Using the hand-pump, increase the applied pressure to just below the switch
operating point. Screw the volume adjuster in until the switch operates (the
operating pressure of the switch is then written to the display).
•
Reduce pressure until the switch releases (indicated by the switch symbol). The
release pressure is then written to the display and the hysteresis displayed.
P
I
-
PRESSURE INT
bar
OUTPUT
NEW VALUE
CURRENT
mA
OK
TASK : I-P
SNAPSHOT MODE
F2
F1
EXIT
Max 30V
CAT II
DPI 615 IS
CD
3
+
-
+
-
C
C
E
D
A
B
PRESSURE INT
bar
F2
F1
RUN
CONTACT STATE
max 30V
TASK : P-SWITCH
SNAPSHOT MODE
DPI 615 IS
A
B
3
C
Taking Measurements
Содержание Druck DPI 615 IS
Страница 3: ...K0430 Issue No 2 ii ...
Страница 66: ...57 K0430 Issue No 2 Hydraulic Calibrator Versions 3 ...
Страница 67: ...K0430 Issue No 2 58 This page intentionally left blank ...
Страница 73: ...K0430 Issue No 2 64 This page intentionally left blank ...
Страница 74: ...65 K0430 Issue No 2 Appendix 1 DATA LOG FILE EXAMPLE ...
Страница 75: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com ...
Страница 78: ...69 K0430 Issue No 2 Appendix 2 ATEX Certificate of Conformity ...
Страница 79: ...K0430 Issue No 2 70 intentionally left blank ...
Страница 80: ...71 K0430 Issue No 2 ...
Страница 81: ...K0430 Issue No 2 72 ...
Страница 82: ...73 K0430 Issue No 2 ...
Страница 83: ...K0430 Issue No 2 74 ...
Страница 84: ...75 K0430 Issue No 2 ...
Страница 85: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com ...
Страница 86: ...77 K0430 Issue No 2 ...
Страница 87: ...K0430 Issue No 2 78 ...
Страница 88: ...79 K0430 Issue No 2 ...
Страница 89: ...K0430 Issue No 2 80 ...
Страница 90: ...81 K0430 Issue No 2 ...
Страница 91: ...K0430 Issue No 2 82 ...
Страница 92: ...83 K0430 Issue No 2 ...
Страница 93: ...K0430 Issue No 2 84 ...
Страница 94: ...85 K0430 Issue No 2 ...
Страница 95: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com ...