47
K0430 Issue No. 2
Voltage Input Range (30 Volts)
Use the following procedure for calibrating the internal 30 Volt range.
(1)
Switch the instrument on, enter
CALIBRATION
mode and select
INT RANGES
from
the
CALIBRATION
menu as shown on page 43.
(2)
Select 30V range for calibration:
(3)
Link the voltage input terminals with a short lead and enter the zero point:
(4)
Remove the shorting link from the voltmeter terminals, connect the Vin terminals
of the instrument to a variable voltage source and connect a digital voltmeter
across the source:
SELE
C
T
C
AL
OP
TI
O
N
SELE
C
T
R
ANGE
30
V
O
L
TS
5
V
O
LTS
SELE
C
T
C
AL
OP
TI
O
N
C
ALI
BR
ATI
O
N
V
O
L
T
A
GE
IN
C
U
RR
ENT IN
C
U
RR
ENT
O
UT
PR
ESSU
R
E INT
TEM
P
E
R
ATU
R
E
C
ALI
BR
ATING V
O
LTAGE
R
ANGE +/-
30
V
A
pp
ly
no
mi
na
l z
e
r
o
v
a
l
ue
P
r
ess
ENTE
R
k
e
y
w
h
en
app
li
ed
v
a
l
ue
i
s stab
l
e
SET VALUE
C
HANGE
VALUE
R
UN
C
ALI
BR
ATING V
O
LTAGE
R
ANGE +/-
30
V
C
ALI
BR
ATING V
O
LTAGE
R
ANGE +/-
30
V
E
nte
r A
pp
li
ed
V
a
l
ue:
St
ore zero poin
t
?
YES
N
O
55mA
Vin
mA in
mA Sink
max 30V
max
+
-
Cal V
DVM
Calibration
Содержание Druck DPI 615 IS
Страница 3: ...K0430 Issue No 2 ii ...
Страница 66: ...57 K0430 Issue No 2 Hydraulic Calibrator Versions 3 ...
Страница 67: ...K0430 Issue No 2 58 This page intentionally left blank ...
Страница 73: ...K0430 Issue No 2 64 This page intentionally left blank ...
Страница 74: ...65 K0430 Issue No 2 Appendix 1 DATA LOG FILE EXAMPLE ...
Страница 75: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com ...
Страница 78: ...69 K0430 Issue No 2 Appendix 2 ATEX Certificate of Conformity ...
Страница 79: ...K0430 Issue No 2 70 intentionally left blank ...
Страница 80: ...71 K0430 Issue No 2 ...
Страница 81: ...K0430 Issue No 2 72 ...
Страница 82: ...73 K0430 Issue No 2 ...
Страница 83: ...K0430 Issue No 2 74 ...
Страница 84: ...75 K0430 Issue No 2 ...
Страница 85: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com ...
Страница 86: ...77 K0430 Issue No 2 ...
Страница 87: ...K0430 Issue No 2 78 ...
Страница 88: ...79 K0430 Issue No 2 ...
Страница 89: ...K0430 Issue No 2 80 ...
Страница 90: ...81 K0430 Issue No 2 ...
Страница 91: ...K0430 Issue No 2 82 ...
Страница 92: ...83 K0430 Issue No 2 ...
Страница 93: ...K0430 Issue No 2 84 ...
Страница 94: ...85 K0430 Issue No 2 ...
Страница 95: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com ...