13
K0430 Issue No. 2
Pressure Converter (Pressure to Pressure) Task
Select the P-P task from the task menu and connect the Unit Under Test (UUT) to the
calibrator as shown below. Testing a converter requires one pressure to be applied to
the unit under test (UUT) and another (converter output) to be measured. The additional
measurement is provided by the external transducer option.
Method
•
Connect the UUT to the instrument as shown below. Plug the external transducer
into the instrument as shown below:
A - External pressure source (indicator instruments only)
B - Pressure regulator
C - External pressure source
D - Pressure to pressure device
E - External transducer
•
Press the TASK key and select the P-P task. Providing the external transducer has
been calibrated and its parameters stored in the instrument, the display will
show
External
pressure in the input window and calibrator
Output
pressure in
the output window. If an error message “NO SENSOR OR CAL INVALID” is
displayed, this indicates that the external transducer has not been entered and/
or calibrated with the instrument. Refer to page 56 for details of adding an
external transducer.
If an external transducer change is made, switch the
calibrator off and then on to load new transducer data.
•
If required, select the input and output units as described on page 10.
•
If applicable, turn on Cal Mode and set-up test parameters as detailed on page
11.
USE
RS232
IN SAFE
AREA
ONLY
3
A
B
C
D
E
Taking Measurements
Содержание Druck DPI 615 IS
Страница 3: ...K0430 Issue No 2 ii ...
Страница 66: ...57 K0430 Issue No 2 Hydraulic Calibrator Versions 3 ...
Страница 67: ...K0430 Issue No 2 58 This page intentionally left blank ...
Страница 73: ...K0430 Issue No 2 64 This page intentionally left blank ...
Страница 74: ...65 K0430 Issue No 2 Appendix 1 DATA LOG FILE EXAMPLE ...
Страница 75: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com ...
Страница 78: ...69 K0430 Issue No 2 Appendix 2 ATEX Certificate of Conformity ...
Страница 79: ...K0430 Issue No 2 70 intentionally left blank ...
Страница 80: ...71 K0430 Issue No 2 ...
Страница 81: ...K0430 Issue No 2 72 ...
Страница 82: ...73 K0430 Issue No 2 ...
Страница 83: ...K0430 Issue No 2 74 ...
Страница 84: ...75 K0430 Issue No 2 ...
Страница 85: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com ...
Страница 86: ...77 K0430 Issue No 2 ...
Страница 87: ...K0430 Issue No 2 78 ...
Страница 88: ...79 K0430 Issue No 2 ...
Страница 89: ...K0430 Issue No 2 80 ...
Страница 90: ...81 K0430 Issue No 2 ...
Страница 91: ...K0430 Issue No 2 82 ...
Страница 92: ...83 K0430 Issue No 2 ...
Страница 93: ...K0430 Issue No 2 84 ...
Страница 94: ...85 K0430 Issue No 2 ...
Страница 95: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com ...