63
K0430 Issue No. 2
Flushing - Replenishing or Changing the Hydraulic Fluid (Figure A3)
If necessary, to remove any contaminants, flush out the hydraulic system as follows.
a.
Connect the bleed hose assembly (1) to the priming port and a bleed hose to the
pressure port as shown below:
1
Bleed hose assembly
2
Priming pump
3
Screw-press
4
Isolation valve
O -
Open
C
-
Closed
Figure A3 - Flushing/Filling Connections
b.
Fill the priming fluid container with clean hydraulic fluid of the required type. Use
the priming pump (2), to pump fresh hydraulic fluid through the system until clean
hydraulic fluid, free of air bubbles, emerges into the container at the output port.
Discard the contaminated fluid expelled during this process.
c.
Remove the bleed hose assembly (1) located to the output port and, to prevent
the ingress of any contaminant, fit a blanking plug in its place.
d.
Close the isolation valve (4) and remove the bleed hose assembly (1) from the
priming inlet.
e.
Clean off any surplus oil that may have spilled onto the instrument casing.
The instrument is now ready for operation or storage. If storing, apply a label detailing
the type of hydraulic fluid contained in the instrument.
For long term storage, it is
recommended that the instrument be drained and stored empty.
1
4
2
3
3
S
Hydraulic Calibrator Versions Operation
Содержание Druck DPI 615 IS
Страница 3: ...K0430 Issue No 2 ii ...
Страница 66: ...57 K0430 Issue No 2 Hydraulic Calibrator Versions 3 ...
Страница 67: ...K0430 Issue No 2 58 This page intentionally left blank ...
Страница 73: ...K0430 Issue No 2 64 This page intentionally left blank ...
Страница 74: ...65 K0430 Issue No 2 Appendix 1 DATA LOG FILE EXAMPLE ...
Страница 75: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com ...
Страница 78: ...69 K0430 Issue No 2 Appendix 2 ATEX Certificate of Conformity ...
Страница 79: ...K0430 Issue No 2 70 intentionally left blank ...
Страница 80: ...71 K0430 Issue No 2 ...
Страница 81: ...K0430 Issue No 2 72 ...
Страница 82: ...73 K0430 Issue No 2 ...
Страница 83: ...K0430 Issue No 2 74 ...
Страница 84: ...75 K0430 Issue No 2 ...
Страница 85: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com ...
Страница 86: ...77 K0430 Issue No 2 ...
Страница 87: ...K0430 Issue No 2 78 ...
Страница 88: ...79 K0430 Issue No 2 ...
Страница 89: ...K0430 Issue No 2 80 ...
Страница 90: ...81 K0430 Issue No 2 ...
Страница 91: ...K0430 Issue No 2 82 ...
Страница 92: ...83 K0430 Issue No 2 ...
Страница 93: ...K0430 Issue No 2 84 ...
Страница 94: ...85 K0430 Issue No 2 ...
Страница 95: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com ...