GE
HEALTHCARE
D
IRECTION
2307224-100, R
EVISION
6
D
ISCOVERY
LS S
YSTEM
S
ERVICE
M
ANUAL
Chapter 6 - Table
Page 173
4. Refer to
.
a.) If necessary, loosen the flag hardware, and turn the adjusters to center the flag over the
notch in the CT and PET detent.
b.) Measure the distance between the leading edges of the CT and PET flags.
c.) If necessary, slightly loosen the PET detent hardware, and turn the PET Detent Adjuster
until the distance equals the value entered on
d.) Tighten the detent hardware.
Figure 6-87:
Adjust PET Detent for Recorded Distance Between Scan Planes
6.2.6.33 Assemble and Install PET Detent Adjuster
1.) Locate and remove the following parts from the upgrade kit:
2.) Refer to
. Use the illustration as a guide to assemble the PET detent adjuster.
Figure 6-88:
PET Detent Adjuster, P/N 2332092
Slightly loosen
detent hardware.
Turn screw to adjust distance
distance between PET and
CT detent locations.
Part Number
Description
Quantity
2332092
Adjuster Basis (Block)
1
2103580-46
M5 x 20 mm Flat-Head Countersunk Screw
2
46-328425P2
M6 Hex Nut
1
2345323-2
M6 x 50 mm Set Screw
1
Содержание Discovery LS
Страница 1: ...GE Healthcare OPERATING DOCUMENTATION 2307224 100 Rev 6 Discovery LS System Service Manual ...
Страница 2: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 2 ...
Страница 24: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 24 Table of Contents ...
Страница 32: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 32 List of Figures ...
Страница 110: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 110 Chapter 5 Console ...
Страница 222: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 222 Chapter 6 Table ...
Страница 242: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 242 Chapter 7 Gantry ...
Страница 357: ......