Appendix B - System Calibration Quick Guide
8. Clean Database
20 Minutes
CT Network Queue
:
1. Image Works
2. Network
3. Select Remote Host
4. Ping Dicom Host
5. Queue
6. Network
7. Delete old exams or images.
CT Recon Queue
:
1. Recon Management
2. Restart Queue
CT Archive
:
1. Image Works
2. Queue
3. Archive
4. Scroll through images.
5. Delete old exams or images.
PET Data
:
1. Utilities
2. Delete
3. Type of Data = Patient
4. Sort by = Date-Time
5. Include = All
6. Select all patients except from the last week.
7. Select
8. Delete Selection.
9. Delete Items.
This process deletes all data from memory
except for data from the last week. It is
recommended to back up all data before
clean-up.
If not able to perform a backup before
the clean-up, perform a partial clean-up
by deleting all CT data entered in the
PET database and for which records
still exist in the CT database.
Содержание Discovery LS
Страница 1: ...GE Healthcare OPERATING DOCUMENTATION 2307224 100 Rev 6 Discovery LS System Service Manual ...
Страница 2: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 2 ...
Страница 24: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 24 Table of Contents ...
Страница 32: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 32 List of Figures ...
Страница 110: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 110 Chapter 5 Console ...
Страница 222: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 222 Chapter 6 Table ...
Страница 242: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 242 Chapter 7 Gantry ...
Страница 357: ......