GE H
EALTHCARE
D
IRECTION
2307224-100, R
EVISION
6
D
ISCOVERY
LS S
YSTEM
S
ERVICE
M
ANUAL
Page 106
Chapter 4 - Service Desktop, Tools and Diagnostics
Refer to
System Resets
function allows resetting and downloading the scanner
hardware as required to prepare the system for the scanning operation. Access
Systems Resets
as follows:
1.) Select
Systems Resets
from the Service Desktop. The reset applications selections include:
a.)
Scan
- Resets and downloads all controllers in the gantry and table.
b.)
Recon
- Resets the recon subsystem (IG, RIP, and SDC boards).
c.)
DAS/COLL
- Resets the DAS control and collimator control board.
d.)
Data Acquisition
- Resets the DAS, control, collimator control, and DIP boards.
2.) Select
Reset
on the displayed screen.
3.) Select
Run
.
The status box displays the status of the selected reset.
Figure 4-4:
Systems Reset
Click the appropriate button to reset an application.
Содержание Discovery LS
Страница 1: ...GE Healthcare OPERATING DOCUMENTATION 2307224 100 Rev 6 Discovery LS System Service Manual ...
Страница 2: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 2 ...
Страница 24: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 24 Table of Contents ...
Страница 32: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 32 List of Figures ...
Страница 110: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 110 Chapter 5 Console ...
Страница 222: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 222 Chapter 6 Table ...
Страница 242: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 242 Chapter 7 Gantry ...
Страница 357: ......