Chapter 8 - System QC, Calibration and Functional Tests
Page 249
5.) Refer to
. From the Patient List, click the entry representing the calibration
acquisition. A list of image sets displays.
Figure 8-8:
Images Database Navigator Showing Image Sets List
6.) From the Image Sets list in
, select the two image sets that represent the CT and
PET slices and click [
OK
]. The Image Alignment Table (shown in
displays.
Figure 8-9:
Image Alignment Table
Note:
The Image Alignment Table displays system messages in a box at the bottom of the panel. These
messages are for the User (Operator) only, and are related to the software realignment process.
PET
Image
CT
Image
Use Z values for CT-PET table distance correction.
Measurements in red are
outside acceptable limits.
Not for service use.
Содержание Discovery LS
Страница 1: ...GE Healthcare OPERATING DOCUMENTATION 2307224 100 Rev 6 Discovery LS System Service Manual ...
Страница 2: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 2 ...
Страница 24: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 24 Table of Contents ...
Страница 32: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 32 List of Figures ...
Страница 110: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 110 Chapter 5 Console ...
Страница 222: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 222 Chapter 6 Table ...
Страница 242: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 242 Chapter 7 Gantry ...
Страница 357: ......