GE
HEALTHCARE
D
IRECTION
2307224-100, R
EVISION
6
D
ISCOVERY
LS S
YSTEM
S
ERVICE
M
ANUAL
Chapter 6 - Table
Page 129
6.2.5.10 Remove the Motor Clutch Assembly (P/N 2309067)
For best results, check the table and secondary base with a level before removing the motor clutch
assembly. Check level again, before fastening the covers into place. If necessary, adjust the table
levelers to re-level the table and secondary base.
To disassemble the secondary base drive belt:
1.) Refer to
. Use two strips of masking tape, or some white correction fluid, to mark
the edges of the belt connector before disassembling the belt.
Figure 6-22:
Mark the Secondary Base Drive Belt
2.) Refer to
. Use a 4 mm Allen wrench to remove the four socket-head cap screws
from the fastener plate, and set aside.
3.) Remove the fastener plate and set aside.
4.) Remove the drive belt from the pulley on the motor clutch assembly.
Figure 6-23:
Disassemble Secondary Base Positioning Belt
Socket-Head Cap Screws (4) and Fastener Plate
Содержание Discovery LS
Страница 1: ...GE Healthcare OPERATING DOCUMENTATION 2307224 100 Rev 6 Discovery LS System Service Manual ...
Страница 2: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 2 ...
Страница 24: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 24 Table of Contents ...
Страница 32: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 32 List of Figures ...
Страница 110: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 110 Chapter 5 Console ...
Страница 222: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 222 Chapter 6 Table ...
Страница 242: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 242 Chapter 7 Gantry ...
Страница 357: ......