Revision A
ProCare
Patient Monitor Service Manual
4-19
2009381-001
Calibration & Maintenance:
4.7 Alarm Code Interpretation
4.7 Alarm Code Interpretation
If any other alarms appear that are not listed in the paragraphs that follow,
record the error message and report the failure to Customer Support. Refer
to the Operation Manual for information about patient alarms and general
procedural alarms.
4.7.1 System Failures
When a system failure is encountered, the error code is displayed on the
screen for five seconds and the system enters failsafe mode. The error code
is logged in the history log.
General system error codes are listed below.
Alarm Conditions and Error Codes
When responding to a Monitor alarm, always CHECK THE PATIENT FIRST
and then check the Monitor, cuff, hose and sensors. Press
SILENCE
to
reset patient alarm conditions.
NIBP
SpO
2
Alarm
Definition
Possible Cause
E89
NIBP No Determination
Unable to make an NIBP determination due to
insufficient signal.
E84
Timeout
Determination time > 2 minutes.
Motion Artifact.
E85
Timeout
One cuff pressure at > 1 minute.
Motion artifact.
E83
Timeout: Inflation
Inflation time is > 40 seconds
Possible air leak is being detected.
E82
Excess Air in Cuff
Residual air in cuff above threshold for successful
auto-zero.
E80
Overpressure
Overpressure condition detected
Alarm
Definition
Possible Cause
E20
SpO
2
No Sensor
SpO
2
sensor not connected.
No sensor code detected.
Sensor failure
E21
SpO
2
Replace Sensor
SpO2 sensor or cable possibly defective.
Cable not connected properly.
Содержание DINAMAP ProCare 100
Страница 1: ......
Страница 4: ...DINAMAP ProCare Monitor Service Manual 2009381 001 Revision A ...
Страница 9: ...ProCare Patient Monitor Service Manual Revision A 2009381 001 Table of Contents ...
Страница 10: ...Revision A ProCare Patient Monitor Service Manual 1 1 2009381 001 1 Introduction ...
Страница 11: ...1 2 ProCare Patient Monitor Service Manual Revision A 2009381 001 ...
Страница 18: ...Revsion A ProCare Patient Monitor Service Manual 1 9 2009381 001 Introduction 1 4 Product Description ...
Страница 19: ...1 10 ProCare Patient Monitor Service Manual Revision A 2009381 001 Introduction 1 4 Product Description ...
Страница 20: ...Revsion A ProCare Partient Monitor Service Manual 2 1 2009381 001 2 Product Description ...
Страница 21: ...2 2 ProCare Partient Monitor Service Manual Revision A 2009381 001 For your notes ...
Страница 35: ...2 16 ProCare Partient Monitor Service Manual Revision A 2009381 001 Product Description 2 6 Specifications ...
Страница 36: ...Revision A ProCare Patient Monitor Service Manual 3 1 2009381 001 3 Principles of Operation ...
Страница 37: ...3 2 ProCare Patient Monitor Service Manual Revision A 2009381 001 For your notes ...
Страница 46: ...Revision A ProCare Patient Monitor Service Manual 4 1 2009381 001 4 Calibration Maintenance ...
Страница 47: ...4 2 ProCare Patient Monitor Service Manual Revision A 2009381 001 For your notes ...
Страница 83: ...PNEUMATIC TUBING ROUTING Exploded Assembly Drawing ProCare Monitor front case page 5 3 4 or ...
Страница 96: ...2009381 001 A ...