4-6
ProCare
Patient Monitor Service Manual
Revision A
2009381-001
Calibration & Maintenance:
4.3 Periodic Maintenance
4.3 Periodic Maintenance
4.3.1 As Required
Perform the following maintenance procedures as required.
4.3.1.1 Integrity of Hoses and Cuffs
When the pneumatic integrity of any NIBP cuff and hose is in doubt, replace
the cuff and hose, and discard the questionable accessories.
4.3.1.2 Cleaning of Monitor
CAUTION: Do not clean Monitor with isopropyl alcohol or other solvents.
Wipe the exterior of the Monitor with a cloth slightly dampened with mild
detergent or normal hospital bactericides. Use dishwashing detergents such
as IVORY and JOY (registered trademarks of Procter & Gamble Corp.), or
PALMOLIVE (registered trademark of Colgate-Palmolive Corp.)
Do not immerse unit.
4.3.1.3 Cleaning of Accessories
Clean the adult cuffs supplied for use with the monitor by hand washing in
warm, soapy water. However, take care to avoid entry of water into the cuff
and hoses at any time. If water enters the cuff, dry the cuff by passing air
through it.
The neonatal cuffs are for single patient use - discard if they become soiled.
Clean cuffs and hoses with a cloth slightly dampened with mild detergent.
Do not immerse hoses.
Do not immerse cuffs without prior application of cuff hose caps.
Clean SpO2 sensor surface before and after each patient use. Clean SpO2
sensor with a cloth slightly dampened with a mild detergent. Wipe SpO2
sensor to ensure all detergent residue has been removed.
Compatible cleaning and disinfecting solutions are:
Chlorine bleach disinfectant, 5.25%, 0.75 cup per gallon of water.
CAUTION: Do not apply isopropyl alcohol to the Monitor - some parts can
become marred and cracked.
Содержание DINAMAP ProCare 100
Страница 1: ......
Страница 4: ...DINAMAP ProCare Monitor Service Manual 2009381 001 Revision A ...
Страница 9: ...ProCare Patient Monitor Service Manual Revision A 2009381 001 Table of Contents ...
Страница 10: ...Revision A ProCare Patient Monitor Service Manual 1 1 2009381 001 1 Introduction ...
Страница 11: ...1 2 ProCare Patient Monitor Service Manual Revision A 2009381 001 ...
Страница 18: ...Revsion A ProCare Patient Monitor Service Manual 1 9 2009381 001 Introduction 1 4 Product Description ...
Страница 19: ...1 10 ProCare Patient Monitor Service Manual Revision A 2009381 001 Introduction 1 4 Product Description ...
Страница 20: ...Revsion A ProCare Partient Monitor Service Manual 2 1 2009381 001 2 Product Description ...
Страница 21: ...2 2 ProCare Partient Monitor Service Manual Revision A 2009381 001 For your notes ...
Страница 35: ...2 16 ProCare Partient Monitor Service Manual Revision A 2009381 001 Product Description 2 6 Specifications ...
Страница 36: ...Revision A ProCare Patient Monitor Service Manual 3 1 2009381 001 3 Principles of Operation ...
Страница 37: ...3 2 ProCare Patient Monitor Service Manual Revision A 2009381 001 For your notes ...
Страница 46: ...Revision A ProCare Patient Monitor Service Manual 4 1 2009381 001 4 Calibration Maintenance ...
Страница 47: ...4 2 ProCare Patient Monitor Service Manual Revision A 2009381 001 For your notes ...
Страница 83: ...PNEUMATIC TUBING ROUTING Exploded Assembly Drawing ProCare Monitor front case page 5 3 4 or ...
Страница 96: ...2009381 001 A ...