4-18
ProCare
Patient Monitor Service Manual
Revision A
2009381-001
Calibration & Maintenance:
4.6 ProCare Patient Monitor Parameter Tests
4.6.13 Printer Output Test
1. Load Thermal Paper into the print mechanism
2. Press PRINT button
3. Verify the printer outputs a record and print quality is good
4.6.14 Communication Port Test
1. Connect unit to a PC terminal emulator.
2. Turn unit on.
3. Type " NC0!E", press "Enter"
4. Verify response from unit " NC+!@"
5. Verify that within 40 seconds the pump starts.
6. Type " ND!5", press "Enter"
7. Verify response from unit " ND+!A"
8. Verify the pump stops.
Содержание DINAMAP ProCare 100
Страница 1: ......
Страница 4: ...DINAMAP ProCare Monitor Service Manual 2009381 001 Revision A ...
Страница 9: ...ProCare Patient Monitor Service Manual Revision A 2009381 001 Table of Contents ...
Страница 10: ...Revision A ProCare Patient Monitor Service Manual 1 1 2009381 001 1 Introduction ...
Страница 11: ...1 2 ProCare Patient Monitor Service Manual Revision A 2009381 001 ...
Страница 18: ...Revsion A ProCare Patient Monitor Service Manual 1 9 2009381 001 Introduction 1 4 Product Description ...
Страница 19: ...1 10 ProCare Patient Monitor Service Manual Revision A 2009381 001 Introduction 1 4 Product Description ...
Страница 20: ...Revsion A ProCare Partient Monitor Service Manual 2 1 2009381 001 2 Product Description ...
Страница 21: ...2 2 ProCare Partient Monitor Service Manual Revision A 2009381 001 For your notes ...
Страница 35: ...2 16 ProCare Partient Monitor Service Manual Revision A 2009381 001 Product Description 2 6 Specifications ...
Страница 36: ...Revision A ProCare Patient Monitor Service Manual 3 1 2009381 001 3 Principles of Operation ...
Страница 37: ...3 2 ProCare Patient Monitor Service Manual Revision A 2009381 001 For your notes ...
Страница 46: ...Revision A ProCare Patient Monitor Service Manual 4 1 2009381 001 4 Calibration Maintenance ...
Страница 47: ...4 2 ProCare Patient Monitor Service Manual Revision A 2009381 001 For your notes ...
Страница 83: ...PNEUMATIC TUBING ROUTING Exploded Assembly Drawing ProCare Monitor front case page 5 3 4 or ...
Страница 96: ...2009381 001 A ...