Calibration & Maintenance:
4.6 ProCare Patient Monitor Parameter Tests
9. Calibration verification is complete. Disconnect the probe simulator and
install the temperature probe. If the Monitor does not pass the calibration
verification, then the Monitor needs repair.
4.6.12 SpO
2
(Perform only if equipped with SpO
2
module)
Note for Monitors equipped with Nellcor SpO
2
:
Nellcor is aware that various Nellcor oximetry platforms have some
interesting behaviors when used with various pulse simulators.
On occasion when testing the integrity of the Nellcor oximetry system,
abnormal results may occur when introducing large changes in the pulse
rate and/or pulse amplitude. Extreme changes in rate sent to the Nellcor
sensor by the SpO
2
simulator may cause the SpO
2
algorithm to completely
miss in finding the pulse rate.
This is an expected result. To workaround this, incrementally step up or
down the settings on your SpO
2
simulator and allow for the Monitor to detect
and display the new pulse rate or saturation.
Nellcor recommends use of the SRCmax Portable Tester for use with
ProCare Monitors equipped with the Nellcor SpO
2
system.
Masimo recommends BIO-TEK SpO
2
simulators.
1. Connect the appropriate SpO
2
simulator and cable to the SpO
2
connector.
2. Verify the unit displays a:
Pulse value
Saturation value
Signal Strength Bar Graph
3. Disconnect the SpO
2
cable
4. Verify the unit generates a
alarm and speaker is sounding
5. Press the
SILENCE
button.
6. Verify the sound has stopped but the error display remains
7. Re-connect the SpO
2
sensor
8. Verify the unit displays a:
Pulse Value
Saturation value
Signal Strength bar Graph
Содержание DINAMAP ProCare 100
Страница 1: ......
Страница 4: ...DINAMAP ProCare Monitor Service Manual 2009381 001 Revision A ...
Страница 9: ...ProCare Patient Monitor Service Manual Revision A 2009381 001 Table of Contents ...
Страница 10: ...Revision A ProCare Patient Monitor Service Manual 1 1 2009381 001 1 Introduction ...
Страница 11: ...1 2 ProCare Patient Monitor Service Manual Revision A 2009381 001 ...
Страница 18: ...Revsion A ProCare Patient Monitor Service Manual 1 9 2009381 001 Introduction 1 4 Product Description ...
Страница 19: ...1 10 ProCare Patient Monitor Service Manual Revision A 2009381 001 Introduction 1 4 Product Description ...
Страница 20: ...Revsion A ProCare Partient Monitor Service Manual 2 1 2009381 001 2 Product Description ...
Страница 21: ...2 2 ProCare Partient Monitor Service Manual Revision A 2009381 001 For your notes ...
Страница 35: ...2 16 ProCare Partient Monitor Service Manual Revision A 2009381 001 Product Description 2 6 Specifications ...
Страница 36: ...Revision A ProCare Patient Monitor Service Manual 3 1 2009381 001 3 Principles of Operation ...
Страница 37: ...3 2 ProCare Patient Monitor Service Manual Revision A 2009381 001 For your notes ...
Страница 46: ...Revision A ProCare Patient Monitor Service Manual 4 1 2009381 001 4 Calibration Maintenance ...
Страница 47: ...4 2 ProCare Patient Monitor Service Manual Revision A 2009381 001 For your notes ...
Страница 83: ...PNEUMATIC TUBING ROUTING Exploded Assembly Drawing ProCare Monitor front case page 5 3 4 or ...
Страница 96: ...2009381 001 A ...