Revision A
ProCare
Partient Monitor Service Manual
2-9
2009381-001
Product Description:
2.4. Host Communications Connector
2.4. Host Communications Connector
All host port signals are NON-ISOLATED and should be connected to
equipment conforming to IEC 601-1 ONLY. Where isolation of data
communication is required, the isolated level converter should be used. If
external alarm control is required, p/n 487208 (Isolated Remote Alarm Cable
Assembly) should ALWAYS be used. Please refer to the Information Sheet
included with the isolated remote alarm cable for operational details.
Note:
When using remote alarm, the
ProCare
Monitor should be considered
the primary alarm source. The secondary alarm is used for secondary
purposes only.
2.4.1. DB15 Connector Pin Assignments
Connection Details
Host Port Connector (rear panel)
WARNING! Auxiliary equipment
connected to the DINAMAP
®
ProCare
Monitor will result in the formation of
an electromedical system and thus
must comply with the requirements of EN 60601-1-1/ IEC 601-1. All host port
signals are NON-ISOLATED and should be connected to equipment
conforming to IEC-601-1, configured to comply with IEC 601-1-1 ONLY.
Where isolation of data communication is required, GE Medical Systems
Information Technologies
part number ILC1926 should be used. If external
alarm control is required, GE Medical Systems
Information Technologies
part number 487208 (Isolated Remote Alarm Cable Assembly) should
ALWAYS be used. When a high-priority condition is displayed on the
Monitor, the remote alarm signal becomes active within 0.5 seconds. The
active state of the alarm signal is an open circuit. In the inactive state the
alarm signal is connected to ground. Please refer to the Information Sheet
included with the isolated remote alarm cable for operational details.
Note:
When using remote alarm, the
ProCare
Monitor should be considered
the primary alarm source. The secondary alarm is used for secondary
purposes only.
Pin #
Function
1
Common
2
Inverted TTL Transmit Data
3
Inverted TTL Receive Data
4
+5 volts
5
No connection
6
No connection
7
Common
8
Remote Alarm
9
No connection
10
No connection
11
RS232 Transmit Data (TxD)
12
No connection
13
RS232 Receive Data (RxD)
14
No connection
15
No connection
Содержание DINAMAP ProCare 100
Страница 1: ......
Страница 4: ...DINAMAP ProCare Monitor Service Manual 2009381 001 Revision A ...
Страница 9: ...ProCare Patient Monitor Service Manual Revision A 2009381 001 Table of Contents ...
Страница 10: ...Revision A ProCare Patient Monitor Service Manual 1 1 2009381 001 1 Introduction ...
Страница 11: ...1 2 ProCare Patient Monitor Service Manual Revision A 2009381 001 ...
Страница 18: ...Revsion A ProCare Patient Monitor Service Manual 1 9 2009381 001 Introduction 1 4 Product Description ...
Страница 19: ...1 10 ProCare Patient Monitor Service Manual Revision A 2009381 001 Introduction 1 4 Product Description ...
Страница 20: ...Revsion A ProCare Partient Monitor Service Manual 2 1 2009381 001 2 Product Description ...
Страница 21: ...2 2 ProCare Partient Monitor Service Manual Revision A 2009381 001 For your notes ...
Страница 35: ...2 16 ProCare Partient Monitor Service Manual Revision A 2009381 001 Product Description 2 6 Specifications ...
Страница 36: ...Revision A ProCare Patient Monitor Service Manual 3 1 2009381 001 3 Principles of Operation ...
Страница 37: ...3 2 ProCare Patient Monitor Service Manual Revision A 2009381 001 For your notes ...
Страница 46: ...Revision A ProCare Patient Monitor Service Manual 4 1 2009381 001 4 Calibration Maintenance ...
Страница 47: ...4 2 ProCare Patient Monitor Service Manual Revision A 2009381 001 For your notes ...
Страница 83: ...PNEUMATIC TUBING ROUTING Exploded Assembly Drawing ProCare Monitor front case page 5 3 4 or ...
Страница 96: ...2009381 001 A ...