2-14
ProCare
Partient Monitor Service Manual
Revision A
2009381-001
Product Description:
2.6. Specifications
Sensor Connect/Disconnect from Monitor
The monitor detects the attachment or disconnection of a
sensor from the Monitor within 5 seconds.
Pulse Detection
The Monitor detects a pulse of enters a no-signal state
within 15 seconds of being attached to a patient.
Loss of Pulse
the monitor detects loss of pulse from patient and enters a
no signal state within 10 seconds.
Nellcor Sensors - SpO
2
Range 70% to 100%
OxiMAX Sensor Models Single Patient Use
MAX-A,* MAX-AL*
±2
MAX-N
†
*
±2
MAX-P*
±2
MAX-I*
±2
MAX-R
‡
±3.5
OxiCliq Sensor Models Single Patient Use
OxiCliq A
±2.5
OxiCliq P
±2.5
OxiCliq N
†
±2.5
Reusable Sensor Models
D-YS
±3
D-YS & D-YSE
±3.5
D-YS & DYSPD
±3.5
DS-100A
±3
OXI-A/N
±3
OXI-P/I
±3
* The accuracy specification under motion conditions is ±3. For a definition of motion contact GE Medical Services
Technical Support.
† The MAX-N and the OxiCliq N were tested on patients > 40 kg.
‡ The accuracy specification has been determined between saturations of 80-100%
SpO
2
- Masimo Oximetry
Saturation Range
1% - 100%
Pulse Rate and Accuracy
25-240 BPM ±3 digits
Perfusion Range
0.02 to 20%
SpO
2
- Nellcor Pulse Oximetry
Содержание DINAMAP ProCare 100
Страница 1: ......
Страница 4: ...DINAMAP ProCare Monitor Service Manual 2009381 001 Revision A ...
Страница 9: ...ProCare Patient Monitor Service Manual Revision A 2009381 001 Table of Contents ...
Страница 10: ...Revision A ProCare Patient Monitor Service Manual 1 1 2009381 001 1 Introduction ...
Страница 11: ...1 2 ProCare Patient Monitor Service Manual Revision A 2009381 001 ...
Страница 18: ...Revsion A ProCare Patient Monitor Service Manual 1 9 2009381 001 Introduction 1 4 Product Description ...
Страница 19: ...1 10 ProCare Patient Monitor Service Manual Revision A 2009381 001 Introduction 1 4 Product Description ...
Страница 20: ...Revsion A ProCare Partient Monitor Service Manual 2 1 2009381 001 2 Product Description ...
Страница 21: ...2 2 ProCare Partient Monitor Service Manual Revision A 2009381 001 For your notes ...
Страница 35: ...2 16 ProCare Partient Monitor Service Manual Revision A 2009381 001 Product Description 2 6 Specifications ...
Страница 36: ...Revision A ProCare Patient Monitor Service Manual 3 1 2009381 001 3 Principles of Operation ...
Страница 37: ...3 2 ProCare Patient Monitor Service Manual Revision A 2009381 001 For your notes ...
Страница 46: ...Revision A ProCare Patient Monitor Service Manual 4 1 2009381 001 4 Calibration Maintenance ...
Страница 47: ...4 2 ProCare Patient Monitor Service Manual Revision A 2009381 001 For your notes ...
Страница 83: ...PNEUMATIC TUBING ROUTING Exploded Assembly Drawing ProCare Monitor front case page 5 3 4 or ...
Страница 96: ...2009381 001 A ...