![GE DINAMAP ProCare 100 Скачать руководство пользователя страница 52](http://html.mh-extra.com/html/ge/dinamap-procare-100/dinamap-procare-100_service-manual_128373052.webp)
Revision A
ProCare
Patient Monitor Service Manual
4-7
2009381-001
Calibration & Maintenance:
4.3 Periodic Maintenance
Cidex Formula 7 (registered trademark of Johnson & Johnson Medical
Products, Inc.) or pHisoHex (registered trademark of Winthrop-Breon
Laboratories.)
Quaternary-based germicidal detergents like VESTAL INSURANCE
(registered trademark of the Vestal Corp.), HI-TOR PLUS (registered
trademark of the Huntington Corp.), or VIREX (registered trademark of S.C.
Johnson & Son Corp.)
For the above, follow manufacturers' recommendations for dilution rate and
use. These recommendations are not an endorsement of the manufacturers
or of the effectiveness of these materials for cleaning or disinfecting.
4.3.1.4 Long-Term Storage
If it becomes necessary to store the Monitor for an extended period of time,
remove all attached accessories. Attach the original packing inserts, and
place the monitor into the original shipping container.
Long-term storage of a discharged lead-acid battery can permanently
degrade its storage capacity. Therefore, the battery should be fully charged
before storage.
Batteries in storage should be charged every six months to maintain their
capacity.
Elevated temperatures will shorten battery life and can lead to permanent
battery damage.
4.3.2 Annual Procedures
Perform the test procedures described in section 4.6 every twelve months,
or whenever the accuracy of of the monitor is in doubt.
Содержание DINAMAP ProCare 100
Страница 1: ......
Страница 4: ...DINAMAP ProCare Monitor Service Manual 2009381 001 Revision A ...
Страница 9: ...ProCare Patient Monitor Service Manual Revision A 2009381 001 Table of Contents ...
Страница 10: ...Revision A ProCare Patient Monitor Service Manual 1 1 2009381 001 1 Introduction ...
Страница 11: ...1 2 ProCare Patient Monitor Service Manual Revision A 2009381 001 ...
Страница 18: ...Revsion A ProCare Patient Monitor Service Manual 1 9 2009381 001 Introduction 1 4 Product Description ...
Страница 19: ...1 10 ProCare Patient Monitor Service Manual Revision A 2009381 001 Introduction 1 4 Product Description ...
Страница 20: ...Revsion A ProCare Partient Monitor Service Manual 2 1 2009381 001 2 Product Description ...
Страница 21: ...2 2 ProCare Partient Monitor Service Manual Revision A 2009381 001 For your notes ...
Страница 35: ...2 16 ProCare Partient Monitor Service Manual Revision A 2009381 001 Product Description 2 6 Specifications ...
Страница 36: ...Revision A ProCare Patient Monitor Service Manual 3 1 2009381 001 3 Principles of Operation ...
Страница 37: ...3 2 ProCare Patient Monitor Service Manual Revision A 2009381 001 For your notes ...
Страница 46: ...Revision A ProCare Patient Monitor Service Manual 4 1 2009381 001 4 Calibration Maintenance ...
Страница 47: ...4 2 ProCare Patient Monitor Service Manual Revision A 2009381 001 For your notes ...
Страница 83: ...PNEUMATIC TUBING ROUTING Exploded Assembly Drawing ProCare Monitor front case page 5 3 4 or ...
Страница 96: ...2009381 001 A ...