Revision A
ProCare
Partient Monitor Service Manual
2-13
2009381-001
Product Description:
2.6. Specifications
Classification Information
Mode of Operation: Continuous degree of Protection
against harmful ingress of water: Drip-Proof IPX1.
Mechanical
NIBP
CUFF PRESSURE RANGE
Adult
Neonate
0 mmHg to 290 mmHg
0 mmHg to 145 mmHg
DEFAULT TARGET:
CUFF INFLATION
Adult
Neonate
150 ± 15 mmHg
110 ± 15 mmHg
TARGET CUFF INFLATION
ADJUSTMENT RANGE
Adult
Neonate
100 - 250 mmHg (5mmHg Steps)
100 - 140 mmHg (5 mmHg Steps)
BLOOD PRESSURE
DETERMINATION TIME
Adult
Neonate
120 seconds maximum
85 seconds maximum
PULSE RATE: RANGE
Adult
Neonate
30 - 200 BPM ±3%
30 - 220 BPM ±3%
BLOOD PRESSURE:
MEASUREMENT RANGES
Systolic
mmHg
MAP
mmHg
Diastolic
mmHg
Adult
Neonate
30 - 290
30 - 140
20 - 260
20 - 125
10 - 220
10 - 110
NIBP ACCURACY
Meets AAMI/ANSI standard SP-10
AAMI/ANSI standard: ±5 mmHg mean error
Intra-arterial method: ± 8 mmHg standard dev.
Temperature - Alaris Thermometry
SCALES
Fahrenheit
Celsius
RANGE
Max
Min
42.2° Celsius 108.0° Fahrenheit
31.6° Celsius 88.9° Fahrenheit
Monitor Mode Accuracy
±0.1° C
± 0.2° F (when tested in a calibrated liquid bath; meets
ASTM E1112, Table 1 in range specified)
Predictive Mode Accuracy
±0.6° C
±1.0° F
Determination Time
Less then 60 seconds
SpO
2
- Nellcor Pulse Oximetry
SpO
2
Range and Accuracy
adult/neonate: 70-100% ±3.5 digits
Pulse Rate Range and Accuracy
30 - 250 BPM ±3 BPM
Saturation Pitch Indicator
Pitch changes with saturation
Sensor Connect/Disconnect from Patient
The Monitor detects the connection or disconnection or a
sensor to the patient within 15 seconds
Содержание DINAMAP ProCare 100
Страница 1: ......
Страница 4: ...DINAMAP ProCare Monitor Service Manual 2009381 001 Revision A ...
Страница 9: ...ProCare Patient Monitor Service Manual Revision A 2009381 001 Table of Contents ...
Страница 10: ...Revision A ProCare Patient Monitor Service Manual 1 1 2009381 001 1 Introduction ...
Страница 11: ...1 2 ProCare Patient Monitor Service Manual Revision A 2009381 001 ...
Страница 18: ...Revsion A ProCare Patient Monitor Service Manual 1 9 2009381 001 Introduction 1 4 Product Description ...
Страница 19: ...1 10 ProCare Patient Monitor Service Manual Revision A 2009381 001 Introduction 1 4 Product Description ...
Страница 20: ...Revsion A ProCare Partient Monitor Service Manual 2 1 2009381 001 2 Product Description ...
Страница 21: ...2 2 ProCare Partient Monitor Service Manual Revision A 2009381 001 For your notes ...
Страница 35: ...2 16 ProCare Partient Monitor Service Manual Revision A 2009381 001 Product Description 2 6 Specifications ...
Страница 36: ...Revision A ProCare Patient Monitor Service Manual 3 1 2009381 001 3 Principles of Operation ...
Страница 37: ...3 2 ProCare Patient Monitor Service Manual Revision A 2009381 001 For your notes ...
Страница 46: ...Revision A ProCare Patient Monitor Service Manual 4 1 2009381 001 4 Calibration Maintenance ...
Страница 47: ...4 2 ProCare Patient Monitor Service Manual Revision A 2009381 001 For your notes ...
Страница 83: ...PNEUMATIC TUBING ROUTING Exploded Assembly Drawing ProCare Monitor front case page 5 3 4 or ...
Страница 96: ...2009381 001 A ...