![GE CGA 351 Скачать руководство пользователя страница 73](http://html.mh-extra.com/html/ge/cga-351/cga-351_user-manual_128345073.webp)
December 2004
4-34
Setup and Calibration
Adjusting the Calibration
Setting (cont.)
At the next prompt, enter the correct value for the oxygen
concentration in ppm. If desired, refer to Appendix A,
The Nernst
Equation
, to determine the corresponding sensor output in mV.
Setup
Cal
Diag
Use the [ ] and [ ] keys to
select
[Cal]
and press
[ENTER]
.
Slct Cal Menu Item
POWER
ESC
ENTER
Setup
Cal
Diag
Use the [
] and [
] keys to
select
[Heater]
and press
[ENTER]
.
Heater
POWER
ESC
ENTER
PPM-Percent
Use the [
] and [
] keys to
select
[PPM]
and press
[ENTER]
.
PPM
Percent
POWER
ESC
ENTER
PPM-Percent PPM
Using the [ ] and [ ] keys,
position the cursor under the
desired character. Then, use
the [
] and [
] keys to incre-
ment the value. When done,
press
[ENTER]
.
PPM
000.00
s
POWER
ESC
ENTER
PPM.. Invalid Output
Use the [ ] and [ ] keys to
select
[Yes]
and press
[ENTER]
.
Modifying value may
cause invalid output
Yes
No
POWER
ESC
ENTER
www.DataSheet4U.com
www.DataSheet4U.com
Содержание CGA 351
Страница 7: ...Chapter 1 www DataSheet4U com www DataSheet4U com ...
Страница 15: ...Chapter 2 www DataSheet4U com www DataSheet4U com ...
Страница 27: ...Chapter 3 www DataSheet4U com www DataSheet4U com ...
Страница 38: ...Chapter 4 www DataSheet4U com www DataSheet4U com ...
Страница 77: ...Chapter 5 www DataSheet4U com www DataSheet4U com ...
Страница 104: ...Chapter 6 www DataSheet4U com www DataSheet4U com ...
Страница 109: ...Appendix A www DataSheet4U com www DataSheet4U com ...
Страница 115: ...Appendix B www DataSheet4U com www DataSheet4U com ...
Страница 116: ...CE Mark Compliance Overview B 1 EMC Compliance B 1 LVD Compliance B 1 www DataSheet4U com www DataSheet4U com ...
Страница 118: ...Appendix C www DataSheet4U com www DataSheet4U com ...