December 2004
4-22
Setup and Calibration
Setting the Zero Trim
Using the procedure under “Testing the Analog Output” on page
4-21, set the
Test Percent
to 0.
At the analog output setup menu, proceed as follows:
Note:
The
[Aout Type], [Zero Setpoint]
and
[Span Trim]
options are also
available in the above list box.
At the next prompt, enter the desired value for the zero trim.
The meter now automatically returns to the “Analog Output” prompt.
Using the procedure under “Testing the Analog Output” on page
4-21, reset the
Test Percent
to its normal value.
Either press
[ESC]
until you exit the
User Program
or proceed to the
appropriate section to continue setting up the analog output.
User Aout
Use the [
] and [
] keys to
select
[Zero Trim]
, and press
[ENTER]
.
Span Setpoint
Test
Zero Trim
POWER
ESC
ENTER
User Aout Zero-Trim
Use [
] and [
] to increment
the value. Press
[ENTER]
when
done.
Zero Trim
s
0
t
POWER
ESC
ENTER
www.DataSheet4U.com
www.DataSheet4U.com
Содержание CGA 351
Страница 7: ...Chapter 1 www DataSheet4U com www DataSheet4U com ...
Страница 15: ...Chapter 2 www DataSheet4U com www DataSheet4U com ...
Страница 27: ...Chapter 3 www DataSheet4U com www DataSheet4U com ...
Страница 38: ...Chapter 4 www DataSheet4U com www DataSheet4U com ...
Страница 77: ...Chapter 5 www DataSheet4U com www DataSheet4U com ...
Страница 104: ...Chapter 6 www DataSheet4U com www DataSheet4U com ...
Страница 109: ...Appendix A www DataSheet4U com www DataSheet4U com ...
Страница 115: ...Appendix B www DataSheet4U com www DataSheet4U com ...
Страница 116: ...CE Mark Compliance Overview B 1 EMC Compliance B 1 LVD Compliance B 1 www DataSheet4U com www DataSheet4U com ...
Страница 118: ...Appendix C www DataSheet4U com www DataSheet4U com ...