![GE CGA 351 Скачать руководство пользователя страница 60](http://html.mh-extra.com/html/ge/cga-351/cga-351_user-manual_128345060.webp)
Setup and Calibration
4-21
December 2004
Testing the Analog Output
Enter the analog output setup menu, as described on page 4-16, and
proceed as follows:
Note:
The
[Aout Type], [Zero Trim]
and
[Span Trim]
options are also
available in the above list box.
At the next prompt, enter the desired “Test Percent” value, which is
the analog output value (expressed as a percent of the analog output
span) used by the meter as a
TEST
signal.
The meter now automatically returns to the “Analog Output” prompt.
Either press
[ESC]
until you exit the
User Program
or proceed to the
appropriate section to continue setting up the analog output.
User Aout
Use the [
] and [
] keys to
select
[Test]
, and press
[ENTER]
.
Zero Setpoint
Span Setpoint
Test
POWER
ESC
ENTER
User Aout Aout-Test
Use [
] and [
] to increment
the value. Press
[ENTER]
when
done.
Test Percent
s
0
t
POWER
ESC
ENTER
www.DataSheet4U.com
www.DataSheet4U.com
Содержание CGA 351
Страница 7: ...Chapter 1 www DataSheet4U com www DataSheet4U com ...
Страница 15: ...Chapter 2 www DataSheet4U com www DataSheet4U com ...
Страница 27: ...Chapter 3 www DataSheet4U com www DataSheet4U com ...
Страница 38: ...Chapter 4 www DataSheet4U com www DataSheet4U com ...
Страница 77: ...Chapter 5 www DataSheet4U com www DataSheet4U com ...
Страница 104: ...Chapter 6 www DataSheet4U com www DataSheet4U com ...
Страница 109: ...Appendix A www DataSheet4U com www DataSheet4U com ...
Страница 115: ...Appendix B www DataSheet4U com www DataSheet4U com ...
Страница 116: ...CE Mark Compliance Overview B 1 EMC Compliance B 1 LVD Compliance B 1 www DataSheet4U com www DataSheet4U com ...
Страница 118: ...Appendix C www DataSheet4U com www DataSheet4U com ...