General Information
The Sensor Enclosure
The sensor enclosure, which is shown in Figure 1-2 below, houses the
sample measurement components. The functions of these components
are as follows:
•
a zirconium oxide (zirconia)
oxygen sensor
converts the oxygen
concentration of the sample gas into a mV output signal.
•
a
sensor furnace
heats the oxygen sensor and the sample gas to the
optimum temperature of 700°C.
•
a
thermocouple
precisely regulates the temperature of the oxygen
sensor and the sample gas.
•
an inlet
needle valve
and an outlet
flowmeter
adjust the sample gas
flow to 1,000 ± 150 cc/min (2.1 ± 0.3 SCFH).
Figure 1-2: Sensor Enclosure
Furnace Bracket
Flowmeter
Thermocouple
Needle Valve
Sensor Furnace
Oxygen Sensor
Manifold
Electrical Conduit
Flowmeter
Furnace Bracket
Flowmeter
Thermocouple
Needle Valve
Sensor Furnace
Oxygen Sensor
Manifold
Electrical Conduit
www.DataSheet4U.com
www.DataSheet4U.com
Содержание CGA 351
Страница 7: ...Chapter 1 www DataSheet4U com www DataSheet4U com ...
Страница 15: ...Chapter 2 www DataSheet4U com www DataSheet4U com ...
Страница 27: ...Chapter 3 www DataSheet4U com www DataSheet4U com ...
Страница 38: ...Chapter 4 www DataSheet4U com www DataSheet4U com ...
Страница 77: ...Chapter 5 www DataSheet4U com www DataSheet4U com ...
Страница 104: ...Chapter 6 www DataSheet4U com www DataSheet4U com ...
Страница 109: ...Appendix A www DataSheet4U com www DataSheet4U com ...
Страница 115: ...Appendix B www DataSheet4U com www DataSheet4U com ...
Страница 116: ...CE Mark Compliance Overview B 1 EMC Compliance B 1 LVD Compliance B 1 www DataSheet4U com www DataSheet4U com ...
Страница 118: ...Appendix C www DataSheet4U com www DataSheet4U com ...